Application value of a new lung ultrasound scoring method in evaluating the evacuation timing of mechanical ventilation for neonatal respiratory distress syndrome
返回论文页
|更新时间:2021-10-13
|
Application value of a new lung ultrasound scoring method in evaluating the evacuation timing of mechanical ventilation for neonatal respiratory distress syndrome
Journal of Clinical Ultrasound in MedicineVol. 23, Issue 4, Pages: 280-283(2021)
JIANG Qiuxia, HUANG Rongsen, SHEN Longyuan, et al. Application value of a new lung ultrasound scoring method in evaluating the evacuation timing of mechanical ventilation for neonatal respiratory distress syndrome. [J]. Journal of Clinical Ultrasound in Medicine 23(4):280-283(2021)
JIANG Qiuxia, HUANG Rongsen, SHEN Longyuan, et al. Application value of a new lung ultrasound scoring method in evaluating the evacuation timing of mechanical ventilation for neonatal respiratory distress syndrome. [J]. Journal of Clinical Ultrasound in Medicine 23(4):280-283(2021) DOI: 10.16245/j.cnki.issn1008-6978.2021.04.011.
Application value of a new lung ultrasound scoring method in evaluating the evacuation timing of mechanical ventilation for neonatal respiratory distress syndrome
Objective method in evaluating the evacuation timing of mechanical ventilation for neonatal respiratory distress syndrome(NRDS).Methods selected. Bedside lung ultrasound score(LUS)was performed before the children extubation when they reached the clinical evacuation standard
LUS was measured by six-zone method
ten-zone method
twelve-zone method and fourteen-zone method in two groups.According to the results of evacuation
the children were divided into the success group(n=49)and the failure group(n=13)
the LUS of the two groups were compared. LUS was administered to the evacuation failure group before reintubation and before re-extubation.The ROC curve was used to analyze the diagnostic efficacy of LUS for the evacuation timing.ResultsThere were significantly difference in LUS between the two groups for different four method(all P<0.05).In the failure group
the following pathology changers were found by lung ultrasonography:4 cases of obvious lung edema
5 cases of extensive lung consolidation
2 cases of extensive lung consolidation at the bottom of the lung
1 case of large atelectasis of the upper right lung
and 1 case of no abnormality in lung ultrasonography but cardiac dysfunction was found.Compared with the other three methods
the fourteen-zone lung ultrasound scoring method had a better diagnostic performance for the evacuation timing
taking40.5 point as the cutoff value to predict the risk of evacuation failure
the area under the curve was 0.965
the sensitivity was92.31%
and the specificity was 93.88%.Conclusion value in evaluating the evacuation timing of mechanical ventilation in NRDS
which can identify high-risk children with early failure of evacuation
Evaluation of the left ventricular myocardial function in preterm infants with neonatal respiratory distresssyndrome by two-dimensional speckle tracking imaging
Clinical value of pulmonary ultrasound score in evaluation of neonatal respiratory distress syndrome
Related Author
施丽景
李晓庆
沈龙源
黄嵘森
吕国荣
江秋霞
陈智毅
廖剑艺
Related Institution
QuanZhou Women''s and Children’s Hospital
Quanzhou medical college maternal and child health service application technology collaborative innovation center
泉州市妇幼保健院·儿童医院
The Third Affiliated Hospital Of Guangzhou Medical University
Department of Ultrasound Medicine,the Third Affiliated Hospital of Guangzhou Medical University