Yi XIANG, Li JIANG, Bin QIN, et al. Precision of bilateral electrode implantation and treatment outcome in Parkinson's disease after subthalamic nucleus deep brain stimulation[J]. Journal of Third Military Medical University, 2020, 42(5): 517-522.
DOI:
Yi XIANG, Li JIANG, Bin QIN, et al. Precision of bilateral electrode implantation and treatment outcome in Parkinson's disease after subthalamic nucleus deep brain stimulation[J]. Journal of Third Military Medical University, 2020, 42(5): 517-522. DOI: 10.16016/j.1000-5404.201910055.
Precision of bilateral electrode implantation and treatment outcome in Parkinson's disease after subthalamic nucleus deep brain stimulation
To investigate whether there is a significant difference in the accuracy of the two-sided electrode implanted during the same period in Parkinson's disease (PD) with subthalamic nucleus deep brain stimulation (STN-DBS)
and whether the accuracy affects the bilateral postoperative effects.
Methods
2
Twenty-nine PD patients admitted in our department from January 2016 to December 2018
who underwent bilateral STN-DBS surgery were re-examined with CT scanning on the head after bilateral electrodes were implanted during the operation. The results were merged with preoperative 3T-MRI images (with planned targets). The implanted electrodes were measured and calculated for error distance from the planned targets. Unified Parkinson's Disease Rating Scale Ⅲ (UPDRSⅢ) was used to evaluate the improvement of the symptoms before and in 6 months after operation. Pearson correlation analysis was used to study the correlation among age
gender
onset age and course of disease with improvement rate and deviation.
Results
2
Finally
27 patients were enrolled
and 54 electrodes were implanted successfully. There were no significant differences in the age
gender
onset age and course of disease and the improvement rate and deviation among these patients (
P
>
0.05). No obvious difference was seen in the mean deviation between the first implanted electrodes and second implanted electrodes (
t
=1.437
P
>
0.05). The symptoms of all patients were obviously improved. In 6 months after surgery
the average improvement rate of systemic symptoms was (78.10±10.25)% with drug on
and the average decrease of the equivalent dose of levodopa was 45.07%. The improvement rate was (83.15± 9.16)% for the first side
and (83.52±11.21)% for the second side. There was no significant difference in the improvement of limb symptoms between the 2 sides (
P
>
0.05).
Conclusion
2
The accuracy of simultaneous implantation of bilateral electrodes in STN-DBS is not significantly different from that of the electrodes implanted sequentially
while still achieving good clinical outcomes.
关键词
Keywords
references
YU H, NEIMAT J S. The treatment of movement disorders by deep brain stimulation[J]. Neurotherapeutics, 2008, 5(1):26-36. DOI: 10.1016/j.nurt.2007.10.072.
ODEKERKEN V J, VAN LAAR T, STAAL M J, et al . Subthalamic nucleus vers us globus pallidus bilateral deep brain stimulation for advanced Parkinson's disease (NSTAPS study): a randomised controlled trial[J ] . Lancet Neurol, 2013, 12(1): 37-44. DOI:10.1016/S1474-4422(12)70264-8.
XIAN W B, PEI Z, ZHOU X Y, et al . Efficacy and safety of bilateral subthalamic nucleus deep brain stimulation for treating parkinson disease: a Meta-analysis[J ] . Chin J Nerv Ment Dis, 2009, 35(5): 289-294. DOI:10.3969/j.issn.1002-0152.2009.05.009.
GURIDI J, RODRÍGUEZ-OROZ M C, CLAVERO P, et al . Critical review of the subthalamic stimulation in Parkinson's disease[J ] . Neurocirugia (Astur), 2009, 20(6): 521-532.
ELIAS W J, FU K M, FRYSINGER R C. Cortical and subcortical brain shift during stereotactic procedures[J]. J Neuro- surg, 2007, 107(5): 983-988. DOI:10.3171/JNS-07/11/0983.
MIYAGI Y, SHIMA F, SASAKI T. Brain shift: an error factor during implantation of deep brain stimulation electrodes[J]. J Neurosurg, 2007, 107(5): 989-997. DOI:10.3171/JNS-07/11/0989.
IVAN M E, YARLAGADDA J, SAXENA A P, et al . Brain shift during bur hole-based procedures using interventional MRI[J ] . J Neurosurg, 2014, 121(1): 149-160. DOI:10.3171/2014.3.JNS121312.
SLOTTY P J, KAMP M A, WILLE C, et al . The impact of brain shift in deep brain stimulation surgery: observation and obviation[J ] . Acta Neurochir (Wien), 2012, 154(11): 2063-2068. DOI:10.1007/s00701-012-1468-y.
AZMI H, MACHADO A, DEOGAONKAR M, et al . Intracranial air correlates with preoperative cerebral atrophy and stereotactic error during bilateral STN DBS Surgery for Parkinson's disease[J ] . Stereotact Funct Neurosurg, 2011, 89(4): 246-252. DOI:10.1159/000327916.
LIU H G, LIU D F, WANG H M, et al . Deep brain stimulation with a PINS device in Parkinson's disease: long-term follow-up[J ] . Chin J Neurosurg, 2017, 33(11): 1133-1136. DOI:10.3760/cma.j.issn.1001-2346.2017.11.012.
BENABID A L, KOUDSIE A, BENAZZOUZ A, et al . Imaging of subthalamic nucleus and ventralis intermedius of the thalamus[J ] . Mov Dis ord, 2002, 17(Suppl 3): S123-S129. DOI:10.1002/mds.10153.
HU X W, ZHOU X P, JIANG X F, et al . Intraoperative verification and adjustment of subthalamic nucleus electrode placement in deep brain stimulation for Parkinson's disease[J ] . J Clin Neurosurg, 2008, 5(3): 113-116. DOI:10.3969/j.issn.1672-7770.2008.03.001.
MONTGOMERY E B Jr. Microelectrode targeting of the subthalamic nucleus for deep brain stimulation surgery[J]. Mov Disord, 2012, 27(11): 1387-1391. DOI:10.1002/mds.25000.
BENAZZOUZ A, BREIT S, KOUDSIE A, et al . Intraoperative microrecordings of the subthalamic nucleus in Parkinson's disease[J ] . Mov Disord, 2002, 17(Suppl 3): S145-S149. DOI:10.1002/mds.10156.
CHEN S D, GAO G D, FENG T, et al . Chinese expert consensus on programming deep brain stimulation for patients with Parkinson's disease[J ] . Transl Neurodegener, 2018, 7: 11. DOI:10.1186/s40035-018-0116-x.
SASAKI T, AGARI T, KUWA HARA K, et al . Efficacy of dural sealant system for preventing brain shift and improving accuracy in deep brain stimulation surgery[J ] . Neurol Med Chir (Tokyo), 2018, 58(5): 199-205. DOI:10.2176/nmc.oa.2017-0242.
ONDO W G, BRONTE-STEWART H, DBS STUDY GROUP. The North American survey of placement and adjustment strategies for deep brain stimulation[J]. Stereotact Funct Neurosurg, 2005, 83(4): 142-147. DOI:10.1159/000088654.
PEZESHKIAN P, DESALLES A A, GORGULHO A, et al . Accuracy of frame-based stereotactic magnetic resonance imaging vs frame-based stereotactic head computed tomography fused with recent magnetic resonance imaging for postimplantation deep brain Stimulator lead localization[J ] . Neurosurgery, 2011, 69(6): 1299-1306. DOI:10.1227/NEU.0b013e31822b7069.
SHAHLAIE K, LARSON P S, STARR P A. Intraoperative computed tomography for deep brain stimulation surgery: technique and accuracy assessment[J]. Neurosurgery, 2011, 68(1 Suppl Operative): 114-124; discussion 124. DOI:10.1227/NEU.0b013e31820781bc.
KREMER N I, OTERDOOM D L M, VAN LAAR P J, et al . Accuracy of intraoperative computed tomography in deep brain stimulation-A prospective noninferiority study[J ] . Neuromodulation, 2019, 22(4): 472-477. DOI:10.1111/ner.12918.
KOCHANSKI R B, KEROLUS M G, PAL G, et al . Use of intraoperative CT to predict the accuracy of microelectrode recording during deep brain stimulation surgery. A proof of concept study[J ] . Clin Neurol Neurosurg, 2016, 150: 164-168. DOI:10.1016/j.clineuro.2016.09.014.
BOT M, VAN DEN MUNCKHOF P, BAKAY R, et al . Accuracy of intraoperative computed tomography during deep brain stimulation procedures: comparison with postoperative magnetic resonance imaging[J ] . Stereotact Funct Neurosurg, 2017, 95(3): 183-188. DOI:10.1159/000475672.
GEEVARGHESE R, O'GORMAN TUURA R, LUMSDEN D E, et al . Registration accuracy of CT/MRI Fusion for localisation of deep brain stimulation electrode position: an imaging study and systematic review[J ] . Stereotact Funct Neurosurg, 2016, 94(3): 159-163. DOI:10.1159/000446609.
HAMID N A, MITCHELL R D, MOCROFT P, et al . Targeting the subthalamic nucleus for deep brain stimulation: technical approach and fusion of pre- and postoperative MR images to define accuracy of lead placement[J ] . J Neurol Neurosurg Psychiatry, 2005, 76(3): 409-414. DOI:10.1136/jnnp.2003.032029.
[PETERSEN E A, HOLL E M, MARTINEZ-TORRES I, et al . Minimizing brain shift in stereotactic functional neurosurgery[J ] . Neurosurgery, 2010, 67(3 Suppl Operative): ons213-221, discussion ons221. DOI: 10.1227/01.NEU.0000380991.23444.08 http://dx.doi.org/10.1227/01.NEU.0000380991.23444.08 ..
MATIAS C M, FRIZON L A, ASFAHAN F, et al . Brain shift and pneumocephalus assessment during frame-based deep brain stimulation implantation with intraoperative magnetic resonance imaging[J ] . Oper Neurosurg (Hagerstown), 2018, 14(6):668-674. DOI: 10.1093/ons/opx170.
Efficacy of subthalamic deep brain stimulation for levodopa-induced dyskinesia in patients with Parkinson’s disease: retrospective analysis of 12 cases