Yang Qiming, Zhang Shiyang, Zhao Weikang, et al. Biocompatibility of graphene/nano-hydroxyapatite/polyamide 66 composite material. [J]. Journal of Chongqing Medical University 42(5):515-520(2017)
DOI:
Yang Qiming, Zhang Shiyang, Zhao Weikang, et al. Biocompatibility of graphene/nano-hydroxyapatite/polyamide 66 composite material. [J]. Journal of Chongqing Medical University 42(5):515-520(2017) DOI: 10.13406/j.cnki.cyxb.001303.
Biocompatibility of graphene/nano-hydroxyapatite/polyamide 66 composite material
Objective:To evaluate the biocompatibility of graphene/nano-hydroxyapatite/polyamide 66 composite material. Methods:G/n-HA/PA66 composite was prepared by solution blending method. Material extract solution was extracted as specified by GB/T16886.5-2003/ISO 10993-5:2009. The experiments were divided into 3 groups:experimental group
negative control group
and positive control group. The biocompatibilities of G/n-HA/PA66 composite were evaluated according to in vitro cytotoxicity test
acute systemic toxicity test
hemolysis test
pyrogen test and intradermal stimulation test. Results:In vitro cytotoxicity test showed that the cell morphology and proliferation had a good condition and there was a significant statistical difference between each experimental group and positive group(P=0.000). The cell cycle of C3H10T1/2 in G2/M and S period could not be affected by composite material(P =0.889). The cell apoptosis could not be promoted by composite material(P=0.988
P=0.919
P=0.781). Acute systemic toxicity test showed that the mice weight of experimental group[(33.30±3.81)g] had no statistical significance with control group[(31.08±2.91)g](P=0.166)and there was no death and toxicity effect in mice. Hemolysis test showed that the hemolysis rate of G/n-HA/PA66 composite was(0.97±0.36)%
which fitted the evaluation criteria of less than 5%(P=0.000). Pyrogen test showed that the increased temperature of experimental animal was(0.25±0.12)℃
which fitted the evaluation criteria of less than 0.6 ℃(P=0.035). Intradermal stimulation test showed that the phenomena such as erythema and edema were not observed around the injection point in the ex-perimental group and control group
and the PII scores indicated the most slightly stimulation. Conclusion:G/n-HA/PA66 composite had excellent biocompatibility and practicability as a bone fixation implant.
Mid-term clinical efficacy of n-HA/PA66 cage in anterior cervical discectomy and fusion
Study on biocompatability of nano-hydroxyapatite crystals and polyamide 66 composite with human embryonic fibroblast
An tentative radiological and basical studies of preventing the constriction after the laminectomy by using the nano-hydroxyapatite/polyamide-66 composite material artificial vertebral lamina
Study on compatibility of the nHA/PA composite implant
Experiment study of intervertebral fusion cage made by a new biomimetic grafting material
Related Author
Zhao Zenghui
Zhu Yong
Liu Bo
Jiang Dianming
Ou Yunsheng
Deng Qianxing
et al
MENG Chunyang
Related Institution
Department of Orthopedics,College of Clinical Medicine,Chongqing Medical University
重庆医科大学组胚教研室
Department of Radiology,the Fist Affiliated Hospital,Chongqing Medical University
重庆医科大学附属第一医院骨科
Department of Orthopaedics,the First Affiliated Hospital,Chongqing Medical University