Objective
2
To investigate the relationship of power and exposure period of 980 nm continuous laser system on the surface sterilization for metal medical equipment.
Methods
2
The sterile steel lancets were artificially contaminated with
Staphylococcus aureus
and
Escherichia coli
respectively. These lancets were exposed and disinfected by 980 nm continuous laser at different powers for different time periods. The inactivation of the bacteria on each lancet was observed and measured by 8-hour enrichment culture.
Results
2
Staphylococcus aureus
were thoroughly killed after the lancets were disinfected by the 980 nm laser at 5.24 W for 29 s
6.87 W for 24 s
8.33 W for 14 s
and 10.11 W for 14 s
respectively. The power and exposure period for
Escherichia coli
were 5.24 W for 22 s
6.87 W for 12 s
8.33 W for 11 s
and 10.11 W for 9 s
respectively.
Escherichia coli
was more likely to be inactivated than
Staphylococcus aureus
. According to the experimental data
the survival rate under same power of infrared laser as a function of irradiation time was fitted to the first-order kinetics equation.
Conclusion
2
980 nm infrared laser can achieve rapid sterilization on metal surface of medical equipment in a very short time period. The larger power the laser is with
the shorter is the exposure period
and the better the sterilizing effect is.
中华人民共和国卫生部 . 医疗机构消毒技术规范: GB/T1. 1-2009 [S ] . 北京 : 中国标准出版社 , 2012 : 25 - 41 . 中华人民共和国卫生部. 医疗机构消毒技术规范: GB/T1. 1-2009[S]. 北京: 中国标准出版社, 2012: 25-41.
The Ministry of Health of the People's Republic of China . Regulation of disinfection technique in healthcare settings: GB/T1.1-2009 [S ] . Beijing : Standards Press of China , 2012 : 25 - 41 . The Ministry of Health of the People's Republic of China. Regulation of disinfection technique in healthcare settings: GB/T1.1-2009[S]. Beijing: Standards Press of China, 2012: 25-41.
黄 秋华 涂 慧慧 吴 莉 三种消毒液对硅胶气管套管消毒效果比较 中国消毒学杂志 2012 29 3 192 193 http://www.cnki.com.cn/Article/CJFDTOTAL-ZGXD201203007.htm 黄秋华, 涂慧慧, 吴莉.三种消毒液对硅胶气管套管消毒效果比较[J].中国消毒学杂志, 2012, 29(3): 192-193.
Huang Q H Tu H H Wu L Comparison of disinfection efficacy of three liquid disinfectants on silicone tracheal cannula Chinese Journal of Disinfection 2012 29 3 192 193 http://www.cnki.com.cn/Article/CJFDTOTAL-ZGXD201203007.htm Huang Q H, Tu H H, Wu L. Comparison of disinfection efficacy of three liquid disinfectants on silicone tracheal cannula[J]. Chinese Journal of Disinfection, 2012, 29(3): 192-193.
李 钟响 吴 红映 金 美琴 常见细菌对紫外线敏感性试验观察 中华医院感染学杂志 2012 22 1 121 122 http://www.cnki.com.cn/Article/CJFDTOTAL-ZHYY201201054.htm 李钟响, 吴红映, 金美琴.常见细菌对紫外线敏感性试验观察[J].中华医院感染学杂志, 2012, 22(1): 121-122.
Li Z X Wu H Y Jin M Q Experimental observation of ultraviolet light sensitivity effect on common bacteria Chinese Journal of Nosocomiology 2012 22 1 121 122 http://www.cnki.com.cn/Article/CJFDTOTAL-ZHYY201201054.htm Li Z X, Wu H Y, Jin M Q. Experimental observation of ultraviolet light sensitivity effect on common bacteria[J]. Chinese Journal of Nosocomiology, 2012, 22(1): 121-122.
周 逞丽 快速生物监测法对高压蒸汽灭菌效果监测的评价 实用预防医学 2008 15 3 885 886 10.3969/j.issn.1006-3110.2008.03.120 周逞丽.快速生物监测法对高压蒸汽灭菌效果监测的评价[J].实用预防医学, 2008, 15(3): 885-886. DOI:10.3969/j.issn.1006-3110.2008.03.120.
Zhou C L Effect Evaluation of a Rapid Biological Monitoring Method in High—Pressure Sterilization Practical Preventive Medicine 2008 15 3 885 886 10.3969/j.issn.1006-3110.2008.03.120 Zhou C L. Effect Evaluation of a Rapid Biological Monitoring Method in High—Pressure Sterilization[J]. Practical Preventive Medicine, 2008, 15(3): 885-886. DOI:10.3969/j.issn.1006-3110.2008.03.120.
Deppe H Horch H H Schrödl V Effect of 308-nm excimer laser light on peri-implantitis-associated bacteria: an in vitro investigation Lasers Med Sci 2007 22 4 223 227 10.1007/s10103-007-0441-2 Deppe H, Horch H H, Schrödl V, et al. Effect of 308-nm excimer laser light on peri-implantitis-associated bacteria: an in vitro investigation[J ] . Lasers Med Sci, 2007, 22(4): 223-227. DOI:10.1007/s10103-007-0441-2.
Mohaidat Q I . Laser-induced breakdown spectroscopy (LIBS): an innovative tool for studying bacteria [D ] . Detroit : Graduate Schoolof Wayne State University , 2011 . Mohaidat Q I. Laser-induced breakdown spectroscopy (LIBS): an innovative tool for studying bacteria[D]. Detroit:Graduate Schoolof Wayne State University, 2011.
Ando Y Aoki A Watanabe H Bactericidal effect of erbium YAG laser on periodontopathic bacteria Lasers Surg Med 1996 19 2 190 200 10.1002/(ISSN)1096-9101 Ando Y, Aoki A, Watanabe H, et al. Bactericidal effect of erbium YAG laser on periodontopathic bacteria[J]. Lasers Surg Med, 1996, 19(2): 190-200. DOI:10.1002/(SICI)1096-9101(1996) 19:2<190::aid-lsm11>3.0.CO; 2-B.
李 若兰 任 煜光 孙 风阳 脉冲Nd:YAG激光对几种细菌杀菌作用的实验研究 口腔医学 2000 20 1 16 17 10.3969/j.issn.1003-9872.2000.01.008 李若兰, 任煜光, 孙风阳.脉冲Nd:YAG激光对几种细菌杀菌作用的实验研究[J].口腔医学, 2000, 20(1): 16-17. DOI:10.3969/j.issn.1003-9872.2000.01.008.
Li R L Ren Y G Sun F Y Experimental study on bactericidal effects of pulsed nd:YAg laser Stomatology 2000 20 1 16 17 10.3969/j.issn.1003-9872.2000.01.008 Li R L, Ren Y G, Sun F Y. Experimental study on bactericidal effects of pulsed nd:YAg laser[J]. Stomatology, 2000, 20(1): 16-17. DOI:10.3969/j.issn.1003-9872.2000.01.008.
胡 鹏 陈 发良 重复频脉冲激光辐照金属材料热效应模拟分析 中国激光 2016 43 10 1001006 10.3788/cjl201643.1001006 胡鹏, 陈发良.重复频脉冲激光辐照金属材料热效应模拟分析[J].中国激光, 2016, 43(10): 1001006. DOI: 10.3788/cjl201643.1001006.
Hu P Chen F L Simulation of thermal effects in repetitive pulsed laser irradiation on metals Chinese Journal of Lasers 2016 43 10 1001006 10.3788/cjl201643.1001006 Hu P, Chen F L. Simulation of thermal effects in repetitive pulsed laser irradiation on metals[J]. Chinese Journal of Lasers, 2016, 43(10): 1001006. DOI: 10.3788/cjl201643.1001006.
Ongeng D Haberbeck L U Mauriello G Modeling the Fate of Escherichia coli O157:H7 and Salmonella enterica in the Agricultural Environment: Current Perspective J Food Sci 2014 79 4 R421 R427 10.1111/1750-3841.12392 Ongeng D, Haberbeck L U, Mauriello G, et al. Modeling the Fate of Escherichia coli O157:H7 and Salmonella enterica in the Agricultural Environment: Current Perspective [J ] . J Food Sci, 2014, 79(4): R421-R427. DOI:10.1111/1750-3841.12392.
Hiatt C W Kinetics of the inactivation of viruses Bacteriol Rev 1964 28 2 150 163 https://www.ncbi.nlm.nih.gov/pubmed/14172021 Hiatt C W. Kinetics of the inactivation of viruses [J]. Bacteriol Rev, 1964, 28 (2): 150-163.
杭 锋 陈 卫 陈 帅 微波加热对大肠杆菌与金黄色葡萄球菌蛋白质与核酸变性的影响 食品与机械 2008 24 5 12 15 http://www.cnki.com.cn/Article/CJFDTOTAL-SPJX200805006.htm 杭锋, 陈卫, 陈帅, 等.微波加热对大肠杆菌与金黄色葡萄球菌蛋白质与核酸变性的影响[J].食品与机械, 2008, 24(5): 12-15.
Hang F Chen W Chen S The effect of microwave heating on the denaturation of E.coli and S.aureus intracelluar protein and nucleic acid Food and Machinery 2008 24 5 12 15 http://www.cnki.com.cn/Article/CJFDTOTAL-SPJX200805006.htm Hang F, Chen W, Chen S, et al. The effect of microwave heating on the denaturation of E.coli and S.aureus intracelluar protein and nucleic acid[J ] . Food and Machinery, 2008, 24(5):12-15.
Woo I S Rhee I K Park H D Differential Damage in Bacterial Cells by Microwave Radiation on the Basis of Cell Wall Structure Appl Environ Microbiol 2000 66 5 2243 2247 10.1128/aem.66.5.2243-2247.2000 Woo I S, Rhee I K, Park H D. Differential Damage in Bacterial Cells by Microwave Radiation on the Basis of Cell Wall Structure[J]. Appl Environ Microbiol, 2000, 66(5): 2243-2247. DOI:10.1128/aem.66.5.2243-2247.2000.
0
Views
0
下载量
0
CSCD
Publicity Resources
Related Articles
Related Author
Related Institution
京公网安备11010802024621