Detection effects of 4 near-infrared cerebral oximeters: a comparative study in simulated high altitude (4 500 m)
To evaluate the detection effects of 4 brands of near-infrared cerebral oximeters (NICOs)
INVOS 5100C (Medtronic)
EGOS-600 (Aegean)
MNIR-P100 (Mingxi) and BRS-1 (Casibrain Technology) in high-altitude environment.
Methods
2
From March 3
2021 to March 16
2021
a total of 174 healthy male volunteers (18~22 years old
172.8±5.1 cm
BMI 22.5±2.2 kg/m
2
) were subjected from Army Medical University
and then randomly divided into 6 groups (n=29). Their regional cerebral oxygen saturation (rcSO
2
)
instrument response time
blood pressure (BP)
heart rate (HR)
and pulse oxygen saturation (SPO
2
) were recorded and compared in the plain (300 m) and the 4 500-m altitude environment. The correlations among the above indicators were evaluated by Pearson correlation analysis. The difference and consistency of above indicators by the 4 devices were investigated.
Results
2
① The effective detection rates of all 4 NICOs were 100% in 4 500 m environment. In the plain environment
the response time was 2.13±0.34
6.60±1.23
2.87±0.48 and 2.65±0.47 s for INVOS 5100C
EGOS-600
MNIR-P100 and BRS-1
respectively
while the time was 2.22±0.41
6.83±1.25
2.86±0.49 and 2.67±0.49 s respectively in the 4 500-m altitude environment. EGOS-600 had the longest response time
with mo
re than 3 s longer than the other 3 NICOs (
P
<
0.05). There was no obvious difference in the response time in MNIR-P100 and BRS-1 between the plain and high altitude environment (
P
>
0.05). But the response time of INVOS 5100C and EGOS-600 was increased by 0.1 and 0.2 s respectively in the 4 500 m altitude environment than the plain (
P
<
0.05). ②In the plain environment
INVOS 5100C and BRS-1 showed better consistency in rcSO
2
detection than EGOS-600 and MNIR-P100. While in the 4 500 m altitude environment
EGOS-600 and INVOS 5100C performed better consistency than BRS-1 and MNIR-P100. When the subjects entered from the plain to 4 500 m altitude environment
the decline range showed better consistency in BRS-1 and INVOS 5100C than EGOS-600 and MNIR-P100. ③When taking the decline range of rcSO
2
reaching 20% as the intervention value of cerebral hypoxia
78%
94% and 76% of the subjects respectively detected by EGOS-600
MNIR-P100 and BRS-1 were under the threshold. When taking the decline range of rcSO
2
reaching 50% as the intervention value of cerebral hypoxia
52.9%
54.6% and 54.3% of the subjects respectively detected by EGOS-600
MNIR-P100 and BRS-1 were under the threshold. ④The correlation coefficients of SpO
2
and rcSO
2
detected by INVOS 5100C
EGOS-600
MNIR-P100 and BRS-1 were 0.876 4 (
P
<
0.000 1)
0.641 7 (
P
<
0.000 1)
0.761 8 (
P
<
0.000 1) and 0.6525 (
P
<
0.0001)
respectively.
Conclusion
2
In terms of rcSO
2
INVOS 5100C shows great consistency with the 3 domestic NICOs
and the 3 domestic instruments also show great specificity but less good sensitivity.
GAO Y Q, HUANG Q Q, LIU F Y, et al . Research progress on measures to promote acclimatization at high altitude[J ] . J Prev Med Chin, 2002, 20(4): 306-309. DOI: 10.3969/j.issn.1001-5248.2002.04.032.
LIAN G F, LI J, LUO Y J, et al . Research progress on the pathogenesis and prevention of high altitude cerebral edema[J ] . People Mil Surg, 2020, 63(4): 343-346, 357. DOI: CNKI:SUN:RMJZ.0.2020-04-013.
IMRAY C H, CLARKE T, FORSTER P J, et al . Carbon dioxide contributes to the beneficial effect of pressurization in a portable hyperbaric chamber at high altitude[J ] . Clin Sci (Lond), 2001, 100(2): 151-157. DOI: 10.1042/CS20000058.
HADOLT I, LITSCHER G. Noninvasive assessment of cerebral oxygenation during high altitude trekking in the Nepal Himalayas (2850-5600 m)[J]. Neurol Res, 2003, 25(2): 183-188. DOI: 10.1179/016164103101201175.
QIU Y, QI S H. Research status of non-invasive monitoring technology of cerebral oxygen saturation[J]. Mod Med J, 2019, 47(8): 1045-1048. DOI: CNKI:SUN:TDYX.0.2019-08-034.
LIU Q, HAN R Q, Advanes in cerebral oxygensaturation monitoring and its application[J]. Int J Anesthesiol Resusc, 2018, 39(3): 234-238. DOI: 10.3760/cma.j.issn.1673-4378.2018.03.009.
BICKLER P E, FEINER J R, ROLLINS M D. Factors affecting the performance of 5 cerebral oximeters during hypoxia in healthy volunteers[J]. Anesth Analg, 2013, 117(4): 813-823. DOI: 10.1213/ANE.0b013e318297d763.
SHAH N, TRIVEDI N K, CLACK S L, et al . Impact of hypoxemia on the performance of cerebral oximeter in volunteer subjects[J ] . J Neurosurg Anesthesiol, 2000, 12(3): 201-209. DOI: 10.1097/00008506-200007000-00001.
MURKIN J M, ADAMS S J, NOVICK R J, et al . Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study[J ] . Anesth Analg, 2007, 104(1): 51-58. DOI: 10.1213/01.ane.0000246814.29362.f4.
LA COUR A, GREISEN G, HYTTEL-SORENSEN S. In vivo validation of cerebral near-infrared spectroscopy: a review[J]. Neurophotonics, 2018, 5(4): 040901. DOI: 10.1117/1.nph.5.4.040901.
BENNI P B, MACLEOD D, IKEDA K, et al . A validation method for near-infrared spectroscopy based tissue oximeters for cerebral and somatic tissue oxygen saturation measurements[J ] . J Clin Monit Comput, 2018, 32(2): 269-284. DOI: 10.1007/s10877-017-0015-1.
BLAND J M, ALTMAN D G. Applying the right statistics: analyses of measurement studies[J]. Ultrasound Obstet Gynecol, 2003, 22(1): 85-93. DOI: 10.1002/uog.122.
QU R C. Accurate measurement of photoelectric sensor response time[J]. Electron Qual, 2014(4): 64-66. DOI: 10.3969/j.issn.1003-0107.2014.04.016.
TOMLIN K L, NEITENBACH A M, BORG U. Detection of critical cerebral desaturation thresholds by three regional oximeters during hypoxia: a pilot study in healthy volunteers[J]. BMC Anesthesiol, 2017, 17(1): 6. DOI: 10.1186/s12871-016-0298-7.
GERKE O. Reporting standards for a bland-Altman agreement analysis: a review of methodological reviews[J]. Diagnostics (Basel), 2020, 10(5): E334. DOI: 10.3390/diagnostics10050334.
A X R, ZHAO S X, ZHAO W J. Clinical application of point-of-care testing in plateau region of China[J]. Chin J Lab Med, 2014, 37(11): 812-814. DOI: 10.3760/cma.j.issn.1009-9158.2014.11.004.
KUSAKA T, ISOBE K, NAGANO K, et al . Quantification of cerebral oxygenation by full-spectrum near-infrared spectroscopy using a two-point method[J ] . Comp Biochem Physiol A Mol Integr Physiol, 2002, 132(1): 121-132. DOI: 10.1016/s1095-6433(01)00538-4.
AINSLIE P N, OGOH S. Regulation of cerebral blood flow in mammals during chronic hypoxia: a matter of balance[J]. Exp Physiol, 2010, 95(2): 251-262. DOI: 10.1113/expphysiol.2008.045575.
YU B W, WANG G L, DENG X M, et al . Clinical anesthesia monitoring guidelines: chinese anesthesiology guidelines and expert consensus[M ] . 2017. Beijing: People's Medical Publishing House, 2017, 168-169.
Protective effect of sevoflurane preconditioning against lung ischemia-reperfusion injury in patients with one lung ventilation
Comparison of cerebral oxygen saturation values measured by different brands of near-infrared spectrometers under hypoxic conditions: a randomized controlled trial in healthy volunteers
Related Author
REN Yukun
ZHANG Xucheng
LI Hong
LIU Yan
LI Jiaxin
WANG Zhuo
Zhenxin DUAN
Guiying YANG
Related Institution
Department of Anesthesiology, Second Affiliated Hospital, Army Medical University (Third Military Medical University)
Department of Anesthesiology, No. Hospital of Joint Logistic Support Force of PLA
Department of Anesthesiology, No. Hospital of Joint Logistic Support Force of PLA
Department of Anesthesiology, First Affiliated Hospital, Army Medical University (Third Military Medical University)
Research Center for Nutrition and Food Safety, Faculty of Military Preventive Medicine, Army Medical University (Third Military Medical University)