目的 通过对国人腰骶椎解剖结构数值的统计分析,设计一种新型腰骶椎前路钢板,并进行生物力学测试。 方法 对350例正常成人腰骶椎正位侧位X线片腰骶角、L
5
S
1
椎体及椎间盘高度进行测量,得出正常国人腰骶椎结构的大致数值范围,再依据此数值设计出改进型腰骶椎前路安全钢板各项参数指标,并将钢板样品在4 种不同生理运动状况下进行脊柱标本的生物力学数据测试。 结果 腰骶角角度范围为125°~135°,L
5
椎体前缘高度(27.06±2.30)mm,S
1
椎体前缘高度(22.3±2.8)mm,L
5
S
1
椎间盘高度(8.2±0.3)mm。改进型钢板固定后的屈曲位移[前屈(4.706±0.012)mm、后伸(4.549±0.298)mm、侧屈(5.412±0.384)mm
]
及压缩刚度[400 N下刚度(105.8±7.6)N/mm
]
比相同条件下PACH固定后[前屈(5.241±0.113)mm、后伸(5.662±0.246)mm、侧屈(6.767±0.017)mm、400 N下刚度(87.4±4.2)N/mm
]
差异具有统计学意义(P0.05)。 结论 研制的改进型腰骶椎前路钢板不仅在形态上完全适应国人腰骶椎解剖结构,且具有良好的生物力学稳定性能,适合腰骶椎前路脊柱重建的需要。
Objective To design a new lumbosacral anterior plate by statistical analysis for the numerical value of the anatomic structure in Chinese. Methods The numerical value of lumbosacral angle and height of lumbosacral vertebrae
and discus intervertebralis were measured in 350 orthotropia and lateral view Xray pictures of lumbosacral vertebrae of adult normal individuals. According to the acquired amplitude of lumbosacral vertebrae in Chinese
we designed the modified lumbosacral anterior p
late
and then tested the plate in 4 different states of working condition on 8 fresh spinal columns to assay biomechanical data
and the results were compared with those of lumbosacral anterior secure plate (PACH). Results The amplitude of lumbosacral angle were 125 to 135
the heights of anterior border L
5
and S
1
vertebrae were (27.06±2.30) and (22.3±2.8)mm
the height of L
5
S
1
discus intervertebralis was (8.2±0.3)mm. The flex bias and the compression rigidity had significant deviation for the vertebral column fixed the modified plate [flexation (4.706±0.012)mm
extension (4.549±0.298)mm
lateral bending (5.412±0.384)mm
average value of rigidity in 400 N (105.8±7.6)N/mm
]
and PACH[flexation (5.241±0.113) mm
extension (5.662±0.246)mm
lateral bending (6.767±0.017)mm
and average value of rigidity (87.4±4.2)N/mm
]
respectively
with statistical significance among them (P0.05). Conclusion The modified lumbosacral anterior plate system accommodates the anatomic structure of lumbosacral vertebrae in Chinese
possesses conspicuous stability
and is suitable for reconstruction of lumbosacral vertebrae in indicated patients.