余璞, 左浩, 段茂, 等. 三维经食管超声心动图评估人工机械瓣血管翳梗阻的价值研究[J]. 临床超声医学杂志, 2024,26(2).
the value of three-dimensional transesophageal echocardiography in evaluating obstruction of artificial mechanical valve pannus[J]. Journal of Clinical Ultrasound in Medicine, 2024,26(2).
目的 研究三维经食管超声心动图(three-dimensional transesophageal echocardiography
3D-TEE)在评估人工机械瓣膜血管翳的狭窄程度中的应用价值。方法 选取2019年10月-2023年3月在我院经食管超声心动图诊断为人工瓣膜血管翳
并且经手术证实的病例46例
其中人工二尖瓣血管翳22例
人工主动脉瓣血管翳24例。观察人工瓣血管翳在3D-TEE上的成像特征
在三维图像上测量血管翳面积、瓣环内侧缘面积与瓣环外侧缘面积之比、血管翳宽度与瓣膜外径之比(pannus width/valve external diameter
W/D)等相关参数
经胸超声心动图( transthoracic echocardiography
TTE)上测量平均跨瓣压差(the mean transaortic pressure gradient
mTPG)
分析并比较血管翳的三维测量参数与mTPG的相关性。结果 ①3D-TEE结果显示当血管翳环绕人工瓣环的角度在大于等于180°时更容易引起瓣口的明显狭窄
角度大小对瓣口狭窄程度的影响具有显著性差异(p<0.05)。②对于人工主动脉瓣膜
三维图像上测得的瓣环内侧缘面积与瓣环外侧缘面积之比(0.33±0.08)、血管翳W/D(0.41±0.09)、血管翳面积(1.79 ±0.68)cm2 均与mTPG(44.13±13.59)mmHg具有良好的相关性
相关性分别为R=-0.782
P<0.001
R=0.820
P<0.001
R=0.741
P<0.001。对于人工二尖瓣膜
人工二尖瓣瓣环内侧缘面积与瓣环外侧缘面积之比(0.31±0.07)、血管翳W/D(0.35±0.09)、血管翳面积(1.88 ±0.71)cm2 均与人工二尖瓣mTPG(13.87±6.59)mmHg具有良好的相关性
相关性分别为R=-0.793
P<0.001
R=0.878
P<0.001
R=0.806
P<0.001。③人工二尖瓣血管翳相关参数与mTPG的相关性优于人工主动脉瓣。结论 3D-TEE在评估人工瓣膜血管翳的梗阻程度方面具有重要价值
尤以W/D指标能更准确地评估瓣膜的梗阻程度。3D-TEE在人工二尖瓣血管翳梗阻程度的评估价值优于人工主动脉瓣。
Objective: To investigate the efficacy of three-dimensional transesophageal echocardiography (3D-TEE) in evaluating the extent of stenosis in pannus on artificial mechanical valves. Methods: A total of 46 cases with pannus on the prosthetic heart valve diagnosed via transesophageal echocardiography (TEE)
confirmed surgically in our hospital between October 2019 and March 2023
were recruited. This cohort included 12 cases of pannus on the prosthetic mitral valve and 22 cases of pannus on the prosthetic aortic valve. The 3D-TEE imaging characteristics of pannus were observed
and various parameters were measured
including pannus area
the ratio of the areas of the medial and lateral edges of the annulus
and the ratio of the pannus width to the valve external diameter (W/D). Additionally
the mean transvalvular pressure gradient was measured via transthoracic echocardiography (TTE). The study analyzed and compared the correlation between these parameters of the pannus and the mean transvalvular pressure gradient (mTPG). Results: ① The 3D-TEE results revealed that a pannus angle encompassing the prosthetic valve annulus
equal to or exceeding 180°
was associated with a higher likelihood of pronounced stenosis in the valve orifice. The angle size was significantly associated with the degree of stenosis (P < 0.05). ② In prosthetic aortic valves
it was found that the three key measurements - the ratio of the areas of the medial and lateral edges of the annulus being 0.33±0.08
W/D of the pannus at 0.41±0.09
and the area of the pannus being 1.79±0.68 cm2 - strongly correlated with the mTPG of 44.13±13.59 mmHg. The correlation coefficients for these combined parameters were R=-0.782
P<0.001
R=0.820
P<0.001
and R=0.741
P<0.001
respectively. In artificial mitral valves
it was observed that the concurrent presence of the following three measurements - the ratio of the areas of the medial and lateral edges of the annulus at 0.31±0.07
the W/D of the pannus at 0.35±0.09
and the area of the pannus at 1.88±0.71 cm2 - significantly correlated with the mTPG of 13.87±6.59 mmHg. The combined correlation coefficients for these parameters were R=-0.793
P<0.001
R=0.878
P<0.001
and R=0.806
P<0.001
respectively. ③The correlation between parameters related to the pannus in prosthetic mitral valves and the mTPG gradient was stronger compared to that in prosthetic aortic valves. Conclusion: 3D-TEE is valuable for assessing the extent of stenosis in pannus on prosthetic valves. Particularly
the pannus W/D index can accurately assess the degree of valve obstruction. The study suggests that 3D-TEE is more valuable for evaluating the degree of pannus obstruction in prosthetic mitral valves than that of prosthetic aortic valves.
三维经食管超声心动图人工瓣血管翳瓣膜梗阻
three-dimensional transesophageal echocardiographyartificial valve pannusvalvular obstruction