Clinical value of intravascular ultrasound in diagnosis and treatment of peripheral arteriostenosis
|更新时间:2025-01-15
|
Clinical value of intravascular ultrasound in diagnosis and treatment of peripheral arteriostenosis
Journal of Army Medical UniversityVol. 33, Issue 8, Pages: 853-856(2011)
作者机构:
第三军医大学西南医院心血管内科,重庆市介入心脏病学研究所
作者简介:
基金信息:
DOI:
CLC:R540.46;R445.1;R543.5
Published:2011,
Published Online:30 April 2011,
稿件说明:
移动端阅览
HONG YUKUN, CHI LUXIANG, SONG ZHIYUAN, et al. Clinical value of intravascular ultrasound in diagnosis and treatment of peripheral arteriostenosis. [J]. Journal of army medical university, 2011, 33(8): 853-856.
DOI:
HONG YUKUN, CHI LUXIANG, SONG ZHIYUAN, et al. Clinical value of intravascular ultrasound in diagnosis and treatment of peripheral arteriostenosis. [J]. Journal of army medical university, 2011, 33(8): 853-856.DOI:
Clinical value of intravascular ultrasound in diagnosis and treatment of peripheral arteriostenosis
Objective To evaluate the clinical value of intravascular ultrasound (IVUS) in diagnosis and intravascular interventional therapy of peripheral arteriostenosis. Methods Ninety-four inpatients [58 males and 36 females
aged 26 to 77 (58.4±18.3)
]
with peripheral arteriostenosis who received diagnosis and treatment in Southwes
t Hospital
China
from Oct. 2006 to Sep. 2010 were enrolled. The patients were divided into a DSA group [43 patients examined with digital subtraction angiography (DSA)
]
and a DSA+IVUS group (51 patients examined with DSA plus IVUS). The examination results were compared. Stents were implanted under the guidance of DSA and IVUS
and the stent implantation effects were observed. General ultrasound and CTA follow-up were carried out. DSA and IVUS examination were conducted after the operation if necessary. Results In DSA+IVUS group
77 sites of stenotic vessels were detected with DSA
including 42 eccentric stenosis sites and 35 concentric stenosis sites; 82 sites of stenotic vessels were detected with IVUS
including 63 eccentric plaques and 19 concentric plaques. The diameter stenosis rate measured with IVUS [(67.1±12.2)%
]
was significantly higher than that measured with DSA [(54.5±11.4)%
]
(
P
<
0.05). The area stenosis rate measured with IVUS [(89.3±12.3)%
]
was significantly higher than that measured with DSA [(77.1±13.1)%
]
(
P
<
0.05). Stent implantation was successfully used to treat 82 sites of angiopathy in DSA+IVUS group and 53 sites in DSA group. After 3- to 48-month follow-up
the restenosis rate of DSA group (15.1%
8/53) was significantly higher than that of DSA+IVUS group (3.7%
3/82) (
P
<
0.05). Conclusion As compared with DSA
IVUS can judge pathological nature and stenosis severity more precisely
and can guide and evaluate stent implantation more effectively.