宋敏, 范娅涵, 王世春, 等. 间接ELISA测定人血IgM型抗-A抗体方法的建立及应用[J]. 第三军医大学学报, 2018,40(7):556-560.
Min SONG, Yahan FAN, Shichun WANG, et al. Establishment and application of an indirect ELISA for determining IgM anti-A antibodies in human plasma[J]. Journal of Third Military Medical University, 2018, 40(7): 556-560.
宋敏, 范娅涵, 王世春, 等. 间接ELISA测定人血IgM型抗-A抗体方法的建立及应用[J]. 第三军医大学学报, 2018,40(7):556-560. DOI: 10.16016/j.1000-5404.201711078.
Min SONG, Yahan FAN, Shichun WANG, et al. Establishment and application of an indirect ELISA for determining IgM anti-A antibodies in human plasma[J]. Journal of Third Military Medical University, 2018, 40(7): 556-560. DOI: 10.16016/j.1000-5404.201711078.
目的
2
建立间接酶联免疫吸附实验(enzyme linked immunosorbent assay,ELISA)检测方法,快速定量检测人血IgM抗-A抗体。
方法
2
通过制备A型红细胞膜蛋白作为捕获抗原,采用棋盘滴定实验筛选出最佳抗原抗体工作浓度,建立人血IgM型抗-A抗体定量检测方法。
结果
2
所构建的间接ELISA定量检测方法检测下限为0.02 μg/mL;批内变异系数及批间变异系数均小于10%;通过ROC曲线分析,当临界
D
(450)为0.259时,ROC曲线下面积最大,为0.947,检测灵敏度为87.7%,特
异性为100%。采用本研究自建的ELISA检测57例B型血浆标本的IgM型抗-A抗体浓度为(1.71±1.95)μg/mL,定量结果与经典半定量效价评分呈正相关(
r
=0.71,
P
<
0.01)。
结论
2
成功建立了人血IgM型抗-A抗体定量检测方法。
Objective
2
To develop an indirect enzyme linked immunosorbent assay (ELISA) for rapid and quantitative detection of IgM anti-A antibodies in human plasma.
Methods
2
An in-house indirect ELISA was developed using the ghost protein of group A erythrocyte as coating antigen. The optimal combination of antigens and antibodies was selected by checkerboard titration. The methodological performance was evaluated
and the plasma concentrations of IgM anti-A antibodies of clinical samples were determined.
Results
2
The detection limit of the in-house indirect ELISA was 0.02 μg/mL. The coefficient of variation for inner-batch and inter-batch were both less than 10%. According to the receiver operating characteristic (ROC) curve analysis
the largest area under the curve accounted for 0.947 when the cut-off value at
D
(450) was 0.259
and the detection sensitivity and specificity were 87.7% and 100% respectively. The plasma level of IgM anti-A antibodies in 57 samples of type B plasma was 1.71±1.95 μg/mL. The quantitative results were positively correlated to the titer scores from the classical semi-quantitative detection of IgM anti-A antibodies (
r
=0.71
P
<
0.01).
Conclusion
2
An indirect ELISA is developed successfully for quantitative detection of IgM anti-A antibodies.
COOLING L. ABO and platelet transfusion therapy[J]. Immunohematology, 2007, 23 (1): 20-33.
FUNG M K, DOWNES K A, SHULMAN I A. Transfusion of platelets containing ABO-incompatible plasma: a survey of 3 156 North American laboratories[J]. Arch Pathol Lab Med, 2007, 131 (6): 909-916. DOI: 10.1043/1543-2165(2007) 131[909:TOPCAP]2.0.CO; 2.
HARRIS S B, JOSEPHSON C D, KOST C B, et al . Nonfatal intravascular hemolysis ina pediatric patient after transfusion of a platelet unit with high-titer anti-A[J ] . Transfusion, 2007, 47 (8): 1412-1417. DOI: 10.1111/j.1537-2995.2007.01283.x.
KOBAYASHI T, SAITO K. A series of surveys on assay for anti-A/B antibody by Japanese ABO-incompatible Transplantation Committee[J]. Xenotransplantation, 2006, 13(2): 136-140. DOI: 10.1111/j.1399-3089. 2006.00296.x.
PESCIOTTA E N, LAM H S, KOSSENKOV A, et al . In-depth, label-free analysis of the erythrocyte cytoplasmic proteome in Diamond Blackfan anemia identifies a unique inflammatory signature[J ] . PLoS ONE, 2015, 10 (10): e140036. DOI: 10.1371/journal.pone.0140036.
CROWTHER J R. The ELISA guidebook[M]. 2nd ed. New York: Humana Press, 2009. 566.
LEE M F, LIN T M, LIU S W, et al . A rapid method of detecting autoantibody against FcepsilonRIalpha for chronic spontaneous urticaria[J ] . PLoS ONE, 2014, 9(10): e109565. DOI: 10.1371/journal.pone.0109565.
DENG Q, QIU M, WANG Y, et al . A sensitive and validated immunomagnetic-bead based enzyme-linked immunosorbent assay for analyzing total T-2 (free and modified) toxins in shrimp tissues[J ] . Ecotoxicol Environ Saf, 2017, 142: 441-447. DOI: 10.1016/j.ecoenv. 2017.04.037.
兰炯采, 吴涛, 张国鹏, 等.用ELISA双抗夹心法检测sHLA-Ⅰ在不同贮存成分血中的浓度[J].中国输血杂志, 2003, 16(6): 402-404. DOI: 10.3969/j.issn.1004-549X.2003.06.007.
LAN J C, WU T, ZHANG G P, et al . Detection of soluble HLA-��concentration in blood components at different storage by ELISA double antigen sandwich assay[J ] . Chin J Blood Transfus, 2003, 16 (6): 402-404.
中华人民共和国卫生部医政司.全国临床检验操作规程[M].3版.南京:东南大学出版社, 2006: 791-798.
Department of Medical Administration of Ministry of Health of People��s Republic of China. National guide to clinical laboratory prodedures[M]. 3 ed. Nanjing: Southeast University Press, 2006: 791-798.
NORRIS J M, DIKEMANJ, GREENE D, et al . Managing change[J ] . Transfusion, 2009, 49 (11 Pt 2): 2524-2535. DOI: 10.1111/j.1537-2995. 2009.02470.x.
MORIOKA D, TOGO S, KUMAMOTO T, et al . Six consecutive cases of successful adult ABO-incompatible living donor liver transplantation: a proposal for grading the severity of antibody-mediated rejection[J ] . Transplantation, 2008, 85(2): 171-178. DOI: 10.1097/TP.0b013e31815e9672.
BELLAMY C O, HERRIOT M M, HARRISON D J, et al . C4d immunopositivity is uncommon in ABO-compatible liver allografts, but correlates partially with lymphocytotoxic antibody status[J ] . Histopathology, 2007, 50 (6): 739-749. DOI: 10.1111/j.1365-2559.2007.02677.x.
YURUGI K, KIMURA S, ASHIHARA E, et al . Rapid and accurate measurement of anti-A/B IgG antibody in ABO-unmatched living donor liver transplantation by surface plasmon resonance[J ] . Transfus Med, 2007, 17 (2): 97-106. DOI: 10.1111/j.1365-3148.2007.00737.x.
SATOH A, KAWAGISHI N, MINEGISHI M, et al . Development of a novel ELISA for detection of anti-A and anti-B antibodies in recipients of ABO-incompatible living donor liver grafts[J ] . Tohoku J Exp Med, 2007, 211(4): 359-367. DOI: 10.1620/tjem.211.359.
张敏, 芦慧霞, 辛琪. IgG抗-A/BELISA检测方法的初探[J].中国输血杂志, 2011, 24(5): 414-415. DOI: 10.13303/j.cjbt.issn.1004-549x.2011.05.018.
ZHANG M, LU H X, XIN Q. Primary investigation of ELISA for determination of IgG anti-A/B[J]. Chin J Blood Transfus, 2011, 24(5): 414-415. DOI: 10.13303/j.cjbt.issn.1004-549x.2011.05.018.
卢新奇, 张志峰.定量测定ABO血型IgG抗体新方法的研究[J].医学检验与临床, 2012, 23(1): 72-73. DOI: 10.3969/j.issn.1673-5013.2012.01.031.
LU X Q, ZHANG Z F. A new method for quantitative determination of the ABO blood group IgG antibody[J]. Med Lab Sci Clin, 2012, 23 (1): 72-73. DOI: 10.3969/j.issn.1673-5013.2012.01.031.
SPALTER S H, KAVERI S V, BONNIN E, et al . Normal human serum contains natural antibodies reactive with autologous ABO blood group antigens[J ] . Blood, 1999, 93 (12): 4418-4424.
0
浏览量
0
下载量
0
CSCD
关联资源
相关文章
相关作者
相关机构
京公网安备11010802024621