白露华, 罗婷婷, 唐家伟, 等. 不同超声脉冲宽度和脉冲重复频率组合激励微泡空化对肿瘤血流灌注及释药的影响[J]. 陆军军医大学学报, 2022,44(9):935-942.
Luhua BAI, Tingting LUO, Jiawei TANG, et al. Effects of ultrasound combinations of different pulse lengths and pulse repetition frequencies to stimulate microbubble cavitation on tumor perfusion and drug delivery in mice[J]. Journal of Army Medical University, 2022, 44(9): 935-942.
白露华, 罗婷婷, 唐家伟, 等. 不同超声脉冲宽度和脉冲重复频率组合激励微泡空化对肿瘤血流灌注及释药的影响[J]. 陆军军医大学学报, 2022,44(9):935-942. DOI: 10.16016/j.2097-0927.202202130.
Luhua BAI, Tingting LUO, Jiawei TANG, et al. Effects of ultrasound combinations of different pulse lengths and pulse repetition frequencies to stimulate microbubble cavitation on tumor perfusion and drug delivery in mice[J]. Journal of Army Medical University, 2022, 44(9): 935-942. DOI: 10.16016/j.2097-0927.202202130.
目的
2
机械指数(mechanical index,MI)和占空比固定时,探讨超声不同脉冲宽度(pulse length,PL)和脉冲重复频率(pulse repetition frequency,PRF)组合激励微泡空化(ultrasound stimulated microbubble,USMB)对C57BL/6小鼠肿瘤组织血流灌注和化疗药物释放的影响。
方法
2
健康6~8周龄雌性C57BL/6荷瘤小鼠24只,体质量(17.74±0.43)g,随机分为3组(
n
=8):对照组(A组,超声假照),PL3.0+PRF2000+DOX组(B组,较高PRF较低PL超声辐照),PL34.5+PRF200+DOX(C组,较高PL较低PRF超声辐照)。USMB治疗前后分别进行超声造影(contrast-enhanced ultrasound,CEUS),比较各组肿瘤峰值强度(peak intensity,PI)、曲线下面积(area under curve,AUC)和肿瘤血流灌注面积百分比。治疗后缓慢推注阿霉素(doxorubicin,DOX),检测肿瘤组织中DOX药物浓度;病理切片观察小鼠肿瘤组织微结构。
结果
2
治疗后,A组的PI、AUC值与治疗前相比无显著差别(
P
>
0.05);B组的PI、AUC值较治疗前分别增加19.3%、18.0%(
P
<
0.05);C组的PI、AUC值较治疗前分别减少16.0%、15.8%(
P
<
0.05)。治疗后,A组的肿瘤灌注面积较治疗前无显著差别(
P
>
0.05);B组的肿瘤灌注面积较治疗前增加21.0%(
P
<
0.05);C组的肿瘤灌注面积较治疗前减少8.2%(
P
<
0.05)。高效液相分析发现B组DOX药物浓度分别是A组的5.0倍和C组的4.8倍(
P
<
0.05);A组和C
组DOX药物浓度无显著差别(
P
>
0.05)。病理结果发现A组和B组肿瘤组织微血管呈轻度和中度扩张,而C组肿瘤微血管破裂出血。
结论
2
在相同的MI和相似占空比条件下,较低脉冲宽度和较高脉冲重复频率超声组合激励微泡空化可增强肿瘤血供、促进化疗药物释放;较高脉冲宽度和较低脉冲重复频率超声组合可阻断肿瘤血供。
Objective
2
To determine the effects of ultrasound stimulated microbubble (USMB) by different ultrasound combinations of pulse length (PL) and pulse repetition frequency (PRF) under fixed mechanical index and duty cycle on tumor perfusion and drug delivery in C57BL/6 mice.
Methods
2
Twenty-four female C57BL/6 mice (6~8 weeks old
average body weight 17.74±0.43 g) were randomly divided into 3 groups (
n
=8): control group (group A)
DOX+PL3.0+PRF2000 group (group B) and DOX+PL30.5+PRF200 group (group C) after being carried with MC38 xenograft tumor. Contrast-enhanced ultrasound (CEUS) was performed before and after USMB treatment to compare the peak intensity (PI)
area under the curve (AUC) and tumor perfusion percentage. Doxorubicin (DOX) was injected slowly after USMB treatment
and then the concentration of DOX in the tumor was detected with high-performance liquid chromatography (HPLC). Microstructure of the tumor was observed pathologically.
Results
2
The values of PI and AUC after treatment had no significant differences when compared with those before treatment in group A (
P
>
0.05)
the values were increased by 19.3% and 18.0%
respectively in group B (
P
<
0.05)
and decreased by 16.0% and 15.8% in group C (
P
<
0.05). Compared with the tumor perfusion area before treatment
there was no significant difference in the area in group A (
P
>
0.05)
and the area increased by 21.0% in group B (
P
<
0.05)
and decreased by 8.2% in group C after treatment (
P
<
0.05). The results of HPLC showed that the concentration of Dox in gro
up B was 5.0 times higher than that of group A
and 4.8 times higher than that in group C (
P
<
0.05). No statistical difference was seen in the DOX concentration between group A and group C (
P
>
0.05). The pathological results showed that the tumor microvessels in group A and B were slightly and moderately dilated
while those in group C were ruptured and bleeding.
Conclusion
2
Under the fixed mechanical index and duty cycle
USMB with higher PRF and lower PL can enhance tumor blood perfusion and promote drug delivery. However
higher PL and lower PRF can block tumor blood perfusion.
GRAHAM K, UNGER E. Overcoming tumor hypoxia as a barrier to radiotherapy, chemotherapy and immunotherapy in cancer treatment[J]. Int J Nanomed, 2018, 13: 6049-6058. DOI: 10.2147/IJN.S140462.
JING X M, YANG F M, SHAO C C, et al . Role of hypoxia in cancer therapy by regulating the tumor microenvironment[J ] . Mol Cancer, 2019, 18(1): 157. DOI: 10.1186/s12943-019-1089-9.
FILIPCZAK N, JOSHI U, ATTIA S A, et al . Hypoxia-sensitive drug delivery to tumors[J ] . J Control Release, 2022, 341: 431-442. DOI: 10.1016/j.jconrel.2021.11.034.
TRINER D, SHAH Y M. Hypoxia-inducible factors: a central link between inflammation and cancer[J]. J Clin Invest, 2016, 126(10): 3689-3698. DOI: 10.1172/JCI84430.
DEPREZ J, LAJOINIE G, ENGELEN Y, et al . Opening doors with ultrasound and microbubbles: beating biological barriers to promote drug delivery[J ] . Adv Drug Deliv Rev, 2021, 172: 9-36. DOI: 10.1016/j.addr.2021.02.015.
SNIPSTAD S, VIKEDAL K, MAARDALEN M, et al . Ultrasound and microbubbles to beat barriers in tumors: improving delivery of nanomedicine[J ] . Adv Drug Deliv Rev, 2021, 177: 113847. DOI: 10.1016/j.addr.2021.113847.
FERIL L B Jr, KONDO T, TAKAYA K, et al . Enhanced ultrasound-induced apoptosis and cell Lysis by a hypotonic medium[J ] . Int J Radiat Biol, 2004, 80(2): 165-175. DOI: 10.1080/09553000310001654684.
CHEN H, KREIDER W, BRAYMAN A A, et al . Blood vessel deformations on microsecond time scales by ultrasonic cavitation[J ] . Phys Rev Lett, 2011, 106(3): 034301. DOI: 10.1103/PhysRevLett.106.034301.
POSTEMA M, GILJA O H. Contrast-enhanced and targeted ultrasound[J]. World J Gastroenterol, 2011, 17(1): 28-41. DOI: 10.3748/wjg.v17.i1.28.
FENG S, QIAO W, TANG J W, et al . Chemotherapy augmentation using low-intensity ultrasound combined with microbubbles with different mec hanical indexes in a pancreatic cancer model[J ] . Ultrasound Med Biol, 2021, 47(11): 3221-3230. DOI: 10.1016/j.ultrasmedbio.2021.07.004.
LI N S, TANG J W, YANG J, et al . Tumor perfusion enhancement by ultrasound stimulated microbubbles potentiates PD-L1 blockade of MC38 colon cancer in mice[J ] . Cancer Lett, 2021, 498: 121-129. DOI: 10.1016/j.canlet.2020.10.046.
MARTIN K H, DAYTON P A. Current status and prospects for microbubbles in ultrasound theranostics[J]. Wiley Interdiscip Rev Nanomed Nanobiotechnol, 2013, 5(4): 329-345. DOI: 10.1002/wnan.1219.
CHOWDHURY S M, ABOU-ELKACEM L, LEE T, et al . Ultrasound and microbubble mediated therapeutic delivery: underlying mechanisms and future outlook[J ] . J Control Release, 2020, 326: 75-90. DOI: 10.1016/j.jconrel.2020.06.008.
TZU-YIN W, WILSON K E, MACHTALER S, et al . Ultrasound and microbubble guided drug delivery: mechanistic understanding and clinical implications[J ] . Curr Pharm Biotechnol, 2013, 14(8): 743-752. DOI: 10.2174/1389201014666131226114611.
KODAMA T, TOMITA Y, KOSHIYAMA K I, et al . Transfection effect of microbubbles on cells in superposed ultrasound waves and behavior of cavitation bubble[J ] . Ultrasound Med Biol, 2006, 32(6): 905-914. DOI: 10.1016/j.ultrasmedbio.2006.03.004.
KHOKHLOVA T D, HAIDER Y, HWANG J H. Therapeutic potential of ultrasound microbubbles in gastrointestinal oncology: recent advances and future prospects[J]. Ther Adv Gastroenterol, 2015, 8(6): 384-394. DOI: 10.1177/1756283X15592584..
唐娜娇, 唐家伟, 张毅, 等. 微泡超声空化增强乏血供肿瘤血流灌注的实验研究[J]. 中华超声影像学杂志, 2021, 30(2): 167-172. DOI: 10.3760/cma.j.cn131148-20200902-00713.
TANG N J, TANG J W, ZHANG Y, et al . Perfusion augmentation in hypovascular tumor by ultrasound stimulated microbubble cavitation[J ] . Chin J Ultrason, 2021, 30(2): 167-172. DOI: 10.3760/cma.j.cn131148-20200902-00713.
LI N S, LIAO Y Y, TANG J W. Characteristic blood-perfusion reduction of walker 256 tumor induced by diagnostic ultrasound and microbubbles[J]. Ultrasound Med Biol, 2020, 46(11): 3069-3079. DOI: 10.1016/j.ultrasmedbio.2020.07.009.
冯爽, 乔伟, 益磋, 等. 不同机械指数诊断超声联合微泡对裸鼠胰腺癌移植瘤化疗的影响[J]. 中国介入影像与治疗学, 2019(2): 107-111.
FENG S, QIAO W, YI C, et al . Impact of diagnostic ultrasound with different mechanical index en hanced by microbubbles on chemotherapy of nude mice bearing pancreatic tumor[J ] . Chin J Interv Imaging Ther, 2019(2): 107-111.
MORSE S V, MISHRA A, CHAN T G, et al . Liposome delivery to the brain with rapid short-pulses of focused ultrasound and microbubbles[J ] . J Control Release, 2022, 341: 605-615. DOI: 10.1016/j.jconrel.2021.12.005.
ARANGO-RESTREPO A, RUBI J M, KJELSTRUP S, et al . Enhancing carrier flux for efficient drug delivery in cancer tissues[J ] . Biophys J, 2021, 120(23): 5255-5266. DOI: 10.1016/j.bpj.2021.10.036.
KELLER S, BRUCE M, AVERKIOU M A. Ultrasound imaging of microbubble activity during sonoporation pulse sequences[J]. Ultrasound Med Biol, 2019, 45(3): 833-845. DOI: 10.1016/j.ultrasmedbio.2018.11.011.
MASON O R, DAVIDSON B P, SHEERAN P, et al . Augmentation of tissue perfusion in patients with peripheral artery disease using microbubble cavitation[J ] . JACC Cardiovasc Imaging, 2020, 13(3): 641-651. DOI:10.1016/j.jcmg.2019.06.012.
0
浏览量
0
下载量
0
CSCD
关联资源
相关文章
相关作者
相关机构
京公网安备11010802024621