车凤莲,陈靓雯,赵婉等.减压沸腾清洗联合低温真空干燥技术在牙科手机再处理中的应用[J].局解手术学杂志,2026,35(07):638-641.
CHE Feng-lian,CHEN Liang-wen,ZHAO Wan,et al.Application of depressurized boiling cleaning combined with low-temperature vacuum drying technology in reprocessing of dental handpieces[J].JOURNAL OF REGIONAL ANATOMY AND OPERATIVE SURGERY,2026,35(07):638-641.
车凤莲,陈靓雯,赵婉等.减压沸腾清洗联合低温真空干燥技术在牙科手机再处理中的应用[J].局解手术学杂志,2026,35(07):638-641. DOI: 10.11659/jjssx.07E025081.
CHE Feng-lian,CHEN Liang-wen,ZHAO Wan,et al.Application of depressurized boiling cleaning combined with low-temperature vacuum drying technology in reprocessing of dental handpieces[J].JOURNAL OF REGIONAL ANATOMY AND OPERATIVE SURGERY,2026,35(07):638-641. DOI: 10.11659/jjssx.07E025081.
目的
2
对比不同牙科手机清洗与干燥技术联合应用的效果,确定最优再处理方案,保障器械处置质量与临床使用安全。
方法
2
选取2024年3月我院使用的牙科手机800件,分别采用喷淋清洗消毒+压力气枪干燥(喷淋+气枪组)、喷淋清洗消毒+低温真空干燥(喷淋+真空组)、减压沸腾清洗消毒+压力气枪干燥(沸腾+气枪组)、减压沸腾清洗消毒+低温真空干燥(沸腾+真空组)处理,每组200件。比较4组牙科手机的清洗效果、干燥效果、灭菌后油性湿包情况以及工作人员对4组处理方案的操作满意度。
结果
2
沸腾+气枪组与沸腾+真空组牙科手机目测合格率及蛋白残留定量检测合格率均高于喷淋+气枪组、喷淋+真空组(
P
<
0.05)。4组牙科手机清洗消毒后干燥合格率比较,差异无统计学意义(
P
>
0.05);喷淋+真空组、沸腾+真空组牙科手机清洗消毒后额外器械干燥合格率显著高于喷淋+气枪组、沸腾+气枪组(
P
<
0.05)。喷淋+真空组、沸腾+真空组牙科手机油性湿包发生率显著低于喷淋+气枪组、沸腾+气枪组(
P
<
0.05)。工作人员对沸腾+真空组的操作满意度最高(
P
<
0.01)。
结论
2
采用减压沸腾清洗消毒联合低温真空干燥技术处理牙科手机,既可有效保证清洗及灭菌质量,又可实现消毒供应中心高效处理复用医疗器械的工作目标,工作人员对此处理方案的操作满意度较高。
Objective
2
To compare the effects of combined application of different cleaning and drying technologies for dental handpieces, determine the optimal reprocessing scheme , and ensure the quality of instrument reprocessingt and clinical safety.
Methods
2
A total of 800 dental handpieces used in our hospital in March 2024 were selected and equally divided into four groups with 200 pieces per group: spray cleaning and disinfection combined with compressed air gun drying group (spray+air gun group), spray cleaning and disinfection combined with low-temperature vacuum drying group (spray+vacuum group), depressurized boiling cleaning and disinfection combined with compressed air gun drying group (boiling+air gun group), depressurized boiling cleaning and disinfection combined with low-temperature vacuum drying group (boiling+vacuum group). The four groups were compared in terms of cleaning effect, drying effect, incidence of oily wet packs after sterilization, and staff´s operational satisfaction with the four processing scheme.
Results
2
The qualified rates of visual inspection and quantitative detection of protein residues of dental handpieces in the boiling+air gun group and boiling+vacuum group were higher than those in the spray+air gun group and spray+vacuum group (
P
<
0.05). There was no statistically significant difference in the drying qualified rates of dental handpieces after cleaning and disinfection among the four groups (
P
>
0.05). The qualified rates of supplementary mechanical drying of dental handpieces after cleaning and disinfection in the spray+vacuum group and boilin
g+vacuum group were significantly higher than those in the spray+air gun group and boiling+air gun group (
P
<
0.05). The incidences of oily wet packs after sterilization in the spray+vacuum group and boiling+vacuum group were significantly lower than those in the spray+air gun group and boiling+air gun group (
P
<
0.05). The boiling+vacuum group obtained the highest staff´s operational satisfaction (
P
<
0.01).
Conclusion
2
Adopting depressurized boiling cleaning and disinfection combined with low-temperature vacuum drying technology for dental handpieces can effectively guarantee cleaning and sterilization qualities, enable the goal of efficient reprocessing of reusable medical devices in central sterile supply department, and achieve the high staff´s operational satisfaction with the processing protocol.
Deasy EC , Scott TA , Swan JS , et al . Effective cleaning and decontamination of the internal air and water channels,heads and head-gears of multiple contra-angle dental handpieces using an enzymatic detergent and automated washer-disinfection in a dental hospital setting [J]. J Hosp Infect , 2022 , 128 : 80 - 88 . doi: 10.1016/j.jhin.2022.07.019 http://dx.doi.org/10.1016/j.jhin.2022.07.019 .
Coleman DC , Deasy EC , Swan JS , et al . Direct evidence for effective cleaning and decontamination of the internal air and water channels,heads and head-gears of multiple contra-angle dental handpieces using an enzymatic detergent and automated washer-disinfection in a dental hospital setting [J]. J Hosp Infect , 2023 , 135 : 204 - 205 . doi: 10.1016/j.jhin.2023.03.001 http://dx.doi.org/10.1016/j.jhin.2023.03.001 .
Schalli M , Kogler B , Miorini T , et al . High-speed dental instruments: an investigation of protein-contaminated dental handpieces with the bicinchoninic acid assay in dental offices in Styria,Austria [J]. Int J Environ Res Public Health , 2023 , 20 ( 3 ): 1670 . doi: 10.3390/ijerph20031670 http://dx.doi.org/10.3390/ijerph20031670 .
张文清 , 郭燕娥 , 马如梦 . 2种不同的清洗方法在眼科手术器械清洗中的效果评价 [J]. 中国医疗器械信息 , 2023 , 29 ( 24 ): 142 - 144 . doi: 10.3969/j.issn.1006-6586.2023.24.047 http://dx.doi.org/10.3969/j.issn.1006-6586.2023.24.047 .
李燕 , 赵强 , 王春琴 . 表面蛋白质残留测定法在复用医疗器械清洗效果中评价作用 [J]. 中国医疗器械信息 , 2023 , 29 ( 5 ): 59 - 62 . doi: 10.3969/j.issn.1006-6586.2023.05.017 http://dx.doi.org/10.3969/j.issn.1006-6586.2023.05.017 .
Cobbold A , Lord S . A critical guideline appraisal overview: Choice Framework for local Policy and Procedures (CFPP) 01-01: management and decontamination of surgical instruments (medical devices) used in acute care.Part A: the formulation of local policy and choices manual Version 1.0:England [J]. J Perioper Pract , 2012 , 22 ( 12 ): 376 - 379 . doi: 10.1177/175045891602201201 http://dx.doi.org/10.1177/175045891602201201 .
应艳艳 , 项蜜程 , 陈慧伦 . 气枪二次吹干内腔管路对牙科手机清洗消毒效果的影响 [J]. 中国基层医药 , 2025 , 32 ( 6 ): 922 - 925 . doi: 10.3760/cma.j.cn341190-20241031-01421 http://dx.doi.org/10.3760/cma.j.cn341190-20241031-01421 .
刘钕镝 , 许小红 , 杨小梅 . 牙科手机干燥方法对油包发生率的影响 [J]. 医疗装备 , 2023 , 36 ( 23 ): 17 - 18 . doi: 10.3969/j.issn.1002-2376.2023.23.005 http://dx.doi.org/10.3969/j.issn.1002-2376.2023.23.005 .
王思园 . 不同注油方式对牙科手机灭菌后发生油包的影响 [J]. 临床口腔医学杂志 , 2024 , 40 ( 7 ): 429 - 430 . doi: 10.3969/j.issn.1003-1634.2024.07.011 http://dx.doi.org/10.3969/j.issn.1003-1634.2024.07.011 .
刘漫丽 , 谢婉青 , 宋思平 , 等 . 脉动真空清洗消毒机对牙科手机清洗效果的影响 [J]. 四川大学学报(医学版) , 2023 , 54 ( 4 ): 788 - 791 . doi: 10.12182/20230760201 http://dx.doi.org/10.12182/20230760201 .
黄蓉 , 江丽 , 彭佳欣 , 等 . 牙科手机清洗质量管理的最佳证据总结 [J]. 中华医院感染学杂志 , 2025 , 35 ( 8 ): 1256 - 1260 . doi: 10.11816/cn.ni.2025-241572 http://dx.doi.org/10.11816/cn.ni.2025-241572 .
郝宝莲 , 卢倩 , 侯锐 , 等 . 新型冠状病毒肺炎疫情期口腔医院防护标准(八)——口腔诊疗器械处理 [J]. 实用口腔医学杂志 , 2020 , 36 ( 2 ): 188 - 191 . doi: 10.3969/j.issn.1001-3733.2020.02.009 http://dx.doi.org/10.3969/j.issn.1001-3733.2020.02.009 .
车凤莲 , 金敏智 , 季侃雯 , 等 . 精密导线器械盘绕支架的设计与应用 [J]. 局解手术学杂志 , 2022 , 31 ( 1 ): 74 - 76 . doi: 10.11659/jjssx.08E021041 http://dx.doi.org/10.11659/jjssx.08E021041 .
杜芳 , 赵经纶 , 吴秉伦 , 等 . 深圳市口腔科门诊患者满意度及其影响因素研究 [J]. 中国卫生经济 , 2022 , 41 ( 12 ): 71 - 75 .
林洁 , 刘帆 . 基于多酶清洗剂控制的口腔专科器械清洗消毒模式的探索 [J]. 国际口腔医学杂志 , 2022 , 49 ( 3 ): 324 - 327 . doi: 10.7518/gjkq.2022052 http://dx.doi.org/10.7518/gjkq.2022052 .
0
浏览量
0
下载量
0
CSCD
关联资源
相关文章
相关作者
相关机构
京公网安备11010802024621