Shen Dan, Fu Xiaoming, Wang Chao, et al. Three-dimensional finite element analysis of edentulous mandibular four-implant overdenture via ball cap and Locator connection. [J]. Journal of Chongqing Medical University 42(9):1060-1065(2017)
DOI:
Shen Dan, Fu Xiaoming, Wang Chao, et al. Three-dimensional finite element analysis of edentulous mandibular four-implant overdenture via ball cap and Locator connection. [J]. Journal of Chongqing Medical University 42(9):1060-1065(2017) DOI: 10.13406/j.cnki.cyxb.001088.
Three-dimensional finite element analysis of edentulous mandibular four-implant overdenture via ball cap and Locator connection
Objective:To analyze the influence of two connection ways via ball cap and Locator on stress distribution of mandible
implants and abutments under different loading modes by establishing finite element models of edentulous mandibular four-implant-supported overdenture. Methods:The cone beam computerized tomography(CBCT)data of edentulous mandibular patients were collected
and three-dimensional finite element models of implant-supported overdenture via ball cap and Locator connections were established respectively by Mimics
Gomagic Studio
SolidWorks software
and stress distribution on the mandible
implants and abutments under different loading modes was analyzed combined with ANSYS software. Results:(1)Under different loading modes
stress concentration was found on cortical bone of the neck of implants and lower margin of mandible median lingual region via the two connection ways
but the max equivalent stress(von Mises stress)was located on cortical bone of lower margin of mandible median lingual region.(2)Stress concentration was found on the neck of implants and the contact zone of implants and abutments.(3)Stress of the above areas of stress concentration via Locator connection was lower than that of ball cap connection under the same loading mode. Conclusion:(1)The max von-Mises tress of four-implantsupported overdenture is located on cortical bone of lower margin of mandible median lingual region
rather than cortical bone the neck of implants.(2)The von-Mises tress of overdenture by Locator connection is lower than that of ball cap connection;overdenture via Locator connection is more favorable to the physical health of alveolar bone and long-term persistence of implants.
关键词
下无牙颌4种植体覆盖义齿球帽locator三维有限元
Keywords
edentulous mandiblefour implantsoverdentureball caplocatorthree-dimensional finite element