李阳, 彭滔滔, 左都坤, 等. 低氧条件下不同品牌近红外光谱仪测定脑氧饱和度数值的比较——一项健康志愿者的随机对照研究[J]. 陆军军医大学学报, 2023,45(8):772-779.
Yang LI, Taotao PENG, Dukun ZUO, et al. Comparison of cerebral oxygen saturation values measured by different brands of near-infrared spectrometers under hypoxic conditions: a randomized controlled trial in healthy volunteers[J]. Journal of Army Medical University, 2023, 45(8): 772-779.
李阳, 彭滔滔, 左都坤, 等. 低氧条件下不同品牌近红外光谱仪测定脑氧饱和度数值的比较——一项健康志愿者的随机对照研究[J]. 陆军军医大学学报, 2023,45(8):772-779. DOI: 10.16016/j.2097-0927.202209257.
Yang LI, Taotao PENG, Dukun ZUO, et al. Comparison of cerebral oxygen saturation values measured by different brands of near-infrared spectrometers under hypoxic conditions: a randomized controlled trial in healthy volunteers[J]. Journal of Army Medical University, 2023, 45(8): 772-779. DOI: 10.16016/j.2097-0927.202209257.
目的
2
通过健康志愿者降氧试验对比INVOS 5100C(美敦力)
EGOS-600(爱琴)和BRS-1(中科搏锐)脑氧测量值的变化情况,仪器的一致性、相关性以及检测关键去饱和事件的能力。
方法
2
2021年12月6日至2022年1月18日,招募陆军军医大学第二附属医院健康志愿者20名。采用随机数字表法将受试者分为两组,其中E组使用INVOS 5100C和EGOS-600的单个传感器,B组使用INVOS 5100C和BRS-1的传感器,每组10人。将各自的单个传感器贴于受试者前额后,通过调节吸入氮气和氧气的比例来诱导和逆转缺氧,并连续记录局部脑组织氧饱和度(regional cerebral oxygen saturation,rSO
2
),受试者的心率(heart rate,HR)、脉搏氧饱和度(pulse oxygen saturation,SPO
2
)随时间的变化情况。统计方法使用Wilcoxon检验评估各设备之间rSO
2
基线值及变化情况。Bland-Altman和Deming回归分析两组仪器的一致性及相关性。受试者工作特征曲线(receiver operator characteristic curve
ROC)分析不同仪器检测低氧的效能。
结果
2
所有设备都对吸入氧浓度的变化做出了相似的反应。①EGOS-600和BRS-1的rSO
2
变化幅度及速率均小于INVOS 5100C,INVOS 5100C、EGOS-600和BRS-1的ΔrSO
2
分别为(16%~34.15%)、(6%~20.81%)和(5%~23.81%),绝对中位数变化率分别为(2.81~4.40%/min)、(1.50~2.14%/min)及(1.77~2.41%/min)。②Bland-Altman分析显示EGOS-600和BRS-1与INVOS 5100C的差异均值分别为5.7、2.5;Deming回归显示Person系数分别为0.722 4、0.845 9,
P
<
0.001,与INVOS 5100C的一致性及相关性均较好。③以INVOS 5100C相对降低10%、20%和50%的绝对值作为关键的干预性rSO
2
阈值方面,EGOS-600检出率为90%、48%和40%,BRS-1检出率为80%、48%和50%。④ROC结果显示EGOS-600和BRS-1的AUC值分别为0.893、0.715,缺氧预测阈值(cut-off)分别为59.3%、56%。
结论
2
EGOS-600和BRS-1对氧合变化的反应与INVOS 5100C相似,相关性及一致性较好,可以准确地预测脑氧变化。
Objective
2
By comparing the differences in cerebral oxygen parameters with 3 brands of near-infrared spectrometers
INVOS 5100C (Medtronic)
EGOS-600 (Aegean) and BRS-1 (Casibrain Technology)
to investigate their consistency
correlation and ability to detect key desaturation events in healthy volunteers in a hypoxic condition.
Methods
2
From December 6
2021 to January 18
2022
20 healthy volunteers were recruited from our hospital. They were randomly divided into group E (a single sensor of INVOS 5100C at the left forehead and that of EGOS-600 at the right) and group B (INVOS 5100C and BRS-1
as the other group)
with 10 subjects in each group. After the sensors were applied to the forehead of the subject
hypoxia was induced and reversed by adjusting the ratio of inhaled nitrogen to oxygen. Regional cerebral oxygen saturation (rSO
2
)
heart rate and pulse oxygen saturation (SPO
2
) were continuously monitored and recorded in the whole process. Wilcoxon test was used to assess the baseline value and changes in rSO
2
between different devices. Bland-Altman and Deming regression analyses were employed to analyze the consistency and correlation between the 2 devices. Receiver operating characteristic (ROC) curve was plotted to analyze the performance of the different devices in detection of hypoxia.
Results
2
All 3 devices responded similarly to the changes in inhaled oxygen concentration. ① The magnitude and rate of changes in rSO
2
for EGOS-600 and BRS-1 were less than those of INVOS 5100C. The ΔrSO
2
for INVOS 5100C
EGOS-600 and BRS-1 was 16%~34.15%
6%~20.81% and 5%~23.81%
with an absolute median rate of change of (2.81~4.40)%/min
(1.50~2.14)%/min and (1.77~2.41)%/min
respectively. ②Bland-Altman analysis showed that the mean difference of EGOS-600 and BRS-1 to INVOS 5100C were 5.7 and 2.5 respectively. Deming regression displayed that the Person coefficient was 0.722 4 and 0.845 9
respectively (
P
<
0.001)
which were in good consistency and correlation with INVOS 5100C. ③In terms
of absolute values of 10%
20% and 50% relative reduction in INVOS 5100C as the key interventional rSO
2
thresholds
EGOS-600 detection rates were 90%
48% and 40% and BRS-1 detection rates were 80%
48% and 50%. ④The ROC curve results indicated the AUC value was 0.893 and 0.715 for EGOS-600 and BRS-1
respectively
and the cut-off threshold for hypoxia prediction was 59.3% and 56.0%
respectively.
Conclusion
2
The response of EGOS-600 and BRS-1 to oxygenation changes is similar to that of INVOS 5100C
and the correlation and consistency are good
which can accurately predict brain oxygen changes.
麻醉学领域发展科学和临床规范《2017版中国麻醉学指南与专家共识》发布[J]. 中华医学信息导报, 2018(4): 7.
The 2017 Chinese Anesthesiology Guidelines and Expert Consensus was released as a scientific and clinical norm for the development of anesthesiology[J ] . China Med News, 2018(4): 7..
UYSAL S, LIN H M, TRINH M, et al . Optimizing cerebral oxygenation in cardiac surgery: A randomized controlled trial examining neurocognitive and perioperative outcomes[J ] . J Thorac Cardiovasc Surg, 2020, 159(3): 943-953. DOI: 10.1016/j.jtcvs.2019.03.036
OH C S, SA M, PARK H J, et al . Effects of remote ischemic preconditioning on regional cerebral oxygen saturation in patients in the beach chair position during shoulder surgery: a double-blind randomized controlled trial[J ] . J Clin Anesth, 2020, 61: 109661. DOI: 10.1016/j.jclinane.2019.109661.
SCHOLKMANN F, KLEISER S, METZ A J, et al . A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodology[J ] . NeuroImage, 2014, 85: 6-27. DOI: 10.1016/j.neuroimage.2013.05.004.
PISANO A, GALDIERE N, IOVINO T P, et al . Direct comparison between cerebral oximetry by INVOS TM and EQUANOX TM during cardiac surgery: a pilot study. [J ] . Heart, lung and vessels, 2014, 6(3)..
HESSEL T W, HYTTEL SORENSEN S, GREISEN G. Cerebral oxygenation after birth-a comparison of INVOS ® and FORE-SIGHT TM near-infrared spectroscopy oximeters[J ] . Acta Paediatrica, 2014, 103(5): 488-493. DOI: 10.1111/apa.12567..
TOMLIN K L, NEITENBACH A, BORG U. Detection of critical cerebral desaturation thresholds by three regional oximeters during hypoxia: a pilot study in healthy volunteers[J]. BMC Anesthesiology, 2017, 17(1). DOI: 10.1186/s12871-016-0298-7.
VALENCIA L, RODRIGUEZ-PEREZ A, OJEDA N, et al . Baseline cerebral oximetry values depend on non-modifiable patient characteristics[J ] . Anaesth Crit Care Pain Med, 2015, 34(6): 345-348. DOI: 10.1016/j.accpm.2015.06.008.
DIX L M L, VAN BEL F, BAERTS W, et al . Comparing near-infrared spectroscopy devices and their sensors for monitoring regional cerebral oxygen saturation in the neonate[J ] . Pediatr Res, 2013, 74(5): 557-563. DOI: 10.1038/pr.2013.133.
BICKLER P E, FEINER J R, ROLLINS M D. Factors affecting the performance of 5 cerebral oximeters during hypoxia in healthy volunteers[J]. Anesth Analg, 2013, 117(4): 813-823. DOI: 10.1213/ANE.0b013e318297d763.
SCHNEIDER A, MINNICH B, HOFSTATTER E, et al . Comparison of four near-infrared spectroscopy devices shows that they are only suitable for monitoring cerebral oxygenation trends in preterm infants[J ] . Acta Paediatr, 2014, 103(9): 934-938. DOI: 10.1111/apa.12698.
HYTTEL-SORENSEN S, SORENSEN L C, RIERA J, et al . Tissue oximetry: a comparison of mean values of regional tissue saturation, reproducibility and dynamic range of four NIRS-instruments on the human forearm[J ] . Biomed Opt Express, 2011, 2(11): 3047-3057. DOI: 10.1364/BOE.2.003047.
GOMERSALL C D, LEUNG P L, GIN T, et al . A comparison of the Hamamatsu NIRO 500 and the INVOS 3100 near-infrared spectrophotometers[J ] . Anaesth Intensive Care, 1998, 26(5): 548-557. DOI: 10.1177/0310057X9802600512.
MOHANDAS B S, JAGADEESHA M, VIKRAM S B. Impact of monitoring cerebral oxygen saturation on the outcome of patients undergoing open heart surgery[J]. Ann Card Anaesth, 2013, 16(2): 102-106. DOI: 10.4103/0971-9784.109740.
DESCHAMPS A, HALL R, GROCOTT H, et al . Cerebral oximetry monitoring to maintain normal cerebral oxygen sat uration during high-risk cardiac surgery: a randomized controlled feasibility trial[J ] . Anesthesiology, 2016, 124(4): 826-836. DOI: 10.1097/ALN.0000000000001029.
HYTTEL-SORENSEN S, HESSEL T W, GREISEN G. Peripheral tissue oximetry: comparing three commercial near-infrared spectroscopy oximeters on the forearm[J]. J Clin Monit Comput, 2014, 28(2): 149-155. DOI: 10.1007/s10877-013-9507-9.
KOBAYASHI K, KITAMURA T, KOHIRA S, et al . Cerebral oximetry for cardiac surgery: a preoperative comparison of device characteristics and pitfalls in interpretation[J ] . J Artif Organs, 2018, 21(4): 412-418. DOI: 10.1007/s10047-018-1052-3.
KIM H S, HA S O, YU K H, et al . Cerebral qxygenation as a monitoring parameter for mortality during venoarterial extracorporeal membrane oxygenation[J ] . ASAIO J, 2019, 65(4): 342-348. DOI: 10.1097/MAT.0000000000000827.
ITO H, TAKAHASHI K, HATAZAWA J, et al . Changes in human regional cerebral blood flow and cerebral blood volume during visual stimulation measured by positron emission tomography[J ] . J Cereb Blood Flow Metab, 2001, 21(5): 608-612. DOI: 10.1097/00004647-200105000-00015.