Dong Wei, Jia Junling, Li Zhen, et al. Comparison of different treatment methods for Vancouver type B2 fracture based on finite element analysis[J]. Journal of Chongqing Medical University, 2016,41(5):472-476.
Dong Wei, Jia Junling, Li Zhen, et al. Comparison of different treatment methods for Vancouver type B2 fracture based on finite element analysis[J]. Journal of Chongqing Medical University, 2016,41(5):472-476. DOI: 10.13406/j.cnki.cyxb.000820.
Objective:To compare different treatment methods for Vancouver type B2 fracture based on finite element analysis.Methods:The data of left femurs of 30 healthy aged volunteers
M/L Taper prosthesis and Wagner S/L prosthesis were collected.The data were inputted into finite element analysis software and Vancouver type B2 fracture was simulated. The offsets of fractures under different loads were recorded and analysed by statistic software. Results:Under both load conditions
significant statistical difference could be found among the four groups(P <0.05). With pair-wised comparisons
significant statistical differences could be found between the first group and the rest 3 groups(P<0.05)
but no statistical difference could be found among the last 3 groups.Conclusion:Wagner SL prosthesis could be used for the treatment of Vancouver type B2 fractures. The optimum length of the distal fixed part should be twice longer than the diameter of the femur. If the prosthsis with ideal length could not be used
a shorter prosthesis could be considered
but a combination of locking compressed plate is recommended.
关键词
股骨假体周围骨折Vancouver分型内固定有限元分析
Keywords
femurperiprosthetic fracturesVancouver classificationinternal fixationfinite element analysis