FANG Wen-hao, et al. Optimization of scanning parameters for 64-slice volume CT digital subtraction angiography[J]. Journal of Chongqing Medical University, 2009,34(11):1542-1545.
FANG Wen-hao, et al. Optimization of scanning parameters for 64-slice volume CT digital subtraction angiography[J]. Journal of Chongqing Medical University, 2009,34(11):1542-1545.DOI:
3D VR)评价减影图像质量。结果:变化mA、同步与非同步变化kV、变化层厚和同步变化螺距减影后图像质量均为Ⅰ级(100%)
非同步变化螺距减影后图像质量为Ⅲ级(100%)。随mA、kV、层厚降低和螺距增大
减影后全图CT值增大
图像噪声增加而图像质量降低。非同步变化kV的双能量减影
随kV的增加
减影后全图CT值降低
图像噪声降低而图像质量提高。结论:应用双能量、厚层采集薄层重建、降低毫安和增加螺距
不仅能保证VCTDSA的图像质量
同时可降低X线辐射剂量。
Abstract
Objective:To investigate the influence of scanning parameters(including mA
kV
slice thickness and pitch)on image quality and radiation dose of volume computed tomographic digital subtraction angiography(VCTDSA).Methods:VCTDSA were performed on a skull specimen.Scans were obtained with different scanning parameters
including current of 100
85
75
65
50
25 and 10 mA
synchronic voltage modification of 140
120
100 and 80 k V
non-synchronic voltage modification of 120~140
100~140
80~140
100~120
80~120 and 80~100 kV
slice thickness of 0.625
1.25
2.5
3.75 and 5 mm
and pitch of 0.531
0.969.CT numbers of the overall view were measured on subtraction images obtained from Add/Sub software.The image quality of subtraction images were analyzed
which were reformatted and displayed with 3D volume rendering(3D VR).Results:Except subtracted image qualitys were evaluated gradeⅢ(100%)with non-synchronic pitch
other subtracted image qualitys were evaluated grade Ⅰ(100%)undergoing different scanning parameters(mA
kV
slice thickness and synchronic pitch).With the decrease of mA
kV
slice thickness and the increase of pitch
CT numbers and imaging noise increased
and imaging quality after subtraction decreased.Image quality was improved and noise was reduced by dual energy subtraction.With the increase of k V
CT numbers increased.Conclusion:Image quality was improved and radiation dose was reduced by dual energy subtraction technique