1.陆军军医大学第二附属医院急诊科,重庆 400038
2.重庆市綦江区赶水镇中心卫生院,重庆 401437
Received:30 December 2025,
Published:25 April 2026
移动端阅览
蒋杨,田师宇,周人杰等.穿戴式无线脑血氧监测头带应用于急诊环境的可行性研究[J].局解手术学杂志,2026,35(04):359-365.
JIANG Yang,TIAN Shi-yu,ZHOU Ren-jie,et al.A feasibility study on the application of wearable wireless cerebral oximetry monitoring headbands in emergency environment[J].JOURNAL OF REGIONAL ANATOMY AND OPERATIVE SURGERY,2026,35(04):359-365.
蒋杨,田师宇,周人杰等.穿戴式无线脑血氧监测头带应用于急诊环境的可行性研究[J].局解手术学杂志,2026,35(04):359-365. DOI: 10.11659/jjssx.12E025142.
JIANG Yang,TIAN Shi-yu,ZHOU Ren-jie,et al.A feasibility study on the application of wearable wireless cerebral oximetry monitoring headbands in emergency environment[J].JOURNAL OF REGIONAL ANATOMY AND OPERATIVE SURGERY,2026,35(04):359-365. DOI: 10.11659/jjssx.12E025142.
目的
2
评估无线式脑血氧监测头带在急诊复杂环境中的应用,验证其是否可为急救场景提供快速、连续、精准的局部脑氧饱和度(rSO
2
)。
方法
2
比较有线脑血氧、无线脑血氧监测的建立时间,并招募健康志愿者评估体位变化与运动状态下无线脑血氧监测的稳定性。在急诊场景中,监测不同急诊状态患者在诊疗操作前、中、后的rSO
2
变化。
结果
2
无线脑血氧监测建立时间显著短于有线脑血氧监测(
P
<
0.01)。在动态测试中,受试者蹲踞后即刻rSO
2
较站立时显著升高(
P
<
0.01),而恢复站立后rSO
2
与站立基线值相比无明显差异(
P
>
0.01),表明设备在动态动作中具有良好的响应稳定性。跑步运动全程无线式脑血氧监测头带监测信号连续无中断。救护车转运场景下,无线式脑血氧监测头带与有线脑血氧监测设备在信号脱落和信号干扰方面无显著差异(
P
>
0.01),两者在静态、运动扰动时、颠簸时测得的rSO
2
值呈线性相关。患者在急诊环境中的有创操作、放射检查及电除颤/电复律前、中、后三个阶段的rSO
2
测量值无显著变化(
P
>
0.01)。
结论
2
无线式脑血氧监测头带适用于急救场景,可提供连续可靠的rSO
2
监测数据,具有急诊应用价值。
Objective
2
To evaluate the application of wireless cerebral oximetry monitoring headbands in the complex emergency environment and verify whether they can provide rapid,continuous and accurate regional cerebral oxygen satu
ration (rSO
2
) for emergency scenarios.
Methods
2
The establishment times of wired and wireless cerebral oximetry monitoring were compared,and healthy volunteers were recruited to evaluate the stability of wireless cerebral oximetry monitoring under changes in body position and during exercise. In emergency scenarios, the changes of rSO
2
of patients were monitored before, during and after diagnosis and treatment procedures on different emergency conditions.
Results
2
The establishment time of wireless cerebral oximetry monitoring was significantly shorter than that of wired cerebral oximetry monitoring (
P
<
0.01).In the dynamic test,the rSO
2
of subjects immediately after squatting was significantly higher than that at standing (
P
<
0.01),but there was no significant difference in rSO
2
of subjects after returning to standing compared with the baseline value of standing (
P
>
0.01),indicating that the device had good response stability during dynamic movements. The wireless cerebral oximetry monitoring headband throughout the running exercise provides continuous and uninterrupted monitoring signals. In the ambulance transfer scenario,there was no significant difference between wireless cerebral oximetry monitoring headband and wired cerebral oximetry monitoring devices in terms of signal shedding and signal interference(
P
>
0.01), the rSO
2
values measured by the two were linearly correlated under static, motion disturbances, and jolts conditions. There was no significant change in the rSO
2
values of patients before,during and after invasive operation,radiological examination or electrical defibrillation/electrical cardioversion in the emergency environment (
P
>
0.01).
Conclusion
2
Wireless cerebral oximetry monitoring headbands are suitable for emergency scenarios and can provide continuous and reliable rSO₂ monitoring data,which has emergency application value.
Pries PJ , Mutch WAC , Funk DJ . Characterizing drivers of change in intraoperative cerebral saturation using supervised machine learning [J]. J Clin Monit Comput , 2025 , 39 ( 3 ): 559 - 569 . doi: 10.1007/s10877-025-01265-3 http://dx.doi.org/10.1007/s10877-025-01265-3 .
Bertini P , Marabotti A , Paternoster G , et al . Regional cerebral oxygen saturation to predict favorable outcome in extracorporeal cardiopulmonary resuscitation: a systematic review and meta-analysis [J]. J Cardiothorac Vasc Anesth , 2023 , 37 ( 7 ): 1265 - 1272 . doi: 10.1053/j.jvca.2023.01.007 http://dx.doi.org/10.1053/j.jvca.2023.01.007 .
Van den Bempt S , Wauters L , Dewolf P . Pulseless electrical activity: detection of underlying causes in a prehospital setting [J]. Med Princ Pract , 2021 , 30 ( 3 ): 212 - 222 . doi: 10.1159/000513431 http://dx.doi.org/10.1159/000513431 .
Moerman A , Wouters P . Near-infrared spectroscopy (NIRS) monitoring in contemporary anesthesia and critical care [J]. Acta Anaesthesiol Belg , 2010 , 61 ( 4 ): 185 - 194 .
Jaeger M , Soehle M , Meixensberger J . Brain tissue oxygen (PtiO 2 ):a clinical comparison of two monitoring devices [J]. Acta Neurochir Suppl , 2005 , 95 : 79 - 81 . doi: 10.1007/3-211-32318-x_17 http://dx.doi.org/10.1007/3-211-32318-x_17 .
Cheng A , Magid DJ , Auerbach M , et al . Part 6:Resuscitation Education Science:2020 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care [J]. Circulation , 2020 , 142 : s551 - s579 . doi: 10.1161/CIR.0000 000000000903 http://dx.doi.org/10.1161/CIR.0000000000000903 .
Gregory AJ , Hatem MA , Yee K , et al . Optimal placement of cerebral oximeter monitors to avoid the frontal sinus as determined by computed tomography [J]. J Cardiothorac Vasc Anesth , 2016 , 30 ( 1 ): 127 - 133 . doi: 10.1053/j.jvca.2015.07.004 http://dx.doi.org/10.1053/j.jvca.2015.07.004 .
Juanning SI , Zhang X , Meng LI , et al . Wearable wireless real-time cerebral oximeter for measuring regional cerebral oxygen saturation [J]. Information Sciences , 2021 , 64 ( 1 ): 112203 . doi: 10.3389/fneur. 2024.1445563 http://dx.doi.org/10.3389/fneur.2024.1445563 .
McInnis KJ , Balady GJ . Comparison of submaximal exercise responses using the Bruce vs modified Bruce protocols [J]. Med Sci Sports Exerc , 1994 , 26 ( 1 ): 103 - 107 .
El-Baba RM , Schury MP . Neuroanatomy,frontal cortex [M]. StatPearls. Treasure Island (FL) : StatPearls Publishing , 2025 .
Hogue CW , Levine A , Hudson A , et al . Clinical applications of near-infrared spectroscopy monitoring in cardiovascular surgery [J]. Anesthesiology , 2021 , 134 ( 5 ): 784 - 791 . doi: 10.1097/ALN.00000 00000003700 http://dx.doi.org/10.1097/ALN.0000000000003700 .
Tribuddharat S , Ngamsaengsirisup K , Mahothorn P , et al . Correlation and agreement of regional cerebral oxygen saturation measured from sensor sites at frontal and temporal areas in adult patients undergoing cardiovascular anesthesia [J]. Peer J , 2022 , 10 : e14058 . doi: 10.7717/peerj.14058 http://dx.doi.org/10.7717/peerj.14058 .
李阳 . 近红外光谱技术测定国人脑氧饱和度基线值及其影响因素的研究 [D]. 重庆 : 中国人民解放军陆军军医大学 , 2023 .
Hirose T , Shiozaki T , Nomura J , et al . Pre-hospital portable monitoring of cerebral regional oxygen saturation (rSO 2 ) in seven patients with out-of-hospital cardiac arrest [J]. BMC Res Notes , 2016 , 9 ( 1 ): 428 . doi: 10.1186/s13104-016-2239-4 http://dx.doi.org/10.1186/s13104-016-2239-4 .
Hamanaka K , Shimoto M , Hitosugi M , et al . Cerebral oxygenation monitoring during resuscitation by emergency medical technicians:a prospective multicenter observational study [J]. Acute Med Surg , 2020 , 7 ( 1 ): e528 . doi: 10.1002/ams2.528 http://dx.doi.org/10.1002/ams2.528 .
Jeong H , Jeong S , Lim HJ , et al . Cerebral oxygen saturation measured by near-infrared spectroscopy and jugular venous bulb oxygen saturation during arthroscopic shoulder surgery in beach chair position under sevoflurane-nitrous oxide or propofol-remifentanil anesthesia [J]. Anesthesiology , 2012 , 116 ( 5 ): 1047 - 1056 . doi: 10.1097/ALN.0b013e31825154d2 http://dx.doi.org/10.1097/ALN.0b013e31825154d2 .
Papadopoulos G , Karanikolas M , Liarmakopoulou A , et al . Cerebral oximetry and cognitive dysfunction in elderly patients undergoing surgery for hip fractures:a prospective observational study [J]. Open Orthop J , 2012 , 6 : 400 - 405 . doi: 10.2174/1874325001206010400 http://dx.doi.org/10.2174/1874325001206010400 .
Budohoski KP , Czosnyka M , Kirkpatrick PJ , et al . Clinical relevance of cerebral autoregulation following subarachnoid haemorrhage [J]. Nat Rev Neurol , 2013 , 9 ( 3 ): 152 - 163 . doi: 10.1038/nrneurol. 2013.11 http://dx.doi.org/10.1038/nrneurol.2013.11 .
Ahn A , Yang J , Inigo-Santiago L , et al . A feasibility study of cerebral oximetry monitoring during the post-resuscitation period in comatose patients following cardiac arrest [J]. Resuscitation , 2014 , 85 ( 4 ): 522 - 526 . doi: 10.1016/j.resuscitation.2013.12.007 http://dx.doi.org/10.1016/j.resuscitation.2013.12.007 .
Shellen S , Parnia S , Huppert EL , et al . Integrating rSO 2 and EEG monitoring in cardiopulmonary resuscitation:a novel methodology [J]. Resusc Plus , 2024 , 18 : 100644 . doi: 10.1016/j.resplu.2024. 100644 http://dx.doi.org/10.1016/j.resplu.2024.100644 .
Tobias JD . Cerebral oxygenation monitoring:near-infrared spectroscopy [J]. Expert Rev Med Devices , 2006 , 3 ( 2 ): 235 - 243 . doi: 10.1586/17434440.3.2.235 http://dx.doi.org/10.1586/17434440.3.2.235 .