Experimental study of the fully biodegradable occluder in a swine model of atrial septal defect
degradation and bio-compatibility for closure of atrial septal defect(ASD)in an swine model. Methods:The model of ASD was built under echocardiography’s guidance after thoracotomy. Fifteen swines were involved and all swines were sacrificed at 1
3
6
9 months after surgery. Doppler echocardiography was used to observe efficacy and measure diameter of occluder disc.Gross anatomy was done to assess tissue coverage and degration. Hematoxylin-eosin staining was done to observe inflammatory reaction. ELISA was used to test the level of IL-6. Results:Except 3 swines died of modeling
12 swines with ASD were built successfully and implanted with biodegradable ASD occluders.No infection or shedding of occluder was observed intra-operative and post-operative. Occluders were placed well and no residual shunt was observed by doppler echocardiography.Occluders’ surface was partly covered with neointima by gross anatomy at 1 month after surgery
while totally covered at 3
6
9 months.Compared with that of pre-operation
diameter of occluder disc decreased at 1
3
6 months after surgery(P <0.05) and decreased significantly with the extension of time(P<0.05);compared with that of pre-operation
occluders’ height decreased at 3 months
6months after surgery(P<0.05)
however
there was little change at 1 month after surgery(P >0.05). Compared with that pre-operation
double disc area of occluder decreased at 1
3
6
9 months after surgery(P<0.05)and decreased significantly with the extension of time(P<0.05). Compared with that pre-operation
serum IL-6 level increased at 3 months and 6 months(P<0.05)
and declined to normal at 9 months after surgery(P>0.05). Conclusion:With satisfactory effectiveness
bio-compatibility and degradation
the fully biodegradable ASD occluder is ideal for swine model of ASD.
Minimally invasive occlusion via thorax for atrial septal defect
Research on effect of chitosan-polylactic acid film on rabbit colonic fistula for primary repair
Minimally invasive surgical closure in the treatment of infants with atrial septal defect and ventricular septal defect
Efficacy and safety of the interventional therapy and surgical repair for atrial septal defects: a systematic review
Efficacy of Parylene coating occluder in the treatment of children with congenial heart disease
Related Author
ZHANG Cheng
JIANG Yingjiu
LI Qiang
WU Qingchen
CHEN Dan
XIE Liang-xian2
LI Ying2
CHEN Kang2
Related Institution
Department of Cardiothoracic Surgery,the First Affiliated Hospital,Chongqing Medical University
Institue for Acupuncture,School of Traditional Chinese Medicine,Chongqing Medical University
Department of General Surgery,PLA Hospital
Mathematics Teaching and Research Room,College of Basic Medicine,Chongqing Medical University
Department of Cardiology,the Children’s Hospital,Chongqing Medical University,Key Laboratory of Child and Development and Disorder Co-founded by Prouincial Government and Ministry Education,Key Laboratory of Pediatrics in Chongqing,Chongqing International Science and Cooperation Center for Child Development and Disorder