刘会佳,张瑜.肺实性结节半自动体积测量提高观察者间Lung-RADS评分一致性[J].重庆医科大学学报,2023,48(5):544-547.
Liu Huijia,Zhang Yu.Semi-automatic volumetry of solid pulmonary nodules improves the interobserver consistency of Lung-RADS score[J].Journal of Chongqing Medical University,2023,48(5):544-547.
刘会佳,张瑜.肺实性结节半自动体积测量提高观察者间Lung-RADS评分一致性[J].重庆医科大学学报,2023,48(5):544-547. DOI: 10.13406/j.cnki.cyxb.003227.
Liu Huijia,Zhang Yu.Semi-automatic volumetry of solid pulmonary nodules improves the interobserver consistency of Lung-RADS score[J].Journal of Chongqing Medical University,2023,48(5):544-547. DOI: 10.13406/j.cnki.cyxb.003227.
目的
2
研究3D-Slicer软件实性肺结节体积测量对不同观察者在肺部影像报告数据系统(lung CT screening reporting and data system,Lung-RADS)分类一致性中的影响。
方法
2
纳入76例患者中的76个实性结节。由3位放射科医师分别采用手动和3D-Slicer软件半自动体积测量方法获得结节的直径与体积,并转化为相应的Lung-RADS评分,其中2分为阴性,3分及以上为阳性。采用同类相关系数(intraclass correlation coefficient,ICC)及Bland-Altman指数来评价观察者间直径与体积测量的一致性,Kappa分析评价观察者间Lung-RADS评分及阳性/阴性组间的一致性。
结果
2
ICC分析结果显示手动直径测量的一致性(0.994~0.996)明显低于半自动体积测量的一致性(0.997~0.998),同时Bland-Altman指数分析结果显示手动直径测量的偏倚高于半自动体积测量。采用半自动体积测量,能够比手动直径测量明显提高观察者间Lung-RADS评分及阳性/阴性之间的一致性(0.963~0.975及0.957~0.977 vs. 0.833~0.866及0.863~0.892)。
结论
2
3D-Slicer半自动体积测量能够提高实性肺结节观察者间大小测量的一致性,相应的Lung-RADS分类一致性也随之提高。
Objective
2
To study the effect of solid pulmonary nodules volume measurement by 3D-Slicer software on the interobserver consistency of lung CT screening reporting and data system(Lung-RADS) categories.
Methods
2
Seventy-six solid nodules from 76 patients were included in this study. The diameter and volume of nodules were obtained by three radiologists using manual and 3D-Slicer software semi-automatic volume measurement methods,respectively,and the raw data was converted to the corresponding Lung-RADS score,of which 2 points was negative and 3 points or above was positive. Intraclass correlation coefficient(ICC) and Bland-Altman index were used to evaluate the interobserver consistency of diameter and volume measurements. Kappa analysis was used to evaluate the interobserver consistency of Lung-RADS score and the intergroup consistency of positive/negative.
Results
2
ICC analysis showed that the consistency of manual diameter measurement(0.994-0.996) was significantly lower than that of semi-automatic volume measurement(0.997-0.998),and Bland-Altman index analysis showed that the bias of manual diameter measurement was higher than that of semi-automatic volume measurement. Compared with manual diameter measurement,the use of semi-automatic volume measurement can significantly improve the consistency of Lung-RADS score and positive/negative classifications(0.963-0.975 and 0.957-0.977 vs. 0.833-0.866 and 0.863-0.892).
Conclusion
2
3D-Slicer semi-automatic volume measurement can improve the interobserver consistency of the size measurement of solid pulmonary nodules, as well as the corresponding Lung-RADS categories.
Bray F , Ferlay J , Soerjomataram I , et al . Global cancer statistics 2018:GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries [J ] . CA A Cancer J Clin , 2018 , 68 ( 6 ): 394 - 424 .
American College of Radiology Web Site . Lung CT screening reporting and data system (lung-RADS) version 1.1 [EB/OL ] . ( 2019-05-31 ) [ 2022-03-01 ] . http://www.acr.org/Quality-Safety/Resources/LungRADS http://www.acr.org/Quality-Safety/Resources/LungRADS .
Wood DE , Kazerooni EA , Baum SL , et al . Lung cancer screening,version 3.2018,NCCN clinical practice guidelines in oncology [J ] . J Natl Compr Canc Netw , 2018 , 16 ( 4 ): 412 - 441 .
Gierada DS , Rydzak CE , Zei M , et al . Improved interobserver agreement on lung-RADS classification of solid nodules using semiautomated CT volumetry [J ] . Radiology , 2020 , 297 ( 3 ): 675 - 684 .
Cohen JG , Goo JM , Yoo RE , et al . Software performance in segmenting ground-glass and solid components of subsolid nodules in pulmonary adenocarcinomas [J ] . Eur Radiol , 2016 , 26 ( 12 ): 4465 - 4474 .
Revel MP , Bissery A , Bienvenu M , et al . Are two-dimensional CT measurements of small noncalcified pulmonary nodules reliable? [J ] . Radiology , 2004 , 231 ( 2 ): 453 - 458 .
Revel MP , Mannes I , Benzakoun J , et al . Subsolid lung nodule classification:a CT criterion for improving interobserver agreement [J ] . Radiology , 2018 , 286 ( 1 ): 316 - 325 .
Yip SSF , Parmar C , Blezek D , et al . Application of the 3D slicer chest imaging platform segmentation algorithm for large lung nodule delineation [J ] . PLoS One , 2017 , 12 ( 6 ): e0178944 .
Xu FY , Zhu WC , Shen Y , et al . Radiomic-based quantitative CT analysis of pure ground-glass nodules to predict the invasiveness of lung adenocarcinoma [J ] . Front Oncol , 2020 , 10 : 872 .
Heuvelmans MA , Walter JE , Vliegenthart R , et al . Disagreement of diameter and volume measurements for pulmonary nodule size estimation in CT lung cancer screening [J ] . Thorax , 2018 , 73 ( 8 ): 779 - 781 .
Han DW , Heuvelmans MA , Vliegenthart R , et al . Influence of lung nodule margin on volume- and diameter-based reader variability in CT lung cancer screening [J ] . Br J Radiol , 2018 , 91 ( 1090 ): 20170405 .