[1]张菁菁,徐海飞,李统怀,郑哲岚.声辐射力脉冲成像联合应力式弹性成像诊断子宫肌瘤变性的应用价值[J].临床超声医学杂志,2020,22(11):843-846.
ZHANG Jingjing, XU Haifei, LI Tonghuai, et al. Application value of acoustic radiation force impulse imaging combined with stress elastography in the diagnosis of hysteromyoma degeneration[J]. Journal of Clinical Ultrasound in Medicine, 2020,22(11):843-846.
[1]张菁菁,徐海飞,李统怀,郑哲岚.声辐射力脉冲成像联合应力式弹性成像诊断子宫肌瘤变性的应用价值[J].临床超声医学杂志,2020,22(11):843-846. DOI: 10.16245/j.cnki.issn1008-6978.2020.11.015.
ZHANG Jingjing, XU Haifei, LI Tonghuai, et al. Application value of acoustic radiation force impulse imaging combined with stress elastography in the diagnosis of hysteromyoma degeneration[J]. Journal of Clinical Ultrasound in Medicine, 2020,22(11):843-846. DOI: 10.16245/j.cnki.issn1008-6978.2020.11.015.
目的探讨声辐射力脉冲成像(ARFI)技术联合应力式弹性成像在子宫肌瘤变性诊断中的应用价值。方法选取我院经手术病理结果证实的子宫肌瘤患者98例
共115个病灶
根据子宫肌瘤是否变性分为变性组22个
未变性组93个。应用常规超声观察病灶大小、位置、血流情况及内部回声;应力式弹性成像获取病灶弹性成像评分;ARFI技术测量病灶剪切波速度(VTQ);比较两组上述各参数差异。绘制受试者工作特征(ROC)曲线分析ARFI技术、应力式弹性成像及两者联合诊断效能
并对其进行比较。结果变性组与未变性组在子宫肌瘤位置、内部回声及血流情况比较差异均无统计学意义;两组直径≥5 cm者与<5 cm者比较差异有统计学意义(P<0.05)。变性组弹性成像评分和VTQ值均低于未变性组
差异均有统计学意义(均P<0.05)。ARFI技术判断子宫肌瘤变性的ROC曲线下面积为0.71
VTQ截断值为2.52 m/s时判断子宫肌瘤变性的敏感性、特异性及准确率分别为65.52%、79.10%及88.69%。应力式弹性成像判断肌瘤变性的敏感性为63.63%、特异性为93.55%、准确率为87.83%。ARFI技术联合应力式弹性成像评估子宫肌瘤变性的敏感性、特异性及准确率分别为86.36%、95.59%、90.43%
均较单独技术诊断效能高
差异均有统计学意义(均P<0.05)。结论 ARFI技术联合应力式弹性成像可提高子宫肌瘤变性与否的鉴别诊断价值。
ObjectiveTo investigate the application value of acoustic radiation force impulse imaging(ARFI)combined with stress elastography in the diagnosis of hysteromyoma degeneration.MethodsA total of 115 lesions were selected from 98 patients with hysteromyoma degeneration confirmed by surgical pathological results in our hospital
including 22 lesions in degeneration group and 93 lesions in undenatured group. Conventional ultrasound was used to observe the size
location
blood flow and internal echo of the lesion. Stress elastography was used to observe the lesion elasticity score. ARFI was used to measured shear wave velocity(VTQ). The differences of above parameters were compared between the two groups. The receiver operating characteristic(ROC)curve was drawn to analyze and compare the efficacy of ARFI
stress elastography and combined diagnosis.ResultsThere were no significant difference of location
blood flow and internal echo of hysteromyoma between degeneration group and undenatured group. The patients in two groups with a diameter ≥5 cm or <5 cm were statistically different(P<0.05). The elastic score and VTQ in degeneration group were lower than those in undenatured group
and the differences were statistically significant(both P<0.05). The area under the ROC curve of ARFI for judging hystero myoma degeneration was 0.71
the sensitivity
specificity and accuracy were 65.52%
79.10% and 88.69%
respectively
when VTQcutoff value was 2.52 m/s. The sensitivity
specificity and accuracy of stress elastography for judging hysteromyoma degeneration were 63.63%
93.55% and 87.83%
respectively. The sensitivity
specificity and accuracy of ARFI combined with stress elastography for assessing of hysteromyoma degeneration were 86.36%
95.59% and 90.43%
respectively. Compared with the single technique
the difference of diagnostic efficacy was statistically significant(all P<0.05).ConclusionARFI combined with stress elastography can significantly improve the differential diagnostic value in the diagnosis of hysteromyoma degeneration or not.
声辐射力脉冲成像应力式弹性成像子宫肌瘤变性
Acoustic radiation force impluse imagingStress elastographyHysteromyomaDegeneration