SUN BING-JIE, CAI YU-QING, YUAN GAO-FENG, et al. Study on the ultrasonic cleaning effect and influencing factors of emergency laparoscopic surgical instruments at night. [J]. Journal of regional anatomy and operative surgery, 2025, 34(2): 121-124.
DOI:
SUN BING-JIE, CAI YU-QING, YUAN GAO-FENG, et al. Study on the ultrasonic cleaning effect and influencing factors of emergency laparoscopic surgical instruments at night. [J]. Journal of regional anatomy and operative surgery, 2025, 34(2): 121-124. DOI: 10.11659/jjssx.03E024115.
Study on the ultrasonic cleaning effect and influencing factors of emergency laparoscopic surgical instruments at night
To explore the ultrasonic cleaning effect and influencing factors of emergency laparoscopic surgical instruments at night.
Methods
2
A total of 698 emergency laparoscopic surgical instruments cleaned by ultrasound at night were selected as the research subjects,and the unqualified ultrasonic cleaning of instruments was statistically analyzed. Univariate and multivariate binary
Logistic
regression equations were used to analyze the factors that may lead to unqualified ultrasonic cleaning.
Results
2
Among 698 ultrasonic cleaning sessions for emergency laparoscopic surgical instruments at night,a total of 55 times (7.88%) were unqualified. The results of univariate and multivariate binary
Logistic
regression analysis showed that the time from cleaning to the end of surgery ≥10 hou
rs,luminal surgical instruments,nonstandard pretreatment,unqualified cleaning water quality,and unqualified cleaning medium were the risk factors that lead to the occurrence of unqualified ultrasonic cleaning of emergency laparoscopic surgical instruments at night (
OR
>
1,
P
<
0.05),while manual+ultrasonic cleaning was the protective factor to avoid the occurrence of unqualified ultrasonic cleaning of emergency laparoscopic surgical instruments at night (
OR
<
1,
P
<
0.05).
Conclusion
2
The unqualified rate of ultrasonic cleaning of emergency laparoscopic surgical instruments at night is relatively high,so it should be cleaned as early as possible after the surgery. Before ultrasonic cleaning,manual cleaning should be carried out as much as possible,especially for luminal surgical instruments that are difficult to clean,standardized pretreatment should be carried out,and qualified cleaning water quality and medium should be selected.
Yamashita T,Ando K.Low-intensity ultrasound induced cavitation and streaming in oxygen-supersaturated water:role of cavitation bubbles as physical cleaning agents[J].Ultrason Sonochem,2019,52(4):268-679.doi:10.1016/j.ultsonch.2018.11.025http://dx.doi.org/10.1016/j.ultsonch.2018.11.025.
Effects of different CO2 pneumoperitoneum pressures on delirium and cognitive function after laparoscopic radical resection of rectal cancer in the elderly
Analgesic effect of Esketamine combined with sufentanil after laparoscopic myomectomy
Pathogenic bacterium and risk factors of wound infection in patients with pressure injury in intensive care unit
Influencing factors and prediction model of postoperative gastrointestinal dysfunction in elderly patients with colorectal cancer
Clinical study of laparoscopic surgery combined with anti-infection and dienogest in the treatment of endometriosis
Related Author
WU Hao
GENG Ying
CHEN Yue
ZHONG Ming-jie
LIU Yong
XIANG Yan-qing
LIU Ying-chun
LIN Yi-he
Related Institution
Department of Anesthesiology,Suqian First People's Hospital Affiliated to Nanjing Medical University
Operating Room, Dongying People's Hospital
Department of Anesthesiology, Dongying People's Hospital
EICU, Nanjing Integrated Traditional Chinese and Western Medicine Hospital
Anorectal Department, Guang'an Hospital, West China Hospital of Sichuan University