MIAO Ji-dong, et al. Effect of focused ultrasound exposure patterns on bovine liver damages n vitro[J]. Journal of Chongqing Medical University, 2011,36(4):394-397.
MIAO Ji-dong, et al. Effect of focused ultrasound exposure patterns on bovine liver damages n vitro[J]. Journal of Chongqing Medical University, 2011,36(4):394-397. DOI: 10.13406/j.cnki.cyxb.2011.04.034.
Objective:To compare the destructive characteristics of the bovine liver in vitro by double or single high intensity focused ultrasound exposure patterns.Methods:Two HIFU instruments(frequency of 1.6 MHz
acoustic power of 300 W
in situ intensity of 12 000 W/cm2)were used to ablate in vitro bovine liver ranging from 8 cm and 10 cm.In double exposure mode
two focused ultrasound transducers were fixed on the handle;the directions of focused ultrasound beams were opposite.In the single exposure mode
one focused ultrasound transducer was fixed on one handle
and a metal reflector was fixed on the other handle.The characteristics of coagulation necrosis and temperature change in treated bovine liver were observed.Results:In the double exposure mode
and the coagulation necrosis shape was similar to circular cylinder.However
under the same exposure energy and exposure time
the target area coagulation necrosis volume and the temperature were higher than those in the single side exposure mode.Conclusion:Working in double side exposure mode
HIFU can induce complete coagulation necrosis of predetermined bovine liver targets
and the effect surpasses the single side exposure mode.