DONG Rui, ZHU Qiong, FENG Shuang, et al. Enhanced urokinase thrombolysis by endovascular ultrasound catheter and microbubbles:a experimental study. [J]. Journal of Clinical Ultrasound in Medicine 22(2):85-89(2020)
DONG Rui, ZHU Qiong, FENG Shuang, et al. Enhanced urokinase thrombolysis by endovascular ultrasound catheter and microbubbles:a experimental study. [J]. Journal of Clinical Ultrasound in Medicine 22(2):85-89(2020) DOI: 10.16245/j.cnki.issn1008-6978.2020.02.002.
Enhanced urokinase thrombolysis by endovascular ultrasound catheter and microbubbles:a experimental study
catheter was measured by hydrophone method.Ninety bovine thrombus samples in vitro were randomly divided into six groups:ultrasound(US)+MB+UK(combined group)
US+UK group
US+MB group
US group
UK group and control group
each grouphas 15 samples. The thrombus quality of each group was weighed before and after thrombolysis
and thrombolysis rate wascalculated and compared.After treatment
the thrombus in each group wasstained with HE
and the distribution of erythrocyte andfibrin was observed under a light microscope.ResultsThe PNP of EKOS
?
catheter varied irregularly within the range of 0.4~6.4 MPa.The thrombolysis rate of combined group(41.9±4.5)% was close to that of US+UK group(42.0±3.3)%
which weresignificantly higher than that of the other groups(all P
<
0.05).The thrombolysis rate of US+MB group was(28.4±3.3)%
comparedwith(27.8±3.1)% in UK group
there was no significant difference between the two groups
which were significantly higher thanthat of US group(23.9 ±3.0)% and control group(17.5±3.7)%
there were significant difference(both P
<
0.05).Histologicalexamination of the thrombus in combined group and US+UK group showed that the thrombus contacting with EKOS
?
catheter wasmore disintegrated
and the distribution of red blood cells was more sparse than other groups.Conclusion EKOS
?
catheter canobviously improve the thrombolysis rate of UK to bovine thrombus in vitro
however
MB has no significant effect on thrombolysis rate.