Hongya DAI, Yibing ZHOU, Jindong QIAN, et al. Error analysis of 4-dimensional computed tomography image reconstruction in uniform motion state[J]. Journal of Third Military Medical University, 2021, 43(19): 1913-1918.
Hongya DAI, Yibing ZHOU, Jindong QIAN, et al. Error analysis of 4-dimensional computed tomography image reconstruction in uniform motion state[J]. Journal of Third Military Medical University, 2021, 43(19): 1913-1918. DOI: 10.16016/j.1000-5404.202104025.
目的
2
利用呼吸运动体模研究均匀运动状态的四维CT(four-dimensional computed tomography,4DCT)图像重建的准确性。
方法
2
采用呼吸运动体模模拟肺部肿瘤在头脚方向做余弦波运动,根据呼吸运动幅度(A)和周期(T)分为6组(A=0.5 cm,T=6 s;A=0.5 cm,T=3 s;A=1.0 cm,T=6 s;A=1.0 cm,T=3 s;A=1.5 cm,T=6 s;A=1.5 cm,T=3 s)。采用Philips大孔径CT模拟机进行4DCT扫描。6组运动采用相同的扫描参数分别采集2次。在重建的4DCT图像上勾画内靶区(internal target volume,ITV),统计ITV的体积和长度,并与理论值比较,分析4DCT重建误差。
结果
2
ITV在6组4DCT重建图像上均有误差,TIV体积误差相对值范围为1.13%~27.64%;ITV长度误差范围为-0.95~0.03 cm,ITV长度平均误差为0.25 cm。在A=1.5 cm,T=6 s时,ITV重建误差最大,前后2次扫描的体积误差分别为27.64%和19.04%,长度误差分别为0.11 cm和0.84 cm。呼吸运动参数和扫描参数完全一致的前后2次4DCT图像重建的靶体积和长度也均存在差异,在A=0.5 cm,T=6 s时,前后2次的ITV体积误差分别为14.47%和4.43%,相差达10.00%。
结论
2
均匀运动情况下4DCT图像重建存在误差,且误差大小随机化,具有不可预见性。
Objective
2
To investigate the accuracy of 4-dimensional computed tomography (4DCT) image reconstruction in uniform motion state using respiratory motion phantom.
Methods
2
The cosine wave movements of lung tumor in the craniocaudal direction were simulated on a respiratory motion phantom
and the simulations were divided into 6 groups according to the respiratory motion amplitude (A) and the breathing period (T)
that is
A=0.5 cm
T=6 s; A=0.5 cm
T=3 s; A=1.0 cm
T=6 s; A=1.0 cm
T=3 s; A=1.5 cm
T=6 s; A=1.5 cm T=3 s. 4DCT scans were acquired using Philips big bore CT simulator
and the movements were scanned twice for each group
with same scanning parameters. Finally
the internal target volume (ITV) was delineated on the reconstructed 4DCT images
and the volume and length of ITVs were measured and compared with their theoretical values to analyze reconstruction errors of 4DCT.
Results
2
There were errors in all the 6 groups of ITV delineated on the reconstructed 4DCT images. The relative error range of ITV volume was from 1.13% to 27.64%
and that of ITV length was from -0.95~0.03 cm. The average ITV length error was 0.25 cm. When A=1.5 cm and T=6 s
the reconstruction error showed the maximum
and the volume errors presented 27.64% and 19.04%
while the length errors were 0.11 and 0.84 cm between the 2 scans. There were also differences between the results of 2 scans under the same respiratory motion parameters and scanning parameters. When A=0.5 cm and T=6 s
the volume errors of ITV were 14.47% and 4.43% in the 2 scans
with a difference of 10.00%.
Conclusion
2
There remain errors in 4DCT image reconstruction under uniform motion
and the size of the error is randomized and unpredictable.
KEALL P J, MAGERAS G S, BALTER J M, et al . The management of respiratory motion in radiation oncology report of AAPM Task Group 76[J ] . Med Phys, 2006, 33(10): 3874-3900. DOI:10.1118/1.2349696.
UNDERBERG R W, LAGERWAARD F J, CUIJPERS J P, et al . Four-dimensional CT scans for treatment planning in stereotactic radiotherapy for stage Ⅰ lung cancer[J ] . Int J Radiat Oncol Biol Phys, 2004, 60(4): 1283-1290. DOI: 10.1016/j.ijrobp.2004.07.665.
MAXIM P G, LOO B W Jr, SHIRAZI H, et al . Quantification of motion of different thoracic locations using four-d imensional computed tomography: implications for radiotherapy planning[J ] . Int J Radiat Oncol Biol Phys, 2007, 69(5): 1395-1401. DOI:10.1016/j.ijrobp.2007.05.016.
LIU H H, BALTER P, TUTT T, et al . Assessing respiration-induced tumor motion and internal target volume using four-dimensional computed tomography for radiotherapy of lung cancer[J ] . Int J Radiat Oncol Biol Phys, 2007, 68(2): 531-540. DOI:10.1016/j.ijrobp.2006.12.066.
STROOM J C, HEIJMED B J. Geometrical uncertainties, radiotherapy planning margins, and the ICRU-62 report[J]. Radiother Oncol, 2002, 64(1): 75-83. DOI:10.1016/s0167-8140(02)00140-8.
COLE A J, HANNA G G, JAIN S, et al . Motion management for radical radiotherapy in non-small cell lung cancer[J ] . Clin Oncol (R Coll Radiol), 2014, 26(2): 67-80. DOI:10.1016/j.clon.2013.11.001.
WIKSTRÖM K A, ISACSSON U M, PINTO M C, et al . Evaluation of irregular breathing effects on internal target volume definition for lung cancer radiotherapy[J ] . Med Phys, 2021, 48(5): 2136-2144. DOI:10.1002/mp.14824.
WANG Y, LIU T, CHEN H Q, et al . Comparison of interna l target volumes defined by three-dimensional, four-dimensional, and cone-beam computed tomography images of a motion phantom[J ] . Ann Transl Med, 2020, 8(22): 1488. DOI:10.21037/atm-20-6246.
ANTONY R, LONSKI P, UNGUREANU E, et al . Independent review of 4DCT scans used for SABR treatment planning[J ] . J Appl Clin Med Phys, 2020, 21(3): 62-67. DOI:10.1002/acm2.12825.
尚东平, 李明焕, 李建彬, 等. 四维CT技术确定肺内孤立性病灶靶体积应用[J]. 中华放射肿瘤学杂志, 2011, 20(5): 417-419. DOI:10.3760/cma.j.issn.1004-4221.2011.05.016.
SHANG D P, LI M H, LI J B, et al . The application of four-dimensional CT technique in determining the planning target volume of the solitary pulmonary lesion[J ] . Chin J Radiat Oncol, 2011, 20(5): 417-419. DOI:10.3760/cma.j.issn.1004-4221.2011.05.016.
HURKMANS C W, VAN LIESHOUT M, SCHURING D, et al . Quality assurance of 4D-CT scan techniques in multicenter phase Ⅲ trial of surgery versus stereotactic radiotherapy (radiosurgery or surgery for operable early stage (stage 1A) non-small-cell lung cancer [ROSEL ] study)[J ] . Int J Radiat Oncol Biol Phys, 2011, 80(3): 918-927. DOI:10.1016/j.ijrobp.2010.08.017.
PARK K, HUANG L, GAGNE H, et al . Do maximum intensity projection images truly capture tumor motion? [J ] . Int J Radiat Oncol, 2009, 73(2): 618-625. DOI:10.1016/j.ijrobp.2008.10.008.
UNDERBERG R W, LAGERWAARD F J, SLOTMAN B J, et al . Use of maximum intensity projections (MIP) for target volume generation in 4DCT scans for lung cancer[J ] . Int J Radiat Oncol Biol Phys, 2005, 63(1): 253-260. DOI:10.1016/j.ijrobp.2005.05.045.
MERCIECA S, BELDERBOS J S A, DE JAEGER K, et al . Interobserver variability in the delineation of the primary lung cancer and lymph nodes on different four-dimensional computed tomography reconstructions[J ] . Radiother Oncol, 2018, 126(2): 325-332. DOI:10.1016/j.radonc.2017.11.020.
LI H, NOEL C, GARCIA-RAMIREZ J, et al . Clinical evaluations of an amplitude-based binning algorithm for 4DCT reconstruction in radiation therapy[J ] . Med Phys, 2012, 39(2): 922-932. DOI:10.1118/1.3679015.
STEENBAKKERS R J, DUPPEN J C, FITTON I, et al . Reduction of observer variation using matched CT-PET for lung cancer delineation: a three-dimensional analysis[J ] . Int J Radiat Oncol Biol Phys, 2006, 64(2): 435-448. DOI:10.1016/j.ijrobp.2005.06.034.
NAKAMURA M, NARITA Y, SAWADA A, et al . Impact of motion velocity on four-dimensional target volumes: a phantom study[J ] . Med Phys, 2009, 36(5): 1610-1617. DOI:10.1118/1.3110073.
WATKINS W T, LI R, LEWIS J, et al . Patient-specific motion artifacts in 4DCT[J ] . Med Phys, 2010, 37(6): 2855-2861. DOI:10.1118/1.3432615.
0
Views
0
下载量
0
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution
京公网安备11010802024621