Possibility of automatic detecting instrument application on cell-count in Hvdrothorax or ascites
Objective;To discuss if automatic detecting instrument could be applied on cell count in hydrothorax or ascites instead of traditional method. Methods; 121 cases of hydrothorax or ascites specimens were collected. SysmexUF - 100 automatic urine sediment appearance
SysmexKX -21 blood cell analysis instrument and traditional manual method were used to count white blood cell and totoal cell in all samples. The data were treat with student-T test and relativity analysis. Results;There was significant difference between data collected by SysmexUF - 100 aytinatuc urine sediment appearance and traditional manual method after T test ( P <0. 05). For white blood cell
the correlation coefficient was 0. 645
as for total cell
it was 0. 560. The resualt was although different from manual method when applying SysmentKX - 21 blood cell analysis instrument ( P < 0. 05 )
For white blood cell
the correlation coefficient was 0. 482
as for total cell
it was 0. 305. Conclusion; Both Sysmex UF - 100 automatic urine sediment appearance and SysmexKX - 21 blood cell analysis instrument could not be used for cell count in Hydrothorax or ascites to substitute traditional manual method.