Ping Huang, Meng Zhang, Hongkun Wu, et al. Assessment of fluid responsiveness by variability index of internal jugular vena cava and variability index of inferior vena cava in adult petients with cardiosurgery. [J]. Journal of Shanxi Agricultural Sciences 47(5):614-619(2022) DOI: 10.13406/j.cnki.cyxb.003025.
目的:
2
探索超声测量颈内静脉变异度(variability index of internal jugular vena cava,VJVC)、下腔静脉变异度(variability index of inferior vena cava,VIVC)评估心脏术后容量状态。
方法:
2
前瞻性连续纳入中国科学院大学重庆医院心血管外科成人心脏术后76例患者。超声测量被动抬腿实验(passive leg raising,PLR)前后患者心排指数,心排指数增加10%为标准,增加≥10%为容量有反应者(fluid responders,FR);反之则为容量无反应者(fluid unresponders,FN)。同时记录患者PLR前后心率、收缩压、舒张压、平均压。在PLR前,查氨基末端脑利钠肽(N-terminal pro-brain natriuretic peptide,NT-proBNP),超声测量VIVC和左侧VJVC。
结果:
2
对比PLR前后,FR组中心静脉压在PLR前后无统计学差异[(7.6±1.9)mmHg vs.(10.1±11.4)mmHg,
P
=0.154],心率[(83.0±8.9)次/min vs.(75.3±7.5)次/min,
P
=0.000]、收缩压[(91.1±9.4)mmHg vs.(94.9±9.1)mmHg,
P
=0.000]、舒张压[(52.8±9.8)mmHg vs.(58.9±8.2)mmHg,
P
=0.000]、平均压[(65.6±9.2)mmHg vs.(70.7±8.0)mmHg,
P
=0.000]均有明显变化。FN组PLR后收缩压的变化无统计学差异[(90.0±7.4)mmHg vs.(92.3±6.7)mmHg,
P
=0.076],心率[(83.5±9.1)次/min vs.(80.7±7.4)次/min,
P
=0.005]、舒张压[(50.0±7.1)mmHg vs.(55.2±5.7)mmHg,
P
=0.000]、平均压[(63.3±6.8)mmHg vs.(67.6±5.0)mmHg,
P
=0.000]、中心静脉压[(7.9±1.6)mmHg vs.(10.2±1.4)mmHg,
P
=0.000]均有明显变化。对比FR组和FN组患者被动抬腿实验前(T
0
),心率、收缩压、舒张压、平均压、中心静脉压无统计学差异;被动抬腿实验后(T
1
),心率明显下降,FR组与FN组比较有统计学差异[(75.3±7.5)次/min vs.(80.7±7.4)次/min,
P
=0.001],舒张压和平均压变化有统计学差异[(58.9±8.2)mmHg vs.(55.2±5.7)mmHg,
P
=0.014;(70.7±8.0)mmHg vs.(67.6±5.0)mmHg,
P
=0.024],收缩压和中心静脉压无统计学差异[(94.9±9.1)mmHg vs.(92.3±6.7)mmHg,
P
=0.075;(10.1±11.4)mmHg vs.(10.2±1.4)mmHg,
P
=0.474],2组间血浆NT-proBNP无统计学差异[(1 144.6±9 92.1)pg/mL vs.(1 243.6±1 095.0)pg/mL,
P
=0.335]。FR组和FN组VIVC有统计学差异[(15.2±3.9)% vs.(12.5±4.2)%,
P
=0.003],FR组和FN组VJVC有统计学差异[(17.7±4.3)% vs.(13.9±4.0)%,
P
=0.000]。VIVC诊断容量有反应的临界值为11.5%,敏感性为87.8%,特异性为55%(曲线下面积为0.7,95% CI=0.575~0.826,
P
=0.003)。VJVC诊断容量有反应的临界值为15.5%,敏感性为68.3%,特异性为77.1%(曲线下面积为0.764,95% CI=0.654~0.874,
P
=0.000)。
结论:
2
颈内静脉变异度和下腔静脉变异度都能较好、及时地预测心脏术后患者容量状态,准确指导临床补液治疗。且颈内静脉比下腔静脉更容易测量,位置更加表浅,更加恒定,不受患者肥胖影响,不受腹压影响,无心脏术后引流管影响,重复性更好。临床工作中更推荐使用VJVC评估心脏术后容量状态。
Objective
:
2
To evaluate the variability index of internal jugular vena cava (VJVC) and variability index of inferior vena cava (VIVC) through ultrasound in predicting fluid responsiveness in adult petients with cardiosurgery.
Methods
:
2
Prospectively
76 adult pa tients who had cardiosurgery in the Department of Cardiovascular Surgery
Chongqing General Hospital were collected in the study
and the passive leg raising (PLR) tests were carried out on them of intensive monitoring in the intensive care unit (ICU) ward. Patients with elevated cardiac output index≥10% after PLR were included in the fluid responders (FR) group
while patients with elevated cardiac output index<10% were included in the fluid unresponders (FN) group. Meanwhile
the hemodynamic parameters of heart rate
systolic blood pressure
diastolic blood pressure and mean blood pressure were all recorded before and after PLR tests in both groups. The N-terminal pro-brain natriuretic peptide (NT-proBNP)
VJVC
and VIVC were recorded before PLR tests.
Results
:
2
After PLR tests
in FR group
there were no significant differences in central venous pressure[ (7.6±1.9) mmHg vs. (10.1±11.4) mmHg
P
=0.154]; there were significant differences in heart rate[ (83.0±8.9) bpm vs. (75.3±7.5) bpm
P
=0.000]
systolic blood pressure[ (91.1±9.4) mmHg vs. (94.9±9.1) mmHg
P
=0.000]
diastolic blood pressure[ (52.8±9.8) mmHg vs. (58.9±8.2) mmHg
P
=0.000]
and mean blood pressure[ (65.6±9.2) mmHg vs. (70.7±8.0) mmHg
P
=0.000]. In FN group after PLR
there were no significant differences in systolic blood pressure[ (90.0±7.4) mmHg vs. (92.3±6.7) mmHg
P
=0.076]
and there were significant differences in heart rate[ (83.5±9.1) bpm vs. (80.7±7.4) bpm
P
=0.005]
diastolic blood pressure[ (50.0±7.1) mmHg vs. (55.2±5.7) mmHg
P
=0.000]
mean blood pressure[ (63.3±6.8) mmHg vs. (67.6±5.0) mmHg
P
=0.000]
and central venous pressure[ (7.9±1.6) mmHg vs. (10.2±1.4) mmHg
P
=0.000]. There were no significant differences in heart rate
systolic blood pressure
diastolic blood pressure
mean blood pressure and central venous pressure between FR goup and FN group before PLR tests. After PLR
there were no significant differences in systolic pressure[ (94.9±9.1) mmHg vs. (92.3±6.7) mmHg
P
=0.075]and central venous pressure[ (10.1±11.4) mmHg vs. (10.2±1.4) mmHg
P
=0.474]between the two groups
and there were significant differences in heart rate[ (75.3±7.5) bpm vs. (80.7±7.4) bpm
P
=0.001]
diastolic blood pressure[ (58.9±8.2) mmHg vs. (55.2±5.7) mmHg
P
=0.014]
and mean blood pressure[ (70.7±8.0) mmHg vs. (67.6±5.0) mmHg
P
=0.024]. In addition
There was no significant difference in plasma NT-proBNP between the two groups[ (1 144.6±992.1) pg/mL vs. (1 243.6±1095.0) pg/mL
P
=0.335]
and there were significant differences in VIVC[ (15.2±3.9) % vs. (12.5±4.2) %
P
=0.003]and VJVC[ (17.7±4.3) % vs. (13.9±4.0) %
P
=0.000]. The cutoff value of VIVC diagnostic capacity was 11.5%
the sensitivity was 87.8%
and the specificity was 55% (AUC=0.7
95% CI=0.575-0.826
P
=0.003) . The cutoff value of VJVC diagnostic capacity was 15.5%
the sensitivity was 68.3%
and the specificity was 77.1% (AUC=0.764
95% CI=0.654-0.874
P
=0.000) .
Conclusion
:
2
VIVC and VJVC can easily predict the noninvasive parameter of fuid responsiveness in patients with cardiosurgery and accurately guide the clinical rehydration therapy. Moreover
the internal jugular vein is easier to measure than the inferior vena cava
its position is shallower and more constant
and it is not affected by the obesity of patients
abdominal pressure and drainage tube after cardiosurgery
so the repeatability is better. It is more recommended to use the VJVC to evaluate the fluid responsiveness after cardiosurgery in clinical work.
颈内静脉变异度下腔静脉变异度心脏手术容量状态
variability index of internal jugular vena cavavariability index of inferior vena cavacardiosurgeryfluid responsiveness
Lee J, de Louw E, Niemi M, et al. Association between fluid balance and survival in critically ill patients[J]. J Intern Med, 2015, 277(4): 468-477.
Carsetti A, Cecconi M, Rhodes A. Fluid bolus therapy: monitoring and predicting fluid responsiveness[J]. Curr Opin Crit Care, 2015, 21(5): 388-394.
Cecconi M, Parsons AK, Rhodes A. What is a fluid challenge?[J]. Curr Opin Crit Care, 2011, 17(3): 290-295.
Cecconi M, Dawson D, Grounds RM, et al. Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique[J]. Intensive Care Med, 2009, 35(3): 498-504.
Michard F, Teboul JL. Predicting fluid responsiveness in ICU patients: a critical analysis of the evidence[J]. Chest, 2002, 121(6): 2000-2008.
National Heart L, Wiedemann HP, Wheeler AP, et al. Comparison of two fluid-management strategies in acute lung injury[J]. N Engl J Med, 2006, 354(24): 2564-2575.
Monnet X, Marik PE, Teboul JL. Prediction of fluid responsiveness: an update[J]. Ann Intensive Care, 2016, 6(1): 111.
Michard F, Teboul JL. Using heart-lung interactions to assess fluid responsiveness during mechanical ventilation[J]. Crit Care, 2000, 4(5): 282-289.
Lang RM, Badano LP, Mor-Avi V, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American society of echocardiography and the European association of cardiovascular imaging[J]. J Am Soc Echocardiogr, 2015, 28(1): 1-39.e14.
Monnet X, Marik P, Teboul JL. Passive leg raising for predicting fluid responsiveness: a systematic review and meta-analysis[J]. Intensive Care Med, 2016, 42(12): 1935-1947.
Lima CA, Ritchrmoc MK, Leite WS, et al. Impact of fast-track management on adult cardiac surgery: clinical and hospital outcomes[J]. Rev Bras Ter Intensiva, 2019, 31(3): 361-367.
Bignami E, Guarnieri M, Gemma M. Fluid management in cardiac surgery patients: pitfalls, challenges and solutions[J]. Minerva Anestesiol, 2017, 83(6): 638-651.
Preisman S, Kogan S, Berkenstadt H, et al. Predicting fluid responsiveness in patients undergoing cardiac surgery: functional haemodynamic parameters including the Respiratory Systolic Variation Test and static preload indicators[J]. Br J Anaesth, 2005, 95(6): 746-755.
Cecconi M, Hofer C, Teboul JL, et al. Fluid challenges in intensive care: the FENICE study: a global inception cohort study[J]. Intensive Care Med, 2015, 41(9): 1529-1537.
Boulain T, Achard JM, Teboul JL, et al. Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients[J]. Chest, 2002, 121(4): 1245-1252.
Monnet X, Rienzo M, Osman D, et al. Passive leg raising predicts fluid responsiveness in the critically ill[J]. Crit Care Med, 2006, 34(5): 1402-1407.
Cecconi M, de Backer D, Antonelli M, et al. Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine[J]. Intensive Care Med, 2014, 40(12): 1795-1815.
Wang XT, Liu DW, Yu KJ. Experts'opinions on critical ultrasonography in China[J]. Zhonghua Nei Ke Za Zhi, 2016, 55(11): 900-912.
Dipti A, Soucy Z, Surana A, et al. Role of inferior vena cava diameter in assessment of volume status: a meta-analysis[J]. Am J Emerg Med, 2012, 30(8): 1414-1419.e1.
Vieillard-Baron A, Evrard B, Repessé X, et al. Limited value of end-expiratory inferior vena cava diameter to predict fluid responsiveness impact of intra-abdominal pressure[J]. Intensive Care Med, 2018, 44(2): 197-203.
Brennan JM, Blair JE, Goonewardena S, et al. Reappraisal of the use of inferior vena cava for estimating right atrial pressure[J]. J Am Soc Echocardiogr, 2007, 20(7): 857-861.
Feissel M, Michard F, Faller JP, et al. The respiratory variation in inferior vena cava diameter as a guide to fluid therapy[J]. Intensive Care Med, 2004, 30(9): 1834-1837.
Wang Y, Jiang YH, Wu HN, et al. Assessment of fluid responsiveness by inferior vena cava diameter variation in post-pneumonectomy patients[J]. Echocardiography, 2018, 35(12): 1922-1925.
王志, 刘野, 滕乐, 等. 下腔静脉变异度在容量复苏中的临床应用[J]. 广东医学, 2019, 40(24): 3469-3471, 3475.
Wang Z, Liu Y, Teng L, et al. The clinical application of the variability index of inferior vena cava in volume resuscitation[J]. Guangdong Med J, 2019, 40(24): 3469-3471, 3475.
Cavaliere F, Cina A, Biasucci D, et al. Sonographic assessment of abdominal vein dimensional and hemodynamic changes induced in human volunteers by a model of abdominal hypertension[J]. Crit Care Med, 2011, 39(2): 344-348.
Via G, Tavazzi G, Price S. Ten situations where inferior vena cava ultrasound may fail to accurately predict fluid responsiveness: a physiologically based point of view[J]. Intensive Care Med, 2016, 42(7): 1164-1167.
Vignon P, Repessé X, Bégot E, et al. Comparison of echocardiographic indices used to predict fluid responsiveness in ventilated patients[J]. Am J Respir Crit Care Med, 2017, 195(8): 1022-1032.
Guarracino F, Ferro B, Forfori F, et al. Jugular vein distensibility predicts fluid responsiveness in septic patients[J]. Crit Care, 2014, 18(6): 647.
应海峰, 王明仓, 曹东航, 等. 颈内静脉内径呼吸变异度和血流速度在椎管内麻醉后容量预测中的价值[J]. 临床麻醉学杂志, 2020, 36(5): 445-448.
Ying HF, Wang MC, Cao DH, et al. Ultrasonic measurement of respiratory variability and blood flow velocity of internal jugular vein and the value of intraspinal anesthesia[J]. J Clin Anesthesiol, 2020, 36(5): 445-448.
0
Views
0
下载量
0
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution
京公网安备11010802024621