Qin Donjie, Liang Jing, Wei Yanfei, et al. Determination of Blood Concentration of Fluconazole in Neonates by High Performance Liquid Chromatography. [J]. 26(12):40-43(2020)
DOI:
Qin Donjie, Liang Jing, Wei Yanfei, et al. Determination of Blood Concentration of Fluconazole in Neonates by High Performance Liquid Chromatography. [J]. 26(12):40-43(2020) DOI: 10.13407/j.cnki.jpp.1672.108X.2020.12.012.
Determination of Blood Concentration of Fluconazole in Neonates by High Performance Liquid Chromatography
Objective: To establish the high performance liquid chromatography (HPLC) method for determination of blood concentration of fluconazole in neonates. Methods: HPLC was carried out on Diamonsil C18 column (5 μm
4.6 mm×250 mm). The mobile phase consisted of methanol and phosphate buffer solution (pH 7.0)=45︰55 at the flow of 1.0 mL/min
and the detection wavelength was 261 nm. Column temperature was 30 ℃ and sample amount was 10 μL. The blood sample with internal standard was extracted with ethyl acetate
the organic extract was concentrated to dryness under nitrogen
and the residue was dissolved in methanol and analyzed. Results: The standard curve was good linear from 1.0 to 50.0 μg/mL
and the precision
stability and recovery rate were good. Conclusion: This method is simple
repeatable
accurate and reliable
and can be used for determination of blood concentration of fluconazole in neonates by HPLC.