Individualized 3D printing guide plates-assisted surgical correction for severe kyphosis deformity
Clinical Medicine|更新时间:2026-01-09
|
Individualized 3D printing guide plates-assisted surgical correction for severe kyphosis deformity
Journal of Army Medical UniversityVol. 46, Issue 21, Pages: 2443-2450(2024)
作者机构:
1.510000 广州,华南理工大学医学院
2.510000 广州,广州市第一人民医院骨科
3.510000 广州,中国人民解放军南部战区总医院康复医学科
4.400038 重庆,陆军军医大学(第三军医大学)生物医学工程与影像医学系数字医学教研室
作者简介:
基金信息:
the National Key Research and Development Program of China(2019YFE0126300);the National Key Research and Development Program of China(2021YFB3800800);the General Program of Natural Science Foundation of Guangdong Province(2021A1515012564);the General Program of Natural Science Foundation of Guangdong Province(2023A1515010557);the Key Research and Development Project of Guangzhou(202206040001)
Yuanhao PENG, Kai CHENG, Haotian ZHU, et al. Individualized 3D printing guide plates-assisted surgical correction for severe kyphosis deformity[J]. Journal of Army Medical University, 2024, 46(21): 2443-2450.
DOI:
Yuanhao PENG, Kai CHENG, Haotian ZHU, et al. Individualized 3D printing guide plates-assisted surgical correction for severe kyphosis deformity[J]. Journal of Army Medical University, 2024, 46(21): 2443-2450. DOI: 10.16016/j.2097-0927.202404079.
Individualized 3D printing guide plates-assisted surgical correction for severe kyphosis deformity
本研究为单中心非随机性临床试验研究,样本涵盖2018年12月至2023年6月期间接受经椎体内截骨术的19例患者(男性8例,女性11例)。其中7例患者为导板组,采用计算机辅助设计(CAD)技术进行术前手术规划和术中个性化3D打印导板引导,12例患者为传统组,采用传统置钉矫正。术后评估包括后凸Cobb角、脊柱矫正率、钉位准确性和Oswestry功能障碍指数问卷调查(Oswestry Dysfunction Index
accuracy of pedicle screw fixation and overall clinical efficacy of intravertebral osteotomy and internal fixation surgery with the assistance of 3D printing guide plates in treatment of severe kyphosis.
Methods
2
A single-center nonrandomized clinical pilot study was conducted on 19 patients (8 males and 11 females) with severe kyphosis undergoing intravertebral osteotomy between December 2018 and June 2023. Seven of them (CAD group) had preoperative planning with computer-aided design (CAD) and intraoperative guidance of individualized 3D printing guide plates. And another 12 patients (control group) were corrected with conventional pedicle screw placement. Postoperative evaluation included assessment of posterior Cobb angle
spinal angular correction rate
accuracy of pedicle screw placement and Oswestry Dysfunction Index (ODI) questionnaire.
Results
2
The 19 patients were at a mean age of 48.0 years
and followed up for 26.4 (9~54) months. All of them achieved relatively satisfactory corrective results
with those of the CAD group having a correction rate of 96.83% and those of the control group of 86.61%. There were no statistical differences in average intraoperative blood loss (857
vs
1 045 mL) and average operative time (344
vs
402 min)
but significant difference was observed in average length of hospital stay (11
vs
18 d
P
<
0.05) between the 2 groups. A total of 278 nails were placed in this study
including 70 guide-assisted pedicle screws
97.1% of which were grade A or B. In the control group
208 pedicle screws were placed
93.8% of which were grade A or B. Postoperative CT/X-ray scanning displayed that both groups achieved certain correction for kyphosis. No obvious difference was found in the average spinal angular correction (43.37°
vs
36.10°)
and significantly higher correction rate was seen in the CAD group than the control group
(96.83%
vs
86.61%
P
<
0.01). The ODI value was notably lower in the CAD group than the control group (
P
<
0.05).
Conclusion
2
CAD-assisted preoperative planning
surgical simulation and individualized 3D printing guide plates can promote surgical correction and accuracy of pedicle screw placement and improves the quality of life of patients with severe kyphotic deformity.
关键词
Keywords
references
KOELÉM C, LEMS W F, WILLEMS H C. The clinical relevance of hyperkyphosis: a narrative review[J]. Front Endocrinol, 2020, 11: 5. DOI: 10.3389/fendo.2020.00005.
KADO D M, PRENOVOST K, CRANDALL C. Narrative review: hyperkyphosis in older persons[J]. Ann Intern Med, 2007, 147(5): 330-338. DOI: 10.7326/0003-4819-147-5-200709040-00008.
KADO D M, MILLER-MARTINEZ D, LUI L Y, et al . Hyperkyphosis, kyphosis progression, and risk of non-spine fractures in older community dwelling women: the study of osteoporotic fractures (SOF)[J ] . J Bone Miner Res, 2014, 29(10): 2210-2216. DOI: 10.1002/jbmr.2251.
MAO K Y, WANG Y, XIAO S H, et al . Clinical application of computer-designed polystyrene models in complex severe spinal deformities: a pilot study[J ] . Eur Spine J, 2010, 19(5): 797-802.DOI: 10.1007/s00586-010-1359-0.
DIEBO B G, SHAH N V, SOLOW M, et al . Adult cervical deformity: radiographic and osteotomy classifications[J ] . Orthopade, 2018, 47(6): 496-504. DOI: 10.1007/s00132-018-3581-0.
SCHWAB F, BLONDEL B, CHAY E, et al . The comprehensive anatomical spinal osteotomy classification[J ] . Neurosurgery, 2015, 76(Suppl 1): S33-S41;discussionS41. DOI: 10.1227/01.neu.0000462076.73701.09.
LUAN H P, LIU K, KAHAER A, et al . Pedicle subtraction osteotomy for the corrective surgery of ankylosing spondylitis with thoracolumbar kyphosis: experience with 38 patients[J ] . BMC Musculoskelet Disord, 2022, 23(1): 731. DOI: 10.1186/s12891-022-05693-z.
QIAN B P, HUANG J C, QIU Y, et al . Complications of spinal osteotomy for thoracolumbar kyphosis secondary to ankylosing spondylitis in 342 patients: incidence and risk factors[J ] . J Neurosurg Spine, 2018, 30(1): 91-98. DOI: 10.3171/2018.6.SPINE171277.
GERTZBEIN S D, ROBBINS S E. Accuracy of pedicular screw placement in vivo [J ] . Spine, 1990, 15(1): 11-14. DOI: 10.1097/00007632-199001000-00004.
FARCY J P, WEIDENBAUM M, GLASSMAN S D. Sagittal index in management of thoracolumbar burst fractures[J]. Spine, 1990, 15(9): 958-965. DOI: 10.1097/00007632-199009000-00022.
JIN C, WANG S R, YANG G D, et al . A review of the methods on Cobb angle measurements for spinal curvature[J ] . Sensors (Basel), 2022, 22(9): 3258. DOI: 10.3390/s22093258.
PAN Y L, CHEN Q R, CHEN T T, et al . Evaluation of a computer-aided method for measuring the Cobb angle on chest X-rays[J ] . Eur Spine J, 2019, 28(12): 3035-3043. DOI: 10.1007/s00586-019-06115-w.
FAIRBANK J, PYNSENT P. The Oswestry disability index[J]. Spine, 2000, 25(22): 2940-2952. DOI: 10.1097/00007632-200011150-00017.
CECCHINATO R, BERJANO P, ZERBI A, et al . Pedicle screw insertion with patient-specific 3D-printed guides based on low-dose CT scan is more accurate than free-hand technique in spine deformity patients: a prospective, randomized clinical trial[J ] . Eur Spine J, 2019, 28(7): 1712-1723. DOI: 10.1007/s00586-019-05978-3.
LUO M, WANG W G, YANG N N, et al . Does three-dimensional printing plus pedicle guider technology in severe congenital scoliosis facilitate accurate and efficient pedicle screw placement?[J ] . Clin Orthop Relat Res, 2019, 477(8): 1904-1912. DOI: 10.1097/CORR.0000000000000 739.
CHEN P C, CHANG C C, CHEN H T, et al . The accuracy of 3D printing assistance in the spinal deformity surgery[J ] . Biomed Res Int, 2019, 2019: 7196528. DOI: 10.1155/2019/7196528.
Removal of plaque biofilm by a 3D-printed precision pericoronal irrigator for impacted wisdom teeth: an in vitro study
Clinical application and prospect of 3D printing technology in orthopeadics
Clinical efficacy of 3D printed anatomical titanium alloy tibial prosthesis in reconstruction of bone defects in the lower tibia
Planned prosthesis size in modified 3D printed patient-specific instrumentation-assisted total knee arthroplasty: analysis of consistency with intraoperative size
Current status and perspective of 3D printing technology to construct articular osteochondral scaffolds with boundary structure
Related Author
HUANG Zhenxin
GUO Weiwei
SONG Jianye
ZHOU Liang
WEI Rouyi
WANG Kun
WANG Xiaoting
Guoxian PEI
Related Institution
Department of Stomatology, Second Affiliated Hospital, Army Medical University (Third Military Medical University)
No.9 Department, nstitute of Field Surgery, Army Medical Center of PLA/Daping Hospital of Third Military Medical University
Medical College of Nanchang University
Department of Orthopeadics, Jiangxi Provincial Key Laboratory of Health Digital Orthopeadics, Jiangxi Provincial People's Hospital
Intelligent and Digital Surgery Innovation Center of SUSTech, Southern University of Science and Technology Hospital