孙岩,陈劼,张羽生,等.3 mL低剂量钆对比剂联合3D-TWIST序列在<70 kg人群中CE-MRV诊断准确率与标准剂量等效:一项非随机对照研究[J].陆军军医大学学报,2026,48(17):2408-2417.DOI:,10.16016/j.2097-0927.202604075.
Sun Y,Chen J,Zhang Y S,,et al.A 3 mL low-dose gadolinium-based contrast agent combined with 3D-TWIST sequence yields diagnostic accuracy equivalent to standard-dose CE-MRV in patients weighing <70 kg: a non-randomized controlled study[J].J Army Med Univ,2026,48(17):2408-2417.DOI:,10.16016/j.2097-0927.202604075.
孙岩,陈劼,张羽生,等.3 mL低剂量钆对比剂联合3D-TWIST序列在<70 kg人群中CE-MRV诊断准确率与标准剂量等效:一项非随机对照研究[J].陆军军医大学学报,2026,48(17):2408-2417.DOI:,10.16016/j.2097-0927.202604075. DOI:
Sun Y,Chen J,Zhang Y S,,et al.A 3 mL low-dose gadolinium-based contrast agent combined with 3D-TWIST sequence yields diagnostic accuracy equivalent to standard-dose CE-MRV in patients weighing <70 kg: a non-randomized controlled study[J].J Army Med Univ,2026,48(17):2408-2417.DOI:,10.16016/j.2097-0927.202604075. DOI:
目的
2
针对常规剂量钆对比剂在颅脑对比增强磁共振静脉成像(contrast enhanced-MRV,CE-MRV)中存在的体内沉积风险与安全性问题,探讨低剂量钆对比剂联合3D-时间分辨交叉随机轨迹成像(time-resolved angiography with interleaved stochastic trajectories,TWIST)序列在颅内静脉窦成像中的可行性,评估该方案对图像质量及诊断准确率的影响。
方法
2
采用前瞻性单中心的非随机对照研究,纳入2022年11月至2024年5月在南京医科大学附属苏州医院放射科西门子Skyra 3.0 T磁共振上接受检查的196名患者,所有受检者随机分为A组(
n
=20)、B组(
n
=78)、C组(
n
=98)。B组和C组按体质量分层以10 kg为间隔(40~
<
100 kg),分别对应B4~B9及C4~C9亚组。A组作为金标准:常规剂量(0.1 mmol/kg)钆喷酸葡胺原液,扫描参数为K空间中心区10%/周围区20%的3D-TWIST序列。B组:3 mL钆剂稀释至15 mL注射,扫描参数同A组。C组:3 mL钆剂稀释至15 mL注射,扫描参数为
K空间中心区40%/周围区20%的3D-TWIST序列。分别客观比较B组和C组的不同亚组与A组在颅内静脉窦最大信号强度的信噪比(signal-to-noise ratios,SNR)。主观比较由2名资深放射诊断医生进行评分,分别对A组及B组、C组不同亚组间的颅脑CE-MRV成像质量采用3分标准:1分表示不可见或无法辨别出颅静脉窦血管段,2分表示颅静脉窦血管段中等图像质量显示,3分表示图像对颅静脉窦血管段的高质量显示。进一步分别比较B组、C组不同亚组的颅静脉窦诊断正确率是否与A组间存在差异。
结果
2
A组与B组比较,除B4、B7亚组的下矢状窦外,A组各静脉窦的SNR均高于B组各亚组。A组与C组比较,在C4、C5、C6亚组中,除C4亚组的下矢状窦SNR高于A组(
P
<
0.002 8),其余静脉窦SNR差异均无统计学意义(
P
均
>
0.002 8);在C7、C8、C9亚组中,除C7亚组的下矢状窦外,各静脉窦SNR均低于A组。C组各亚组SNR均高于同体质量B组各亚组。主观图像质量全部评分≥2分,A组与B4~B6、C4~C8各亚组均无差异(
P
均
>
0.004 17)。A组、B组、C组诊断正确率分别为100.00%、98.71%、100.00%,两两比较均无统计学差异(
P
均
>
0.016 7)。
结论
2
低剂量(3 mL)钆对比剂联合优化3D-TWIST序列(K空间中心区40%)在颅脑CE-MRV中可行,可显著降低钆沉积风险。推荐用于体质量<70 kg患者,可替代常规标准剂量方案;70~89 kg患者图像质量主观满足诊断需求,但需进一步提升SNR。
Objective
2
To address the safety concerns regarding gadolinium deposition associated with standard-dose gadolinium-based contrast agents in cerebral contrast-enhanced magnetic resonance venography (CE-MRV)
this study aims to investigate the feasibility of a low-dose gadolinium protocol combined with a 3D time-resolved angiography with stochastic trajectories (TWIST) sequence for intracranial venous sinus imaging
and to evaluate its impact on image quality and diagnostic accuracy.
Methods
2
A prospective
single-center
non-randomized controlled study was conducted on 196 patients who underwent examinations on a Siemens Skyra 3.0 T MR scanner at Department of Radiology of Suzhou Hospital Affiliated to Nanjing Medical University from November 2022 to May 2024. All participants were divided into 3 groups: Group A (
n
=20)
Group B (
n
=78)
and Group C (
n
=98). Groups B and C were further stratified by body weight at 10 kg intervals (40 to
<
100 kg) into subgroups B4
-
B9 and C4
-
C9
respectively. Group A serv
ed as the reference standard: standard-dose (0.1 mmol/kg) gadopentetate dimeglumine administered undiluted
with a 3D-TWIST sequence using 10% central/20% peripheral k-space sampling. Group B received 3 mL of gadolinium-based contrast agent diluted to a total volume of 15 mL
with the same acquisition parameters as Group A. Group C received 3 mL of gadolinium-based contrast agent diluted to 15 mL
with a 3D-TWIST sequence using 40% central/20% peripheral k-space sampling. Objective comparison of the signal-to-noise ratio (SNR) of the maximum signal intensity in intracranial venous sinuses was conducted between different body weight subgroups of Groups B and C versus Group A. Subjective image quality was independently assessed by 2 senior radiologists using a 3-point scoring scale: 1=intracranial venous sinus segments not visible or indistinguishable; 2=moderate image quality with discernible venous sinus segments; 3=high-quality delineation of intracranial venous sinus segments. Furthermore
pairwise comparisons of diagnostic accuracy for intracranial venous sinus evaluation were performed between different subgroups of Group B and Group C versus Group A
respectively.
Results
2
Comparison between Group A and Group B: Except for the inferior sagittal sinus in subgroups B4 and B7
the SNR of all venous sinuses in Group A was higher than that in each subgroup of Group B. Comparison between Group A and Group C showed that among the C4
C5
and C6 subgroups
the SNR of the inferior sagittal sinus in the C4 subgroup was higher than that in group A (
P
<
0.002 8)
while the differences in SNR for the remaining venous sinuses were not statistically significant (all
P
>
0.002 8). In the C7
C8
and C9 subgroups
the SNR of each venous sinus was lower than that in group A
except for the inferior sagittal sinus in the C7 subgroup. The SNR of all subgroups within group C was higher than that of the corresponding subgroups in group B with the same body weight.
All subjective image quality scores were ≥2
with no significant differences were found between group A and subgroups B4
-
B6 and C4
-
C8 (all
P
>
0.004 17). The diagnostic accuracy rates for groups A
B
and C were 100.00%
98.71%
and 100.00%
respectively
with no statistically significant differences between any 2 groups (all
P
>
0.016 7).
Conclusion
2
A low-dose (3 mL) gadolinium-based contrast agent combined with an optimized 3D-TWIST sequence (40% central k-space sampling) is feasible for cerebral CE-MRV and may substantially reduce the risk of gadolinium deposition. This protocol is recommended for patients weighing
<
70 kg as an alternative to the conventional standard-dose regimen. For patients weighing 70
-
89 kg
subjective image quality meets diagnostic requirements
however
further improvements in SNR are warranted.
Lim R P , Hornsey E , Ranatunga D , et al . Upper extremity non-contrast magnetic resonance venography (MRV) compared to contrast enhanced MRV and ultrasound [J]. Clin Imaging , 2017 , 45 : 51 - 57 . DOI: 10.1016/j.clinimag. 2017. 05.020 http://dx.doi.org/10.1016/j.clinimag.2017.05.020 .
戴艳芳 , 叶明 , 吉训明 , 等 . 不同方法的磁共振静脉成像对颅内静脉窦疾病诊断的比较 [J]. 中国脑血管病杂志 , 2012 , 9 ( 6 ): 302 - 306 . DOI: 10.3969/j.issn.1672-5921.2012.06.005 http://dx.doi.org/10.3969/j.issn.1672-5921.2012.06.005 .
Dai Y F , Ye M , Ji X M , et al . Comparison of the diagnosis of intracranial venous sinus disease with different methods of magnetic resonance venography [J]. Chin J Cerebrovascular Dis , 2012 , 9 ( 6 ): 302 - 306 . DOI: 10.3969/j.issn.1672-5921.2012.06.005 http://dx.doi.org/10.3969/j.issn.1672-5921.2012.06.005 .
汤光宇 , 李懋 . 磁共振成像技术与应用 [M]. 上海 : 上海科学技术出版社 , 2023 : 193 - 197 .
Tang G Y , Li M . Magnetic resonance imaging technology and application [M]. Shanghai : Shanghai Science and Technology Publication , 2023 : 193 - 197 .
孔莹 , 李梦双 , 陈柱典 , 等 . 多次使用钆对比剂后MRI平扫小脑齿状核T1WI信号增高的初步研究 [J]. 中华放射学杂志 , 2018 , 52 ( 12 ): 892 - 896 . DOI: 10.3760/cma.j.issn.1005-1201.2018.12.002 http://dx.doi.org/10.3760/cma.j.issn.1005-1201.2018.12.002 .
Kong Y , Li M S , Chen Z D , et al . Increased signal intensity in the dentate nucleus on unenhanced T1-weighted MR images after multiple gadolinium-based contrast material administrations: a preliminary study [J]. Chin J Radiol , 2018 , 52 ( 12 ): 892 - 896 . DOI: 10.3760/cma.j.issn.1005-1201.2018.12.002 http://dx.doi.org/10.3760/cma.j.issn.1005-1201.2018.12.002 .
袁思殊 , 夏黎明 , 杨朝霞 , 等 . 磁共振钆对比剂有关安全性研究现状 [J]. 磁共振成像 , 2020 , 11 ( 8 ): 717 - 720 . DOI: 10.12015/issn.1674-8034.2020.08.030 http://dx.doi.org/10.12015/issn.1674-8034.2020.08.030 .
Yuan S S , Xia L M , Yang Z X , et al . Research status of gadolinium-based contrast on safety [J]. Chin J Magn Reson Imaging , 2020 , 11 ( 8 ): 717 - 720 . DOI: 10.12015/issn.1674-8034.2020.08.030 http://dx.doi.org/10.12015/issn.1674-8034.2020.08.030 .
Mallepally A , Kidd J M , Grizzard J D , et al . Nephrogenic systemic fibrosis in patients with advanced renal dysfunction following gadolinium-based contrast agents [J]. Radiology , 2025 , 317 ( 3 ): e251794 . DOI: 10.1148/radiol.251794 http://dx.doi.org/10.1148/radiol.251794 .
Chandrashekhara S H , Gulati G S , Sharma S , et al . Comparison between time-resolved magnetic resonance angiography and diagnostic digital subtraction angiography in the vascular assessment of nonspecific aorto-arteritis patients: a prospective study [J]. Vasc Endovasc Surg , 2021 , 55 ( 6 ): 586 - 592 . DOI: 10.1177/15385744211010593 http://dx.doi.org/10.1177/15385744211010593 .
Voth M , Haneder S , Huck K , et al . Peripheral magnetic resonance angiography with continuous table movement in combination with high spatial and temporal resolution time-resolved MRA with a total single dose (0.1 mmol/kg) of gadobutrol at 3.0 T [J]. Invest Radiol , 2009 , 44 ( 9 ): 627 - 633 . DOI: 10.1097/RLI.0b013e3181b4c26c http://dx.doi.org/10.1097/RLI.0b013e3181b4c26c .
中华医学会放射学分会磁共振学组 , 中华医学会放射学分会质量控制与安全工作委员会 . 钆对比剂临床安全性应用中国专家建议 [J]. 中华放射学杂志 , 2019 , 53 ( 7 ): 539 - 544 . DOI: 10.3760/cma.j.issn.1005-1201.2019.07.002 http://dx.doi.org/10.3760/cma.j.issn.1005-1201.2019.07.002 .
Magnetic Resonance Group of Chinese Radiology Society , Quality Control and Safety Working Committee of Chinese Radiology Society . Clinical application recommendations on safety of gadolinium contrast agents [J]. Chin J Radiol , 2019 , 53 ( 7 ): 539 - 544 . DOI: 10.3760/cma.j.issn.1005-1201.2019.07.002 http://dx.doi.org/10.3760/cma.j.issn.1005-1201.2019.07.002 .
Woolen S A , Shankar P R , Gagnier J J , et al . Risk of nephrogenic systemic fibrosis in patients with stage 4 or 5 chronic kidney disease receiving a group Ⅱ gadolinium-based contrast agent: a systematic review and meta-analysis [J]. JAMA Intern Med , 2020 , 180 ( 2 ): 223 - 230 . DOI: 10.1001/jamainternmed.2019.5284 http://dx.doi.org/10.1001/jamainternmed.2019.5284 .
中华医学会放射学分会质量控制与安全管理专业委员会 , 卢光明 , 王培军 , 等 . 肾病患者静脉注射钆对比剂应用中国专家共识 [J]. 中华放射学杂志 , 2022 , 56 ( 3 ): 221 - 230 . DOI: 10.3760/cma.j.cn112149-20210330-00294 http://dx.doi.org/10.3760/cma.j.cn112149-20210330-00294 .
Quality Control and Safety Management Committee of Chinese Radiology Society , Lu G M , Wang P J , et al . Chinese expert consensus of gadolinium contrast agent use in patients with renal disease [J]. Chin J Radiol , 2022 , 56 ( 3 ): 221 - 230 . DOI: 10.3760/cma.j.cn112149-20210330-00294 http://dx.doi.org/10.3760/cma.j.cn112149-20210330-00294 .
Domingo J L , Semelka R C . Gadolinium toxicity: mechanisms, clinical manifestations, and nanoparticle role [J]. Arch Toxicol , 2025 , 99 ( 10 ): 3897 - 3916 . DOI: 10.1007/s00204-025-04124-x http://dx.doi.org/10.1007/s00204-025-04124-x .
Vymazal J , Rulseh A M . MRI contrast agents and retention in the brain: review of contemporary knowledge and recommendations to the future [J]. Insights Imaging , 2024 , 15 ( 1 ): 179 . DOI: 10.1186/s13244-024-01763-z http://dx.doi.org/10.1186/s13244-024-01763-z .
van der Molen A J , Quattrocchi C C , Mallio C A , et al . Ten years of gadolinium retention and deposition: ESMRMB-GREC looks backward and forward [J]. Eur Radiol , 2024 , 34 ( 1 ): 600 - 611 . DOI: 10.1007/s00330-023-10281-3 http://dx.doi.org/10.1007/s00330-023-10281-3 .
Mahmood F , Bernchou U , Harboe F S G , et al . Feasibility of a reduced gadolinium dose protocol for MRI-guided radiotherapy in glioblastoma [J]. Acta Oncol , 2025 , 64 : 1185 - 1193 . DOI: 10.2340/1651-226X.2025.44022 http://dx.doi.org/10.2340/1651-226X.2025.44022 .
李少蕊 . Syngo TWIST: 动态时间分辨的MR血管造影技术 [J]. 磁共振成像 , 2012 , 3 ( 5 ): 387 - 391 . DOI: 10.3969/j.issn.1674-8034.2012.05.015 http://dx.doi.org/10.3969/j.issn.1674-8034.2012.05.015 .
Li S R . Syngo TWIST for dynamic time-resolved MR angiography [J]. Chin J Magn Reson Imaging , 2012 , 3 ( 5 ): 387 - 391 . DOI: 10.3969/j.issn.1674-8034.2012.05.015 http://dx.doi.org/10.3969/j.issn.1674-8034.2012.05.015 .
Schabel M C , Parker D L . Uncertainty and bias in contrast concentration measurements using spoiled gradient echo pulse sequences [J]. Phys Med Biol , 2008 , 53 ( 9 ): 2345 - 2373 . DOI: 10.1088/0031-9155/53/9/010 http://dx.doi.org/10.1088/0031-9155/53/9/010 .
Tofts P S . Modeling tracer kinetics in dynamic Gd-DTPA MR imaging [J]. J Magn Reson Imaging , 1997 , 7 ( 1 ): 91 - 101 . DOI: 10.1002/jmri.1880070113 http://dx.doi.org/10.1002/jmri.1880070113 .
Donahue K M , Weisskoff R M , Parmelee T J , et al . Relaxometry of native blood and blood doped with Gd contrast agents [J]. Magn Reson Med , 1995 , 33 ( 5 ): 694 - 701 . DOI: 10.1002/mrm.1910330507 http://dx.doi.org/10.1002/mrm.1910330507 .
Rapacchi S , Natsuaki Y , Plotnik A , et al . Reducing view-sharing using compressed sensing in time-resolved contrast-enhanced magnetic resonance angiography [J]. Magn Reson Med , 2015 , 74 ( 2 ): 474 - 481 . DOI: 10.1002/mrm.25414 http://dx.doi.org/10.1002/mrm.25414 .
Korosec F R , Frayne R , Grist T M , et al . Time-resolved contrast-enhanced 3D MR angiography [J]. Magn Reson Med , 1996 , 36 ( 3 ): 345 - 351 . DOI: 10.1002/mrm.1910360304 http://dx.doi.org/10.1002/mrm.1910360304 .
Hennig J , Scheffler K , Laubenberger J , et al . Time-resolved projection angiography after bolus injection of contrast agent [J]. Magn Reson Med , 1997 , 37 ( 3 ): 341 - 345 . DOI: 10.1002/mrm.1910370306 http://dx.doi.org/10.1002/mrm.1910370306 .
Laub G , Kroeker R . Syngo TWIST for dynamic time-resolved MR angiography [J]. MAGNETOM Flash , 2006 , 34 ( 3 ): 92 - 95 .
颜雨暄 , 岳姣 , 姜永莉 , 等 . 丹参酮ⅡA与灯盏乙素协同抑制PKCδ/NLRC4/Caspase-1通路改善缺血性脑卒中的机制 [J]. 空军军医大学学报 , 2026 , 47 ( 2 ): 273 - 283 . DOI: 10.13276/j.issn.2097-1656.2026.02.016 http://dx.doi.org/10.13276/j.issn.2097-1656.2026.02.016 .
Yan Y X , Yue J , Jiang Y L , et al . Mechanism of tanshinone ⅡA and scutellarin in synergistically ameliorating ischemic stroke via inhibition of the PKCδ/NLRC4/Caspase-1 pathway [J]. J Air Force Med Univ , 2026 , 47 ( 2 ): 273 - 283 . DOI: 10.13276/j.issn.2097-1656.2026.02.016 http://dx.doi.org/10.13276/j.issn.2097-1656.2026.02.016 .
Raczeck P , Fries P , Massmann A , et al . Diagnostic performance of a lower-dose contrast-enhanced 4D dynamic MR angiography of the lower extremities at 3 T using multisegmental time-resolved maximum intensity projections [J]. Magn Reson Imaging , 2021 , 54 ( 3 ): 763 - 774 . DOI: 10.1002/jmri.27631 http://dx.doi.org/10.1002/jmri.27631 .
Endler C H , Jost G , Pietsch H , et al . Effect of contrast agent dose reduction on vascular enhancement and image quality in thoracoabdominal dynamic 3-dimensional magnetic resonance angiography: a systematic intraindividual analysis in pigs [J]. Invest Radiol , 2022 , 57 ( 10 ): 689 - 695 . DOI: 10.1097/RLI.0000000000000882 http://dx.doi.org/10.1097/RLI.0000000000000882 .
甄志铭 , 肖诗蕾 , 刘晨 , 等 . 7.0T MRI SWI序列对脉络丛静脉成像质量的评估价值 [J]. 陆军军医大学学报 , 2023 , 45 ( 15 ): 1651 - 1655 . DOI: 10.16016/j.2097-0927.202302042 http://dx.doi.org/10.16016/j.2097-0927.202302042 .
Zhen Z M , Xiao S L , Liu C , et al . Evaluation value of SWI with 7.0T MRI for choroid plexus vein imaging [J]. J Army Med Univ , 2023 , 45 ( 15 ): 1651 - 1655 . DOI: 10.16016/j.2097-0927.202302042 http://dx.doi.org/10.16016/j.2097-0927.202302042 .
0
浏览量
10
下载量
0
CSCD
关联资源
相关文章
相关作者
相关机构
京公网安备11010802024621