Planned prosthesis size in modified 3D printed patient-specific instrumentation-assisted total knee arthroplasty: analysis of consistency with intraoperative size
Monographic Reports: Clinical application of 3D printing technology in bone and joint surgery|更新时间:2026-01-09
|
Planned prosthesis size in modified 3D printed patient-specific instrumentation-assisted total knee arthroplasty: analysis of consistency with intraoperative size
Journal of Army Medical UniversityVol. 44, Issue 15, Pages: 1523-1530(2022)
Bin SUN, Jixiao ZHANG, Liang YUAN, et al. Planned prosthesis size in modified 3D printed patient-specific instrumentation-assisted total knee arthroplasty: analysis of consistency with intraoperative size[J]. Journal of Army Medical University, 2022, 44(15): 1523-1530.
DOI:
Bin SUN, Jixiao ZHANG, Liang YUAN, et al. Planned prosthesis size in modified 3D printed patient-specific instrumentation-assisted total knee arthroplasty: analysis of consistency with intraoperative size[J]. Journal of Army Medical University, 2022, 44(15): 1523-1530. DOI: 10.16016/j.2097-0927.202202034.
Planned prosthesis size in modified 3D printed patient-specific instrumentation-assisted total knee arthroplasty: analysis of consistency with intraoperative size
To explore the prediction accuracy of prosthesis size of the modified 3D printed patient-specific instrumentation (PSI)-assisted total knee arthroplasty (TKA) by comparing the difference between intraoperative prosthesis size and preoperative planned prosthesis size.
Methods
2
We optimized and improved the workflow
design scheme
shape of guide
operation technology and verification method of the traditional PSI-assisted TKA. A total of 126 patients (137 knees) who received 3D printed PSI-assisted TKA based on preoperative CT images in our center were recruited in this study. The differences between the actual prosthesis size and the preoperative planned size were compared and analyzed.
Results
2
All patients completed the surg
ery successfully
and no serious complications occurred. For the femoral side
the difference between the actual used size and the doctor's planned size was observed in 7 cases
and the difference between the actual used size and the engineer's planned size was in 25 cases (
P
<
0.01)
and the engineer's planned size adjusted by doctors was 19 cases (
P
<
0.01). For the tibial side
the actual used size and the doctor's planned size was 39 cases
the difference between the actual used size and the engineer's planned size was 44 cases
and the engineer's planned size adjusted by doctors was 5 cases.
Conclusion
2
The planned prosthesis size in Our modified 3D printed PSI-assisted TKA is consistent with the intraoperative prosthesis size
and the plan will be more consistent after adjustment by the doctors.
关键词
Keywords
references
BERGER R A, ROSENBERG A G, BARDEN R M, et al . Long-term followup of the Miller-Galante total knee replacement[J ] . Clin Orthop Relat Res, 2001(388): 58-67. DOI: 10.1097/00003086-200107000-00010.
INSALL J N, BINAZZI R, SOUDRY M, et al . Total knee arthroplasty[J ] . Clin Orthop Relat Res, 1985, 192: 13-22. DOI: 10.1097/00003086-198501000-00003.
RITTER M A, KEATING E M, SUEYOSHI T, et al . Twenty-five-years and greater, results after nonmodular cemented total knee arthroplasty[J ] . J Arthroplasty, 2016, 31(10): 2199-2202. DOI: 10.1016/j.arth.2016.01.043.
ARNHOLT C M, MACDONALD D W, MALKANI A L, et al . Corrosion damage and wear mechanisms in long-term retrieved CoCr femoral components for total knee arthroplasty[J ] . J Arthroplasty, 2016, 31(12): 2900-2906. DOI: 10.1016/j.arth.2016.05.006.
YANG B, YUAN L, ZHANG K, et al . The accuracy of bony resection and component size planning of total knee arthroplasty assisted with a novel designed patient-specific instrumentation[J ] . Chin J Orthop, 2021, 41(2): 67-75. DOI: 10.3760/cma.j.cn121113-20200715-00451.
NOBLE J W Jr, MOORE C A,LIU N. The value of patient-matched instrumentation in total knee arthroplasty[J]. J Arthroplasty, 2012, 27(1): 153-155. DOI: 10.1016/j.arth.2011.07.006.
PIETSCH M, DJAHANI O, ZWEIGER C, et al . Custom-fit minimally invasive total knee arthroplasty: effect on blood loss and early clinical outcomes[J ] . Knee Surg Sports Traumatol Arthrosc, 2013, 21(10): 2234-2240. DOI: 10.1007/s00167-012-2284-z.
CUCCHI D, MENON A, COMPAGNONI R, et al . Significant differences between manufacturer and surgeon in the accuracy of final component size prediction with CT-based patient-specific instrumentation for total knee arthroplasty[J ] . Knee Surg Sports Traumatol Arthrosc, 2018, 26(11): 3317-3324. DOI: 10.1007/s00167-018-4876-8.
ISSA K, RIFAI A, MCGRATH M S, et al . Reliability of templating with patient-specific instrumentation in total knee arthroplasty[J ] . J Knee Surg, 2013, 26(6): 429-433. DOI: 10.1055/s-0033-1343615.
SCHOTANUS M G M, SCHOENMAKERS D A L, SOLLIE R, et al . Patient-specific instruments for total knee arthroplasty can accurately predict the component size as used peroperative[J ] . Knee Surg Sports Traumatol Arthrosc, 2017, 25(12): 3844-3848. DOI: 10.1007/s00167-016-4345-1.
GONG S, XU W H, WANG R Y, et al . Patient-specific instrumentation improved axial alignment of the femoral component, operative time and perioperative blood loss after total knee arthroplasty[J ] . Knee Surg Sports Traumatol Arthrosc, 2019, 27(4): 1083-1095.DOI: 10.1007/s00167-018-5256-0.
IVIE C B, PROBST P J, BAL A K, et al . Improved radiographic outcomes with patient-specific total knee arthroplasty[J ] . J Arthroplasty, 2014, 29(11): 2100-2103. DOI: 10.1016/j.arth.2014.06.024.
HEYSE T J, TIBESKU C O. Improved femoral component rotation in TKA using patient-specific instrumentation[J]. Knee, 2014, 21(1): 268-271. DOI: 10.1016/j.knee.2012.10.009.
HEYSE T J, TIBESKU C O. Improved tibial component rotation in TKA using patient-specific instrumentation[J]. Arch Orthop Trauma Surg, 2015, 135(5): 697-701. DOI: 10.1007/s00402-015-2157-2.
KE S, RAN T F, HE Y Q, et al . Does patient-specific instrumentation increase the risk of notching in the anterior femoral cortex in total knee arthroplasty? A comparative prospective trial[J ] . Int Orthop, 2020, 44(12): 2603-2611. DOI: 10.1007/s00264-020-04779-4.
CHAREANCHOLVANICH K, NARKBUNNAM R, PORN-RATTANAMANEEWONG C. A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement[J]. Bone Joint J, 2013, 95-B(3): 354-359. DOI: 10.1302/0301-620X.95B3.29903.
NUNLEY R M, ELLISON B S, ZHU J J, et al . Do patient-specific guides improve coronal alignment in total knee arthroplasty?[J ] . Clin Orthop Relat Res, 2012, 470(3): 895-902. DOI: 10.1007/s11999-011-2222-2.
ZOMAR B O, VASARHELYI E M, SOMERVILLE L E, et al . A randomized trial investigating the cost-utility of patient-specific instrumentation in total knee arthroplasty in an obese population[J ] . J Arthroplasty, 2021, 36(9): 3078-3088. DOI: 10.1016/j.arth.2021.04.029.
CHRISTEN B, TANNER L, ETTINGER M, et al . Comparative cost analysis of four different computer-assisted technologies to implant a total knee arthroplasty over conventional instrumentation[J ] . J Pers Med, 2022, 12(2): 184. DOI: 10.3390/jpm12020184.
ATTARD A, TAWY G F, SIMONS M, et al . Health costs and efficiencies of patient-specific and single-use instrumentation in total knee arthroplasty: a randomised controlled trial[J ] . BMJ Open Qual, 2019, 8(2): e000493. DOI: 10.1136/bmjoq-2018-000493.
KOSSE N M, HEESTERBEEK P J C, SCHIMMEL J J P, et al . Stability and alignment do not improve by using patient-specific instrumentation in total knee arthroplasty: a randomized controlled trial[J ] . Knee Surg Sports Traumatol Arthrosc, 2018, 26(6): 1792-1799. DOI: 10.1007/s00167-017-4792-3.
VAN LEEUWEN J A M J, SNORRASON F, RÖHRL S M. No radiological and clinical advantages with patient-specific positioning guides in total knee replacement[J]. Acta Orthop, 2018, 89(1): 89-94. DOI: 10.1080/17453674.2017.1393732.
NUNLEY R M, ELLISON B S, RUH E L, et al . Are patient-specific cutting blocks cost-effective for total knee arthroplasty?[J ] . Clin Orthop Relat Res, 2012, 470(3): 889-894. DOI: 10.1007/s11999-011-2221-3.
YAMAMURA K, MINODA Y, SUGAMA R, et al . Design improvement in patient-specific instrumentation for total knee arthroplasty improved the accuracy of the tibial prosthetic alignment in the coronal and axial planes[J ] . Knee Surg Sports Traumatol Arthrosc, 2020, 28(5): 1560-1567. DOI:10.1007/s00167-019-05571-7.