XU Xiaolan, GUZAILINUER·Yasen, XIA Nan. Study on the reliability and features of sound touch elastography measuring the elasticity of brachial plexus roots in normal adults. [J]. Journal of Clinical Ultrasound in Medicine 22(4):250-253(2020)
XU Xiaolan, GUZAILINUER·Yasen, XIA Nan. Study on the reliability and features of sound touch elastography measuring the elasticity of brachial plexus roots in normal adults. [J]. Journal of Clinical Ultrasound in Medicine 22(4):250-253(2020) DOI: 10.16245/j.cnki.issn1008-6978.2020.04.004.
Study on the reliability and features of sound touch elastography measuring the elasticity of brachial plexus roots in normal adults
Objective To explore the reliability of sound touch elastography measuring the elasticity of brachial plexus roots in normal adults
and to analyze their characteristics.Methods The sound touch elastography technique was used to proceed repeated measurements for the elasticity of the bilateral C5
C6 and C7 nerve roots of the 44 enrolled healthy volunteers twice in this study by one examiner.The two measurements were summarized and used for evaluation of intra-rater reliability and minimal detectable change.One of the two test results was selected randomly from all subjects
and the data were summarized and used for the analysis of relevant characteristics and influencing factors.Results The nerve root elasticity values of C5
C6 and C7 on the left and right sides of the normal adults were(2.0.64±6.87)kPa
(2.1.34±5.35)kPa
(2.5.48±7.13)kPa and(20.19±7.86) kPa
(22.35±7.99) kPa
(24.61±8.15) kPa
respectively
and the differences of both sides were not significant.Further analysis of the combination of both sides showed that the elastic values of C5
C6 and C7 were(20.41±7.34) kPa
(2.1.85±6.78)kPa
(2.5.03±7.58)kPa
respectively
which showed a significant increasing trend(P<0.01).There were significant differences in pairs(all P<0.05).The internal reliability of the evaluator for the C5
C6
C7 nerve root elasticity measurement in the left and right side of 44 normal adults were 0.70
0.63
0.74 and 0.68
0.60
0.71
respectively
and the corresponding minimal detectable changes were 21.01 kPa
18.16 kPa
20.02 kPa and 24.79 kPa
28.09 kPa
24.36 kPa
respectively.Conclusion Sound touch elastography has a certain reproducibility
good reliability and higher minimal detectable change in measuring the elasticity
of C5~C7 brachial plexus roots in normal adults.Possible errors in measurement should be fully considered when using this technique for quantitative assessment of disease changes or treatment efficacy.