[1]邓相华,李少健,严为棣,吴祖尧,梅芳瑞,柯家珉,姚远志,陈砚候.硅橡胶人工肌腱的临床应用及有关问题探讨[J].重庆医学院学报,1985(02):144-147.
ARTTFIAL SILICON TENDON SUMMY OF CHINICAL APPLICATION AND DISSCUSSION OF RELATED PROBLEMS[J]. Journal of Chongqing Medical University, 1985,(2):144-147.
手部无人区屈指肌腱损伤
用传统的自体肌腱移植治疗
疗效很差。1974年起
作者用国产硅橡胶人工肌腱对31例(37指)屈指肌腱损伤作了肌腱成形术。其中25例(31指)随访一年以上。结果:优良8指、尚可7指、劣16指
有效率48.6%
逊于Hunter报告的结果。硅橡胶人工肌腱质量不稳定
局部组织反应较重是影响疗效的主要原因。其次
人工肌腱置入腕上方者
疗效亦逊于置于掌心部病例。为进一步提高人工肌腱疗效
我们须进一步研制我国的医用硅橡胶制品
并提出我国自己的医用级标准。
Flexor tendon division at any point from the distal palmar crease to the distal phalanx often results in poor function with traditional tendontransplantation. In 1974
we began to apply 3 types (7501
7502
7503) of silicon tendons
which were made by Chongqing Changji-ang Rubber Fatory
to repair severed tendons. To date
31 patients (37 fingers) have been implanted with the siliconrod
and 25 of these cases (31 fingers) have been followed up for more than 1 year.Overall
g fingers showed excellent results
7fingers were good and16 were poor. Generally
our series results are not as good as Hunter’s. This is because the Fatory could not control the quality of the silicon rod
therefore the tissue reaction is much more severe after implanting. We found the quality of type-7501 is much better tha
n type-7503. Ifl addition
we compared 2 different implanting niethods that the conidion of the cases which were implanted proximaly in the forearm are much poorer than those in the palmoIn order to promote the success of tendon repair. We must improve the quality of our medical silicon prothesis and propose a medical standard for our country.