DONG XIAO-JING, HU LI-NA, SUN CAI-XIN, et al. Dosage and safety evaluation of energy controllable steep pulse in treatment of implanted rabbit tumors. [J]. Journal of army medical university, 2007, 29(3): 213-215.
DONG XIAO-JING, HU LI-NA, SUN CAI-XIN, et al. Dosage and safety evaluation of energy controllable steep pulse in treatment of implanted rabbit tumors. [J]. Journal of army medical university, 2007, 29(3): 213-215.DOI:
Objectives To explore the optimal dosage and evaluate the safety of energy controllable steep pulse (ECSP) in the treatment of implanted rabbit tumor. Methods The models of VX
2
breast carcinoma were established in 10 New England rabbits
and then the animals were given ECSP treatment under orthogonality designed dosages of ECSP. The volumes of necrosis tumors were measured by TTC staining. The best dosage was analyzed by intuitionistic analysis. The models of VX
2
liver carcinoma were established in 20 New England rabbits. The normal tissues surrounding the tumor were extracted
the DNA fragmentation
analysis and comet assay were carried out. Results Voltage took the priority among the 4 parameters
and for best dosage they were 150 V for voltage
100 Hz for frequency
0.01μm for pulse width
and 15 min for time if the interaction of these 4 factors was not considered. The electrophoresis of DNA in the peripheral normal tissue was an integrity straining after ECSP treatment. No comet tailing was observed
indicating no DNA lesions after ECSP treatment. Conclusion The best dosage is 150 V for voltage
100 Hz for frequency
0.01 μm for pulse width and treating for 15 min. The ECSP is verified to be safe.
关键词
能量可控陡脉冲VX2肿瘤治疗剂量安全性
Keywords
energy controllable steep pulseVX2 tumordosagesafety