ZHU Zhi-ming,ZHOU Mo-ling,LIU Jin-kang,XIONG Zeng,ZHOU Hui,CHEN Wei.Optimization and standardization of the technical parameter of multi-slice spiral CT pulmonary perfusion imaging[J].J Amry Med Univ (J Third Mil Med Univ),2008,30(20):1893-1896.
ZHU Zhi-ming,ZHOU Mo-ling,LIU Jin-kang,XIONG Zeng,ZHOU Hui,CHEN Wei.Optimization and standardization of the technical parameter of multi-slice spiral CT pulmonary perfusion imaging[J].J Amry Med Univ (J Third Mil Med Univ),2008,30(20):1893-1896.DOI:
Objective To optimize and standardize the technical parameter of multi-slice spiral CT pulmonary perfusion imaging
then to evaluate the effectiveness and stability of this scan protocol. Methods First
to determine the scan protocol according to the results of preliminary experiment and literature review
then based on the previous study
we studied in a larger sample size. We scanned 65 cases of lung single nodule according to the scan protocol of multi-slice spiral CT pulmonary perfusion imaging
then analyzed the main factors concerning the imaging quality
evaluated the effectiveness of scan protocol by determining the ROC curve
diagnostic efficacy and odds ratio
as well as the stability of scan protocol by consistency analysis. Results Forty-seven cases showed optimal images
resulting in correct diagnosis proved pathologically. Nine cases were made indefinite diagnosis. Nine cases were not up to the image quality for definite diagnosis mainly because of the patients’ incompliant respiration. The total sensitivity
specificity
accuracy
likelihood ratio and the area under ROC curve for scan protocol were 76.9%
83.3%
82%
4.620
0.280
0.840 respectively and Kappa value was 0.894. The diagnostic efficacy of multi-slice spiral CT pulmonary perfusion was significantly higher than that by CT plain and enhanced scan. Conclusion The optimized and standardized scan protocol of multi-slice spiral CT pulmonary perfusion imaging is effective in pathologic diagnosis and of well stability