Preparation and preliminary clinical application of 3D printed porous tantalum prosthesis for reconstruction of acetabular bone defect
Monographic Reports: Clinical application of 3D printing technology in bone and joint surgery|更新时间:2026-01-09
|
Preparation and preliminary clinical application of 3D printed porous tantalum prosthesis for reconstruction of acetabular bone defect
Journal of Army Medical UniversityVol. 44, Issue 15, Pages: 1516-1522(2022)
作者机构:
400038 重庆,陆军军医大学(第三军医大学)第一附属医院关节外科
作者简介:
基金信息:
the Key R&D Program of Guangdong Province(2018B090906003);the National Key R&D Program of China(2016YFB1101404);the Individualized Training Project of Excellent Talent Pool of Army Medical University(XZ-2019-505-038)
Huaquan FAN, Fuyou WANG, Rui HE, et al. Preparation and preliminary clinical application of 3D printed porous tantalum prosthesis for reconstruction of acetabular bone defect[J]. Journal of Army Medical University, 2022, 44(15): 1516-1522.
DOI:
Huaquan FAN, Fuyou WANG, Rui HE, et al. Preparation and preliminary clinical application of 3D printed porous tantalum prosthesis for reconstruction of acetabular bone defect[J]. Journal of Army Medical University, 2022, 44(15): 1516-1522. DOI: 10.16016/j.2097-0927.202112205.
Preparation and preliminary clinical application of 3D printed porous tantalum prosthesis for reconstruction of acetabular bone defect
To establish the preparation technology and production method of 3D printed porous tantalum prosthesis for reconstruction of acetabular bone defect and to evaluate its efficacy in preliminary clinical application.
Methods
2
Four patients with Paprosky ⅡB bone defect of the acetabulum who were going to undergo total hip arthroplasty (THA) in our center from June to December 2018 were recruited in this study according to our inclusion and exclusion criteria. Based on the personalized CT data of the patient's complete pelvis
the prosthesis for acetabular bone defect was designed by specialized 3D software and fabricated by powder bed electron beam additive manufacturing. The hip visual analog scale (VAS) score
36-item short-form (SF-36) score and Harris score for hip joint before and after operation were compared
the postoperative complications were observed
and radiographical examination was completed at the last follow-up.
Results
2
All patients obtained good postoperative outcomes after undergoing computer-designed
personalized 3D printed porous tantalum prosthesis assisted THA to repair Paprosky ⅡB acetabular defects. During a follow-up period of 22.5±3.5 months
no surgery-related complications occurred. At the last follow-up
the Harris score was 86.2±0.3
which was significantly higher than that before operation (64.5±12.1)
indicating significantly improved hip function; The VAS score was 2.5±0.7
obviously lower than that before operation (7.5±0.7); the SF-36 score was increased from 39.3±6.3 to 81.6±3.8. Compared to preoperative conditions
significant improvement was achieved in quality of life. Radiological examination showed that the implanted hip prosthesis and the 3D printed porous tantalum prosthesis for the bone defect had good stability
no radiolucent line was observed
and the acetabular bone defect was effectively reconstructed.
Conclusion
2
Our established preparation technology of personalized 3D printed porous tantalum prosthesis has demonstrated excellent effectiveness for reconstruction of acetabular bone defects in preliminary clinical application.
MAO K Y, LIU J H, CUI X. Application of bone tissue engineering materials in the repair of large bone defects[J]. Med J Chin PAPF, 2020, 31(4): 277-280, 283. DOI: 10.14010/j.cnki.wjyx.2020.04.001.
ZHAO D W, LI J L. Fabrication of the porous tantalum and its current status used as orthopedics implants materials[J]. Acta Metallurgica Sinica, 2017, 53(10): 1303-1310. DOI: 10.11900/0412.1961.2017.00260.
YANG K, TANG H P, WANG J, et al . Research development of standardized and addtively manufactured custom-made porous tantalum implant[J ] . Hot Working Technol, 2017, 46(22): 5-8, 12. DOI: 10.14158/j.cnki.1001-3814.2017.22.002.
BODHAK S, BOSE S, BANDYOPADHYAY A. Role of surface charge and wettability on early stage mineralization and bone cell-materials interactions of polarized hydroxyapatite[J]. Acta Biomater, 2009, 5(6): 2178-2188. DOI: 10.1016/j.actbio.2009.02.023.
LI X, WANG L, YU X M, et al . Tantalum coating on porous Ti 6 Al 4 V scaffold using chemical vapor deposition and preliminary biological evaluation[J ] . Mater Sci Eng C Mater Biol Appl, 2013, 33(5): 2987-2994. DOI: 10.1016/j.msec.2013.03.027..
LIU P S, CUI G, CHEN Y M, et al . Macro mesh structural porous tantalum prepared by dipping and sintering organic foams: CN101660076A[P ] . 2010-03-03.
RUAN J M, YE L, XIE J Q, et al . Porous tantalum serving as medical metal implanted material and preparation method thereof: CN102205144A[P ] . 2011-10-05.
WENG X S. Surgical techniques for acetabular defect reconstruction in revision total hip arthroplasty[J]. Chin J Bone Joint Surg, 2022, 15(1): 10-18. DOI: 10.3969/j.issn.2095-9958.2022.01.02.
SUN J T, CAI S T, WANG S H, et al . Application of 3D-printed metal block to repair acetabular bone defects in total hip arthroplasty[J ] . J Tradit Chin Orthop Traumatol, 2020, 32(7): 41-42, 46.
ZHAO M C, HUANG Y, YUAN W J, et al . Application of three-dimensional printing metal implants in orthopedics[J ] . Chin J Tissue Eng Res, 2018, 22(31): 5027-5033. DOI: 10.3969/j.issn.2095-4344.0566.
PAGNANO W, HANSSEN A D, LEWALLEN D G, et al . The effect of superior placement of the acetabular component on the rate of loosening after total hip arthroplasty[J ] . J Bone Joint Surg Am, 1996, 78(7): 1004-1014. DOI: 10.2106/00004623-199607000-00004.
WANG X, LIU H C. Research progress of autologous bone transplantation in repair of bone defects at implant sites[J]. Chin J Prosthodont, 2016, 17(1): 49-52. DOI: 10.3969/j.issn.1009-3761.2016.01.013.
ZHONG Z M, HAN Z M, XIU H, et al . Immune response and treatment of allogeneic bone transplantation[J ] . Acta Acad Med Jiangxi, 2003, 43(6): 146-149. DOI: 10.3969/j.issn.1000-2294.2003.06.056.
TANG C, LI X X, LING C, et al . The medium-term outcomes of 3D printed customized porous titanium acetabular prosthesis replacement for acetabular bone defection[J ] . J Nanjing Med Univ(Nat Sci), 2019, 39(12): 1774-1778.
ZHANG Y M. The researches on titanium and titanium alloy in dental use[J]. J Biomed Eng, 2000, 7(2): 206-208, 217. DOI: 10.3321/j.issn:1001-5515.2000.02.022.
LIU Y, BAO C, WISMEIJER D, et al . The physicochemical/biological properties of porous tantalum and the potential surface modification techniques to improve its clinical application in dental implantology[J ] . Mater Sci Eng C, 2015, 49: 323-329. DOI: 10.1016/j.msec.2015.01.007.
SAMBAZIOTIS C, LOVY A J, KOLLER K E, et al . Histologic retrieval analysis of a porous tantalum metal implant in an infected primary total knee arthroplasty[J ] .J Arthroplasty, 2012, 27(7): 1413. e5-1413. e9. DOI: 10.1016/j.arth.2011.10.025.
BALLA V K, BODHAK S, BOSE S, et al . Porous tantalum structures for bone implants: fabrication, mechanical and in vitro biological properties[J ] . Acta Biomater, 2010, 6(8): 3349-3359. DOI: 10.1016/j.actbio.2010.01.046..
CASSAR-GHEITI A J, MEI X Y, AFENU E A, et al . Midterm outcomes after reconstruction of superolateral acetabular defects using flying buttress porous tantalum augments during revision total hip arthroplasty[J ] . J Arthroplasty, 2021, 36(8): 2936-2941. DOI: 10.1016/j.arth.2021.03.040.
SCHILDHAUER T A, ROBIE B, MUHR G, et al . Bacterial adherence to tantalum versus commonly used orthopedic metallic implant materials[J ] . J Orthop Trauma, 2006, 20(7): 476-484. DOI: 10.1097/00005131-200608000-00005.
MIETTINEN H J, MIETTINEN S S, KETTUNEN J S, et al . Revision hip arthroplasty using a porous tantalum acetabular component[J ] . HIP Int, 2021, 31(6): 782-788. DOI:10.1177/1120700020913294.