[1]王思凯,张雪,钟旭萍,杨彬,杨琴,何其舟.二代双源CT低管电压、大螺距联合迭代重建技术应用于急诊创伤性骨折的可行性研究[J].西南医科大学学报,2021,44(03):218-221.
WANG Si-kai, ZHANG Xue, ZHONG Xu-ping, et al. The feasibility study of the second dual-source CT low-tube-voltage high-pitch with iterative reconstruction in applying to the emergency traumatic fracture[J]. Journal of Southwest Medical University, 2021,44(3):218-221.
[1]王思凯,张雪,钟旭萍,杨彬,杨琴,何其舟.二代双源CT低管电压、大螺距联合迭代重建技术应用于急诊创伤性骨折的可行性研究[J].西南医科大学学报,2021,44(03):218-221. DOI:
WANG Si-kai, ZHANG Xue, ZHONG Xu-ping, et al. The feasibility study of the second dual-source CT low-tube-voltage high-pitch with iterative reconstruction in applying to the emergency traumatic fracture[J]. Journal of Southwest Medical University, 2021,44(3):218-221. DOI:
目的研究二代双源CT低管电压、大螺距联合迭代重建(SAFIRE)技术对急诊创伤性骨折检查的可行性。方法前瞻性纳入因急诊创伤来本院行CT检查的23例患者
每位患者均按以下A、B两种扫描程序扫描。标准扫描程序(首次检查
A)/低剂量扫描程序(复查时
B)、管电压120 kv/100 kv(脊柱、肋骨、骨盆、髋关节)、100 kv/80 kv(四肢)
螺距1.0/3.0
滤波反射投影重建(FBP)/迭代重建(SAFIRE)。比较A、B扫描程序的放射剂量(CT容积剂量指数(CTDIvol)、辐射剂量长度乘积(DLP))、扫描时间、图像质量和诊断效能的差异性。结果 B扫描程序的CTDIvol(2.86±1.61 mGy)、DLP(110.09±87.19 mGy.cm)分别较A扫描程序CTDIvol(8.87±4.47 m Gy)、DLP(301.65±236.04 mGy.cm)降低约67.8%、63.5%
存在统计学差异(t=-4.198、6.022
P <0.001)。B扫描程序的扫描时间(0.90±0.43 s)较A扫描程序(5.90±3.00 s)缩短约84.75%
存在显著性差异(t=8.880
P <0.001)。A、B扫描程序图像质量评分均≥4分
达到诊断要求
无统计学差异(t=-1.582
P> 0.05)
均检出所有骨折。2名医生对图像质量评分的一致性好
K=0.664。结论二代双源CT低管电压、大螺距联合SAFIRE技术应用于急诊创伤性骨折检查
明显降低辐射剂量和缩短扫描时间的同时获取满足临床诊断的图像质量
外伤患者复查了解骨折情况时
可选用低剂量CT扫描技术以降低辐射剂量。
Objective To investigate the feasibility of second-generation dual-source computed tomography(DSCT)with low-tube-voltage
high-pitch
and sinogram-affirmed iterative reconstruction(SAFIRE)in the examination of emergency traumatic fractures. Methods A prospective study was conducted in 23 patients who were admitted to our hospital and required CT examination due to emergency trauma. Each patient was examined with CT using both the following A and B protocols: Standard scanning protocol(first examination
A)/low-dose scanning protocol(re-examination
B)
tube voltage 120 kv/100 kv(for the spine
ribs
pelvis
and hip joint)
100 kv/80 kv(for the limbs)
pitch 1.0/3.0
and filtered back projection/SAFIRE. The A and B protocols were compared in terms of radiation dose(volume CT dose index [CTDIvol]
dose-length product [DLP])
scanning time
image quality
and diagnostic efficacy. Results Compared with the A protocol
the B protocol had significantly reduced CTDIvol(2.86 ± 1.61 mGy vs 8.87 ± 4.47 mGy
t =-4.198
P < 0.001)
DLP(110.09 ± 87.19 m Gy·cm vs 301.65 ±236.04 mGy · cm
t = 6.022
P < 0.001)
and scanning time(0.90 ± 0.43 s vs 5.90 ± 3.00 s
t = 8.880
P <0.001)
and the reduction was 67.8%
63.5%
and 84.75%
respectively. Both A and B protocols had an image quality score of no less than 4 points
both meeting the diagnostic requirement
with no significant difference(t =-1.582
P > 0.05). Both A and B protocols detected all fractures. The two doctors had a high consistency in the image quality scoring(K = 0.664). Conclusion The application of second-generation DSCT with low-tube-voltage
high-pitch
and SAFIRE in the examination of emergency traumatic fractures can significantly reduce the radiation dose and scanning time
with good image quality which satisfies the clinical diagnosis. When the traumatic patients are re-examined to obtain the status of fracture
a low-dose CT scanning protocol can be used to reduce the radiation dose.
低管电压大螺距迭代重建低剂量
Low-tube-voltageHigh-pitchIterative reconstructionLow dose