彭滔滔, 任玉坤, 詹剑, 等. 4种近红外脑氧仪模拟4 500 m高海拔环境的检测效果对比[J]. 第三军医大学学报, 2021,43(18):1813-1820.
Taotao PENG, Yukun REN, Jian ZHAN, et al. Detection effects of 4 near-infrared cerebral oximeters: a comparative study in simulated high altitude (4 500 m)[J]. Journal of Third Military Medical University, 2021, 43(18): 1813-1820.
彭滔滔, 任玉坤, 詹剑, 等. 4种近红外脑氧仪模拟4 500 m高海拔环境的检测效果对比[J]. 第三军医大学学报, 2021,43(18):1813-1820. DOI: 10.16016/j.1000-5404.202103201.
Taotao PENG, Yukun REN, Jian ZHAN, et al. Detection effects of 4 near-infrared cerebral oximeters: a comparative study in simulated high altitude (4 500 m)[J]. Journal of Third Military Medical University, 2021, 43(18): 1813-1820. DOI: 10.16016/j.1000-5404.202103201.
目的
2
评价INVOS 5100C(美敦力)、EGOS-600(爱琴)和MNIR-P100(名希)、BRS-1(中科博锐)等4种近红外脑氧仪在高海拔环境中的检测效果。
方法
2
2021年3月3日~2021年3月16日招募到陆军军医大学男性健康志愿者174名,年龄18~22岁、身高(172.8±5.1)cm、BMI(22.5±2.2)kg/m
2
,用随机数字表法将受试者分为6组,每组29人,分别在平原(海拔300 m)和模拟4 500 m海拔高度时采用4种近红外脑氧仪随机依次测量受试者的脑氧饱和度(regional cerebral oxygen saturation,rcSO
2
),记录仪器的响应时间、受试者的血压(blood pressure,BP)、心率(heart rate,HR)、脉搏氧饱和度(pulse oxygen saturation,SPO
2
),对比4种仪器检测响应时间、rcSO
2
检测值之间的差异和一致性、rcSO
2
与SpO
2
之间的相关性。
结果
2
① 4种脑氧仪在4 500 m高原环境中检测成功率均为100%。INVOS 5100C、EGOS-600、MNIR-P100、BRS-1在平原响应时间分别为(2.13±0.34)s、(6.60±1.23)s、(2.87±0.48)s、(2.65±0.47)s,在高原响应时间分别为(2.22±0.41)、(6.83±1.25)、(2.86±0.49)、(2.67±0.49)s,其中EGOS-600的响应时间较其余3种仪器平均延长3 s以上(
P
<
0.05)。MNIR-P100和BRS-1的响应时间与平原相比无显著差异(
P
>
0.05),INVOS 5100C和EGOS-600的响应时间与平原相比分别平均增加了约0.1 s和0.2 s(
P
<
0.05)。②在平原(300 m)BRS-1和INVOS 5100C的rcSO
2
测量值一致性较EGOS-600和MNIR-P100更好,在模拟4500 m海拔高度,EGOS-600和INVOS 5100C的rcSO
2
测量值一致性较BRS-1和MNIR-P100更好。受试者从平原至模拟高原环境时,BRS-1与INVOS 5100C测量rcSO
2
下降幅度一致性比EGOS-600和MNIR-P100更好。③当INVOS
5100C达到以rcSO
2
测量值下降超过基础值的20%为脑缺氧干预指标时,EGOS-600、MNIR-P100、BRS-1分别有78%、94%、76%的受试者未达到;当INVOS 5100C达到rcSO
2
为50%绝对阈值作为脑缺氧干预指标时,EGOS-600、MNIR-P100、BRS-1分别有52.9%、54.6%、54.3%的受试者未达到。④INVOS 5100C、EGOS-600、MNIR-P100、BRS-1测量rcSO
2
与SpO
2
相关性系数分别为0.876 4(
P
<
0.000 1)、0.641 7(
P
<
0.000 1)、0.761 8(
P
<
0.000 1)、0.652 5(
P
<
0.000 1)。
结论
2
INVOS 5100C与3种国产仪器测量rcSO
2
一致性较好,3种国产仪器反映rcSO
2
降低的特异性较好,但敏感性稍差。
Objective
2
To evaluate the detection effects of 4 brands of near-infrared cerebral oximeters (NICOs)
INVOS 5100C (Medtronic)
EGOS-600 (Aegean)
MNIR-P100 (Mingxi) and BRS-1 (Casibrain Technology) in high-altitude environment.
Methods
2
From March 3
2021 to March 16
2021
a total of 174 healthy male volunteers (18~22 years old
172.8±5.1 cm
BMI 22.5±2.2 kg/m
2
) were subjected from Army Medical University
and then randomly divided into 6 groups (n=29). Their regional cerebral oxygen saturation (rcSO
2
)
instrument response time
blood pressure (BP)
heart rate (HR)
and pulse oxygen saturation (SPO
2
) were recorded and compared in the plain (300 m) and the 4 500-m altitude environment. The correlations among the above indicators were evaluated by Pearson correlation analysis. The difference and consistency of above indicators by the 4 devices were investigated.
Results
2
① The effective detection rates of all 4 NICOs were 100% in 4 500 m environment. In the plain environment
the response time was 2.13±0.34
6.60±1.23
2.87±0.48 and 2.65±0.47 s for INVOS 5100C
EGOS-600
MNIR-P100 and BRS-1
respectively
while the time was 2.22±0.41
6.83±1.25
2.86±0.49 and 2.67±0.49 s respectively in the 4 500-m altitude environment. EGOS-600 had the longest response time
with mo
re than 3 s longer than the other 3 NICOs (
P
<
0.05). There was no obvious difference in the response time in MNIR-P100 and BRS-1 between the plain and high altitude environment (
P
>
0.05). But the response time of INVOS 5100C and EGOS-600 was increased by 0.1 and 0.2 s respectively in the 4 500 m altitude environment than the plain (
P
<
0.05). ②In the plain environment
INVOS 5100C and BRS-1 showed better consistency in rcSO
2
detection than EGOS-600 and MNIR-P100. While in the 4 500 m altitude environment
EGOS-600 and INVOS 5100C performed better consistency than BRS-1 and MNIR-P100. When the subjects entered from the plain to 4 500 m altitude environment
the decline range showed better consistency in BRS-1 and INVOS 5100C than EGOS-600 and MNIR-P100. ③When taking the decline range of rcSO
2
reaching 20% as the intervention value of cerebral hypoxia
78%
94% and 76% of the subjects respectively detected by EGOS-600
MNIR-P100 and BRS-1 were under the threshold. When taking the decline range of rcSO
2
reaching 50% as the intervention value of cerebral hypoxia
52.9%
54.6% and 54.3% of the subjects respectively detected by EGOS-600
MNIR-P100 and BRS-1 were under the threshold. ④The correlation coefficients of SpO
2
and rcSO
2
detected by INVOS 5100C
EGOS-600
MNIR-P100 and BRS-1 were 0.876 4 (
P
<
0.000 1)
0.641 7 (
P
<
0.000 1)
0.761 8 (
P
<
0.000 1) and 0.6525 (
P
<
0.0001)
respectively.
Conclusion
2
In terms of rcSO
2
INVOS 5100C shows great consistency with the 3 domestic NICOs
and the 3 domestic instruments also show great specificity but less good sensitivity.
高钰琪, 黄庆愿, 刘福玉, 等. 促进高原习服措施的研究进展[J]. 解放军预防医学杂志, 2002, 20(4): 306-309. DOI: 10.3969/j.issn.1001-5248.2002.04.032.
GAO Y Q, HUANG Q Q, LIU F Y, et al . Research progress on measures to promote acclimatization at high altitude[J ] . J Prev Med Chin, 2002, 20(4): 306-309. DOI: 10.3969/j.issn.1001-5248.2002.04.032.
吴世政. 高原脑科学研究进展[J]. 高原医学杂志, 2019, 29(1): 47-53. DOI: CNKI:SUN:GYYZ.0.2019-01-010.
WU S Z. Research progress of high altitude brain[J]. J High Alt Med, 2019, 29(1): 47-53. DOI: CNKI:SUN:GYYZ.0.2019-01-010.
廉国锋, 李锏, 罗勇军, 等. 高原脑水肿发病机制及防治研究进展[J]. 人民军医, 2020, 63(4): 343-346, 357. DOI: CNKI:SUN:RMJZ.0.2020-04-013.
LIAN G F, LI J, LUO Y J, et al . Research progress on the pathogenesis and prevention of high altitude cerebral edema[J ] . People Mil Surg, 2020, 63(4): 343-346, 357. DOI: CNKI:SUN:RMJZ.0.2020-04-013.
IMRAY C H, CLARKE T, FORSTER P J, et al . Carbon dioxide contributes to the beneficial effect of pressurization in a portable hyperbaric chamber at high altitude[J ] . Clin Sci (Lond), 2001, 100(2): 151-157. DOI: 10.1042/CS20000058.
HADOLT I, LITSCHER G. Noninvasive assessment of cerebral oxygenation during high altitude trekking in the Nepal Himalayas (2850-5600 m)[J]. Neurol Res, 2003, 25(2): 183-188. DOI: 10.1179/016164103101201175.
邱宇, 戚思华. 无创脑氧饱和度监测技术的研究现状[J]. 现代医学, 2019, 47(8): 1045-1048. DOI: CNKI:SUN:TDYX.0.2019-08-034.
QIU Y, QI S H. Research status of non-invasive monitoring technology of cerebral oxygen saturation[J]. Mod Med J, 2019, 47(8): 1045-1048. DOI: CNKI:SUN:TDYX.0.2019-08-034.
刘强, 韩如泉. 脑氧饱和度监测方法及其应用进展[J]. 国际麻醉学与复苏杂志, 2018, 39(3): 234-238. DOI: 10.3760/cma.j.issn.1673-4378.2018.03.009.
LIU Q, HAN R Q, Advanes in cerebral oxygensaturation monitoring and its application[J]. Int J Anesthesiol Resusc, 2018, 39(3): 234-238. DOI: 10.3760/cma.j.issn.1673-4378.2018.03.009.
BICKLER P E, FEINER J R, ROLLINS M D. Factors affecting the performance of 5 cerebral oximeters during hypoxia in healthy volunteers[J]. Anesth Analg, 2013, 117(4): 813-823. DOI: 10.1213/ANE.0b013e318297d763.
SHAH N, TRIVEDI N K, CLACK S L, et al . Impact of hypoxemia on the performance of cerebral oximeter in volunteer subjects[J ] . J Neurosurg Anesthesiol, 2000, 12(3): 201-209. DOI: 10.1097/00008506-200007000-00001.
MURKIN J M, ADAMS S J, NOVICK R J, et al . Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study[J ] . Anesth Analg, 2007, 104(1): 51-58. DOI: 10.1213/01.ane.0000246814.29362.f4.
LA COUR A, GREISEN G, HYTTEL-SORENSEN S. In vivo validation of cerebral near-infrared spectroscopy: a review[J]. Neurophotonics, 2018, 5(4): 040901. DOI: 10.1117/1.nph.5.4.040901.
BENNI P B, MACLEOD D, IKEDA K, et al . A validation method for near-infrared spectroscopy based tissue oximeters for cerebral and somatic tissue oxygen saturation measurements[J ] . J Clin Monit Comput, 2018, 32(2): 269-284. DOI: 10.1007/s10877-017-0015-1.
BLAND J M, ALTMAN D G. Applying the right statistics: analyses of measurement studies[J]. Ultrasound Obstet Gynecol, 2003, 22(1): 85-93. DOI: 10.1002/uog.122.
陆梦洁, 刘玉秀, 缪华章, 等. Bland-Altman一致性评价的样本含量估计[J]. 中国卫生统计, 2015, 32(3): 537-539.
LU M J, LIU Y X, MIAO H Z, et al . Estimation of sample size in Bland-Altman consistency evaluation[J ] . Chin J Heal Stat, 2015, 32(3): 537-539.
屈仁超. 光电传感器响应时间的准确测量方法[J]. 电子质量, 2014(4): 64-66. DOI: 10.3969/j.issn.1003-0107.2014.04.016.
QU R C. Accurate measurement of photoelectric sensor response time[J]. Electron Qual, 2014(4): 64-66. DOI: 10.3969/j.issn.1003-0107.2014.04.016.
TOMLIN K L, NEITENBACH A M, BORG U. Detection of critical cerebral desaturation thresholds by three regional oximeters during hypoxia: a pilot study in healthy volunteers[J]. BMC Anesthesiol, 2017, 17(1): 6. DOI: 10.1186/s12871-016-0298-7.
GERKE O. Reporting standards for a bland-Altman agreement analysis: a review of methodological reviews[J]. Diagnostics (Basel), 2020, 10(5): E334. DOI: 10.3390/diagnostics10050334.
向小燕. 浅谈高原环境对医疗设备的影响[J]. 西藏医药, 2020, 41(4): 150-151. DOI: CNKI:SUN:YLZB.0.2003-10-036.
XIANG X Y. Effects of medical equipment in plateau environment[J]. Tibet Med, 2020, 41(4): 150-151. DOI: CNKI:SUN:YLZB.0.2003-10-036.
阿祥仁, 赵生秀, 赵文君. 我国高原地区POCT临床应用的现况[J]. 中华检验医学杂志, 2014, 37(11): 812-814. DOI: 10.3760/cma.j.issn.1009-9158.2014.11.004.
A X R, ZHAO S X, ZHAO W J. Clinical application of point-of-care testing in plateau region of China[J]. Chin J Lab Med, 2014, 37(11): 812-814. DOI: 10.3760/cma.j.issn.1009-9158.2014.11.004.
KUSAKA T, ISOBE K, NAGANO K, et al . Quantification of cerebral oxygenation by full-spectrum near-infrared spectroscopy using a two-point method[J ] . Comp Biochem Physiol A Mol Integr Physiol, 2002, 132(1): 121-132. DOI: 10.1016/s1095-6433(01)00538-4.
AINSLIE P N, OGOH S. Regulation of cerebral blood flow in mammals during chronic hypoxia: a matter of balance[J]. Exp Physiol, 2010, 95(2): 251-262. DOI: 10.1113/expphysiol.2008.045575.
于布为, 王国林, 邓小明, 等. 临床麻醉监测指南中国-麻醉学指南与专家共识[M]. 2017版. 北京: 人民卫生出版社, 2017, 168-169.
YU B W, WANG G L, DENG X M, et al . Clinical anesthesia monitoring guidelines: chinese anesthesiology guidelines and expert consensus[M ] . 2017. Beijing: People's Medical Publishing House, 2017, 168-169.