目的 评估Pentacam与SL-OCT测量原发性闭角型青光眼(primary angle-closure glaucoma,PACG)应用激光周边虹膜切开术(laser peripheral iridoplasty,LPI)手术前后中央前房深度(center anterior chamber depth
CACD)的差异性、相关性和一致性。 方法 在我院眼科门诊筛查PACG患者中符合行LPI术的病例共58例(58眼),应用眼前节分析系统Pentacam和裂隙灯光学相干断层成像术(slit-lamp-adapted optical coherence tomography
SL-OCT)分别对PACG患者进行CACD测量,分别记录LPI术前,术后1 d、1周、1个月参数。 结果 Pentacam测得PACG患者LPI术前,术后1 d、1周及1个月的CACD分别为(1.60±0.31)、(1.65±0.28)、(1.62±0.30)、(1.61±0.31)mm。SL-OCT测量LPI术前,术后1 d、1周及1个月的CACD分别为(1.70±0.30)、(1.73±0.30)、(1.71±0.31)、(1.72±0.31)mm。2种测量方式LPI术前与术后各时间比较,差异均无统计学意义(
P
0.05);而2种测量方式之间相比较,术前及术后相同时间点差异均具有统计学意义(
P
0.01)。Pentacam与SL-OCT测量PACG患者LPI术前,术后1 d、1周及1个月的CACD值均存在显著相关关系(r=0.929,
P
0.001;r=0.962,
P
0.001;r=0.961,
P
0.001;r=0.951,
P
0.001)。通过Bland-Altman分析方法,Pentacam与SL-OCT测量LPI术前,术后1 d、1周及1个月CACD差异的95%置信界限分别为-0.33~0.12、-0.29~0.08、-0.25~0.08、-0.26~0.07 mm。 结论 Pentacam与SL-OCT测量PACG患者LPI术后CACD具有高度一致性和相关性,二者均可作为青光眼临床筛查的辅助检查方式,为患者提供了有效的无创且可量化的便捷检查。
Objective
To evaluate the central anterior chamber depth (CACD) before and after laser peripheral iridoplasty (LPI) in patients with primary angle-closure glaucoma (PACG)
and assess the difference
relevance and consistency between Pentacam and slit-lamp-adapted optical coherence tomography (SL-OCT) in measuring CACD. Methods A total of 58 PACG patients (58 eyes) were enrolled in this study. Pentacam anterior segment analysis system and Heidelberg SL-OCT were used separately before and in 1 d
1 week
and 1 month after LPI to measure the CACD in all patients. Results The mean CACD values of PACG before and in 1 d
1 week
and 1 month after LPI measured by Pentacam were 1.60±0.31
1.65±0.28
1.62±0.30
and 1.61±0.31mm
respectively
while those measured by SL-OCT were 1.70±0.30
1.73±0.30
1.71±0.31
and 1.72±0.31 mm
respectively. Using the 2 methods
no statistical differences was observed in the CACD values before and after LPI in PACG (
P
0.05). Significant differences of CACD values
however
existed between the 2 methods at all time points (
P
0.01). Good correlations were found between Pentacam and SL-OCT before and in 1 d
1 week
and 1 month after LPI (r=0.929
P
0.001; r=0.962
P
0.001; r=0.961
P
0.001; and r=0.951
P
0.001
respectively). By Bland-Altman analysis
the 95% limits of differences between Pentacam and SL-OCT before and in 1 d
1 week
and 1 month after LPI were -0.33 to 0.12
-0.29 to 0.08
-0.25 to 0.08
and -0.26 to 0.07 mm
respectively. Conclusion In PACG
Pentacam and SL-OCT
of good consistency and correlation in measuring CACD
can be both used for glaucoma screening
and provide an effective non-invasive and quantitative way for patients.