Xiaohui LI, Guoyan LIU, Xiaohua LIU, et al. Effect of lower tube voltage combined with CARE Dose 4D on radiation dose and image quality of aorta CT angiography[J]. Journal of Third Military Medical University, 2017, 39(21): 2099-2104.
Xiaohui LI, Guoyan LIU, Xiaohua LIU, et al. Effect of lower tube voltage combined with CARE Dose 4D on radiation dose and image quality of aorta CT angiography[J]. Journal of Third Military Medical University, 2017, 39(21): 2099-2104. DOI: 10.16016/j.1000-5404.201706078.
目的
2
采用单源光子CT较低管电压联合Care Dose 4D技术,评价主动脉成像图像质量和辐射剂量。
方法
2
43例患者按照扫描管电压随机分为2组,即120 kV组(
n
=19)和100 kV组(
n
=24)。平扫采用固定管电压(100、120 kV),增强扫描采用自动管电压调节技术(Care kV),参考管电压预设为100、120 kV,平扫和增强扫描均联合Care Dose 4D技术。比较不同扫描条件下2组患者主动脉图像质量、CT值和辐射剂量变化,包括平均容积CT剂量指数(CTDIvol)、剂量长度乘积(DLP)和有效剂量(ED)。
结果
2
平扫,2组主动脉上中下段CT值差异无统计学意义(
P
>
0.05),100 kV组辐射剂量(CTD
Ivol、DLP、ED)明显低于120 kV组。平扫和增强扫描,100 kV组有效剂量ED较120 kV组分别降低41.36%和67.80%。Care 100 kV组主动脉各段的CT值均高于120 kV组(
P
<
0.05),两组主动脉成像图像质量评分差异无统计学意义(
P
>
0.05)。
结论
2
联合Care Dose 4D技术,Care kV低管电压能够明显降低主动脉CT血管成像辐射剂量,而图像质量不受影响。
Objective
2
To evaluate the radiation dose and image quality of aorta CT angiography by combined usage of CARE KV and CARE Dose 4D technology.
Methods
2
A retrospective study was carried out on aorta CT angiography images of 43 patients. According to the tube voltage
they were assigned into 120 kV group (
n
=19) and 100 kV group (
n
=24). The plain scanning was performed at fixed tube voltage (100 and 120 kV)
while the enhancement scanning was perform with CARE KV (tube voltage 100 and 120 kV). Both the plain and enhancement scanning were combined with CARE Dose 4D technology. The imaging quality scores
mean CT values of the plain scaning under different tube voltage
as well as the enhancement scanning
radiation doses
that is
average computed tomography dose index volume (CTDIvol)
dose length product (DLP) and effective dose (ED) were compared between 2 groups.
Results
2
There were no significant differences in the mean CT values at upper
middle and lower sections of aorta in plain scans (
P
>
0.05). The radiation doses (CTDIvol
DLP and ED) of 100 kV group was significantly lower than those of 120 kV group. In addition
ED was reduced by 41.36% and 67.80% in plain and enhanced scans
respectively in the 100 kV group when compared with the 120 kV group. The former group also had higher mean CT values of different sections of aorta (
P
<
0.05). But there was no difference in the imaging quality scores between 2 groups (
P
>
0.05).
Conclusion
2
Aorta CT angiography using CARE KV100 together with CARE Dose 4D can ensure the image quality
and significantly reduce the radiation dose when compared with CARE KV120.
BRENNER D J, HALL E J. Computed tomography--an increasing source of radiation exposure[J]. New Engl J Med, 2007, 357(22):2277-2284.DOI: 10.1056/NEJMra 072149.
MUHOGORA W, REHANI M M. Review of the current status of radiation protection in diagnostic radiology in Africa[J].J Med Imaging (Bellingham), 2017, 4(3):031202. DOI: 10.1117/1.JMI.4.3.031202.
GOLDING S J, SHRIMPTON P C. Radiation dose in CT: are we meeting the challenge[J]. Br J Radiol, 2002, 75(889): 1-4. DOI: 10.1259/bjr.75.suppl_1.750001.
MARINE J E. A new ALARA: Keeping leadless pacing thresholds as low as reasonably achievable[J]. Heart Rhythm, 2017, 14(5):692-693.DOI:10.1016/j.hrthm.2017.01.033.
NIJHOF W H, BALTUSSEN E J, KANT I M, et al. Low-dose CT angiography of the abdominal aorta and reduced contrast medium volume: Assessment of image quality and radiation dose[J]. Clin Radiol, 2016, 71(1):64-73. DOI: 10.1016/j.crad.2015.10.007.
宫凤玲, 于鹏, 张惠英, 等. 256层螺旋CT主动脉成像中不同触发监测点对对比剂剂量的影响[J].中国医学影像技术, 2011, 27(7):1495-1498. DOI:10.13929/j.1003-3289.2011.07.014.
GONG F L, YU P, ZHANG H Y, et al. Impact of different triger monitoring points on contrast media doses in 256-slice spiral CT aortic angiography [J]. Chin J Med Imaging Technol, 2011, 27(7):1495-1498. DOI:10.13929/j.1003-3289.2011.07.014.
GOETTI R, WINKLEHNER A, Gordic S, et al. Automated attenuation-based kilovoltage selection: preliminary observations in patients after endovascular aneurysm repair of the abdominal aorta[J]. AJR Am J Roentgenol, 2012, 199(3): W380-W385.DOI: 10.2214/AJR.11.8029.
SHIN H J, CHUNG Y E, LEE Y H, et al. Radiation dose reduction via sinogram affirmed iterative reconstruction and automatic tube voltage modulation (CARE kV) in abdominal CT[J]. Korean Journal of Radiology, 2013, 14(6):886-893.DOI: 10.3348/kjr.2013.14.6.886.
YANG B, LI Z L, GAO Y, et al.Image quality evaluation for CARE kV technique combined with iterative reconstruction for chest computed tomography scanning[J].Medicine (Baltimore), 2017, 96(11):e6175. DOI: 10.1097/MD.0000000000006175.
李剑, 郑敏文, 石明国, 等.低kV和CARE Dose 4D技术对降低双源CT主动脉成像辐射剂量的价值[J].临床放射学杂志, 2011, 30(10):1528-1531. DOI:10.13437/j.cnki.jcr.2011.10.013.
LI J, ZHENG M W, SHI M G, et al.Reducing radiation dose with Low-kV and CARE Dose4D in dual-source CT aorta angiography[J].J Clin Radiol, 2011, 30(10):1528-1531.DOI:10.13437/j.cnki.jcr.2011.10.013.
LESCHKA S, STOLZMANN P, SCHMID F T, et al. Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose[J]. Eur Radiol, 2008, 18(9):1809-1817.DOI:10.1007/s00330-008-0966-1.
COPPENRATH E, MEINDL T, HERZOG P, et al. Dose reduction in multidetector CT of the urinary tract. Studies in a phantom model.[J]. Eur Radiol, 2006, 16(9):1982-1989.DOI: 10.1007/s00330-005-0138-5.
WINKLEHNER A, GOETTI R, BAUMUELLER S, et al. Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness[J]. Invest Radiol, 2011, 46(12):767-773.DOI:10.1097/RLI.0b013e3182266448.
卢光明, 张龙江.低辐射剂量低对比剂用量CT的应用[J].中华医学杂志, 2014, 94(29):2241. DOI:10.3760/cma.j.issn.0376-2491.2014.29.001.
LU G M, Zhang L J. CT application with low radiation dose and low contrast media.[J].Natl Med J China, 2014, 94(29):2241.DOI:10.3760/cma.j.issn.0376-2491.2014.29.001.
王淑婷, 任克.腹部低剂量CT技术临床应用现状与研究趋势[J].国际医学放射学杂志, 2015, 38(6):552-556. DOI:10.3874/j.issn.1674-1897.2015.06.Z0610.
WANG S T, REN K. Clinical application and research trend of low dose CT technology in the abdominal scan[J].Int J Med Radiol, 2015, 38(6):552-556.DOI:10.3874/j.issn.1674-1897.2015.06.Z0610.
吕培杰, 柴亚如, 阎晓朋, 等. CT能谱智能匹配技术联合自适应统计迭代重组技术对腹部低对比剂量扫描图像质量和辐射剂量的影响[J].中华放射学杂志, 2016, 50(2):122-127. DOI:10.3760/cma.j.issn.1005-1201.2016.02.010.
LU P J, CHAI Y R, YAN X P, et al. Effect of automatic spectral imaging mode selection and adaptive statistical iterative reconstruction at abdominal CT with low contrast agent dose.[J].Chin J Radiol, 2016, 50(2):122-127. DOI:10.3760/cma.j.issn.1005-1201.2016.02.010.
HIGASHIGAITO K, SCHMID T, PUIPPE G, et al. CT angiography of the aorta: prospective evaluation of individualized low-volume contrast media protocols[J].Radiology, 2016, 280(3):960-968. DOI: 10.1148/radiol.201615 1982.
0
Views
0
下载量
0
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution
京公网安备11010802024621