郭乃铭,张鹤,李杰,黄博,郑文杰,周跃.O型臂导航系统与X线正侧位片引导下牛腰椎椎弓根打孔的对比研究[J].陆军军医大学学报(原第三军医大学学报),2013,35(04):277-279.
Guo Naiming,Zhang He,Li Jie,Huang Bo,Zheng Wenjie,Zhou Yue.O-arm based navigation vs X-ray fluoroscopy in accuracy of pedicle drilling in bovine lumber spine specimens[J].J Amry Med Univ (J Third Mil Med Univ),2013,35(04):277-279.
郭乃铭,张鹤,李杰,黄博,郑文杰,周跃.O型臂导航系统与X线正侧位片引导下牛腰椎椎弓根打孔的对比研究[J].陆军军医大学学报(原第三军医大学学报),2013,35(04):277-279. DOI:
Guo Naiming,Zhang He,Li Jie,Huang Bo,Zheng Wenjie,Zhou Yue.O-arm based navigation vs X-ray fluoroscopy in accuracy of pedicle drilling in bovine lumber spine specimens[J].J Amry Med Univ (J Third Mil Med Univ),2013,35(04):277-279. DOI:
目的 对比研究O型臂导航系统与X线正侧位片引导下牛腰椎椎弓根打孔的精准性。 方法 对15具牛脊骨4~5个腰椎节段两侧椎弓根进行打孔,在每具标本一侧节段应用O型臂导航系统引导椎弓根打孔即导航组,另一侧相应节段应用X线正侧位片引导椎弓根打孔即放射组,两组各完成椎弓根打孔61次,共置入克氏针122枚。术中应用O型臂3D扫描模式进行扫描及三维重建,术后进行经克氏针水平的CT扫描,观察克氏针置入的精准度。 结果 术后CT图像显示,放射组克氏针置入Ⅰ级(完全置入椎弓根内)28枚(45.9%),Ⅱ级(穿破侧壁但未完全置于椎弓根外)24枚(39.3%),Ⅲ级(完全置于椎弓根外)9枚(14.8%)。导航组克氏针置入Ⅰ级46枚(75.4%),Ⅱ级15枚(24.6%),无Ⅲ级克氏针置入情况。两组差异显著(P0.05)。术后O型臂图像评价克氏针置入精准度结果与术后CT评价结果差异不显著(P0.05)。放射组打孔时间平均6.5 min,导航组打孔时间平均3.9 min,差异显著(P0.05)。 结论 应用O型臂导航系统辅助,能提高椎弓根打孔的精准性、缩短手术时间。
Objective To evaluate the feasibility and accuracy of bovine lumber spine pedicle drilling assisted by X-ray fluoroscopy and O-arm based navigation system. Methods Sixty-one holes were drilled assisted by X-ray fluoroscopy at one side of pedicles in 4 to 5 lumbar segments from 15 bovine lumber spine specimens
and another 61 holes were drilled assisted by O-arm based navigation system at the other side. Then 122 K-wires were put into the holes in all. K-wire positions were evaluated by O-arm 3 dimensional reconstruction after intraoperative cone beam computed tomography scanning (CBCT) and postoperative CT scanning. Results Postoperative CT scanning revealed that there were 28(45.9%) K-wires in Grade Ⅰ (perfect within the pedicles)
24(39.3%) Grade Ⅱ (breaching the pedicle wall but not completely outside the pedicle boundaries)
and 9(14.8%) Grade Ⅲ (completely outside the pedicle boundaries) in fluoroscopy group
and 46(75.4%) in Grade Ⅰ
15(24.6%) in Grade Ⅱ
and none in Grade Ⅲ in O-arm group. The 2 groups had significantly difference in accuracy (P0.05). There was no significant difference in the accuracy between the assessment of intraoperative O-arm scanning and postoperative CT scanning (P0.05). Average hole drilling time was 6.5 min (ranging from 5.2 to 8.1 min) in fluoroscopy group and 3.9 min (ranging from 2.8 to 5.2 min) in O-arm group (P0.05). Conclusion O-arm based navigation system increases the accuracy of pedicle drilling and reduces the surgical time significantly.