Comparison on quality between FSE T2WI and PROPELLER FSE T2WI for upper abdomen examination with 3.0T MRI
返回论文页
|更新时间:2023-12-18
|
Comparison on quality between FSE T2WI and PROPELLER FSE T2WI for upper abdomen examination with 3.0T MRI
Journal of Chongqing Medical UniversityVol. 38, Issue 2, Pages: 257-261(2014)
作者机构:
1. 华中科技大学同济医学院附属同济医院放射科,武汉,430030
2. 滨州医学院附属医院放射科,滨州,256603
作者简介:
基金信息:
DOI:
CLC:
Published:2014,
扫 描 看 全 文
Wang Qiuxia, Chen Liang, Luo Xin, et al. Comparison on quality between FSE T2WI and PROPELLER FSE T2WI for upper abdomen examination with 3.0T MRI. [J]. Journal of Chongqing Medical University 38(2):257-261(2014)
DOI:
Wang Qiuxia, Chen Liang, Luo Xin, et al. Comparison on quality between FSE T2WI and PROPELLER FSE T2WI for upper abdomen examination with 3.0T MRI. [J]. Journal of Chongqing Medical University 38(2):257-261(2014)DOI:
Comparison on quality between FSE T2WI and PROPELLER FSE T2WI for upper abdomen examination with 3.0T MRI
Objective:To evaluate the effectiveness of periodically rotated overlapping parallel lines with enhanced reconstruction(PROPELLER) technique in reducing motion artifact and improving image quality on FSE T2WI image. Methods:Ninety-seven patients with a mean age of 53.8 years(age range,19-76 years) were underwent upper abdominal examination with 3.0T MRI. The same scanning parameters were used on axial T2-weighted PRO-PELLER sequence and FSE sequence. The respiratory artifacts,gastrointestinal movement artifacts,pulsatility artifacts,overall image quality,and the edge of the liver,pancreas were evalu-ated by two radiologists using a 3-point scale. The Wilcoxon’s rank sum test was used for statistical analysis. The Cohen’s Kappa test was used to determine the agreement between two radiologists. Results:On T2-weighted PROPELLER images,the respiratory artifacts,gastrointestinal movement artifacts,vascular pulsatility artifact were reduced and the overall image quality,sharpness of the liver,pancreas and renal corticomedullary junction area were improved(Z=-5.931,-6.245,-6.481,-6.164,-5.745;P<0.05). The agreement between two radiologists is consistent. Conclusions:Compared those of with FSE T2WI sequence,the respiratory artifacts,gastrointestinal movement artifacts,vascular pulsatility artifacts are significantly reduced on PROPELLER T2WI sequence. PROPELLER T2WI can better display the anatomic structure,improve the image quality and shorten the scanning time.
关键词
磁共振成像螺旋桨技术图像伪影
Keywords
magnetic resonance imagingperiodically rotated overlapping parallel lines with enhanced reconstruction(PROPELLER)image artifacts