GUI LI, TIAN HONG, ZHENG JIAN, et al. Simultaneously determining 4 amino acid neurotransmitters in mice brain by high-performance liquid chromatography with fluorescence detector. [J]. Journal of army medical university, 2009, 31(8): 675-678.
GUI LI, TIAN HONG, ZHENG JIAN, et al. Simultaneously determining 4 amino acid neurotransmitters in mice brain by high-performance liquid chromatography with fluorescence detector. [J]. Journal of army medical university, 2009, 31(8): 675-678.DOI:
Objective To establish a stable method of HPLC-FL to determine multi-amino acids in the mice microdialysate. Methods The sample was collected by microdialysis. Four amino acid neurotransmitters
Asp
Glu
Gly and GABA in the microdialysate were determined by reverse phase HPLC with fluorescence detection after o-Phthaldehyde (OPA) pre-column derivatization. Results Excitatory amino acids
Asp and Glu
and inhibitory amino acids
Gly and gamma-aminobutyric acid (GABA) were totally separated and determined within 18 min. A good linear response was obtained for the 4 standards over the range of 0.625 to 5 μmol/L with correlation coefficients. Conclusion Our method is quick
reliable and sensitive to measure the amino acid neurotransmitters in mice brain.
关键词
高效液相色谱柱前衍生微透析氨基酸神经递质
Keywords
high performance liquid chromatographyprecolumn derivatizationmicrodialysisamino acidneurotransmitters