Comparison of IDEAL and 3D FSE-Cube sequences for normal postganglionic brachial plexus
|更新时间:2025-01-15
|
Comparison of IDEAL and 3D FSE-Cube sequences for normal postganglionic brachial plexus
Journal of Army Medical UniversityVol. 35, Issue 6, Pages: 553-557(2013)
作者机构:
重庆医科大学附属第一医院放射科
作者简介:
基金信息:
DOI:
CLC:R813;R445.2;R322.85
Published:2013,
Published Online:30 March 2013,
稿件说明:
移动端阅览
LI PENG, LU FAJIN, PENG CONG, et al. Comparison of IDEAL and 3D FSE-Cube sequences for normal postganglionic brachial plexus. [J]. Journal of army medical university, 2013, 35(6): 553-557.
DOI:
LI PENG, LU FAJIN, PENG CONG, et al. Comparison of IDEAL and 3D FSE-Cube sequences for normal postganglionic brachial plexus. [J]. Journal of army medical university, 2013, 35(6): 553-557.DOI:
Comparison of IDEAL and 3D FSE-Cube sequences for normal postganglionic brachial plexus
Objective To compare the sequences of iterative decomposition of water and fat with echo asymmetric and least-squares estimation (IDEAL) and 3D FSE-Cube imaging of
brachial plexus. Methods Brachial plexus of 30 health volunteers were imaged on 3.0T MR system with sequences of 3D FSE-Cube and IDEAL respectively. All of the volunteers underwent the same MR scanning for 2 times with 3D FSE-Cube sequence with the assistive device was used in the second time. Images were post-processed with MIP and CPR
and then the display rate was calculated
and image quality was assessed
signal noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. Results The display rates of the ganglion
supra clavicular nerve
and infra clavicular nerve on 3D FSE-Cube sequence were 73.33%
56.67% and 40.00% respectively
while the rates were 100.00%
97.67% and 93.33% respectively after use of the assistive device
with statistically difference between them (
P
<
0.05). The display rate of ganglion were 96.67% and 100.00% on IDEAL and 3D FSE-Cube respectively
and that of supra clavicular nerve was 86.67% and 96.67%
respectively
but there was no significantly difference in both of them (
P
>0.05). The infra clavicular nerve was shown better on 3D FSE-Cube than IDEAL (93.33%
vs
70.00%
P
<
0.05). In levels of ganglion
the imaging quality on IDEAL and 3D FSE-Cube had no significant difference(
P
>0.05). In the levels of interscalene area
costoclavicular space
and axillary
the quality of 3D FSE-Cube was better than IDEAL (
P
<
0.05). In the level of ganglion
the signals of IDEAL were higher than those of 3D FSE-Cube (
P
<
0.05). There was no significant difference in other level of brachial plexus between the 2 sequences (
P
>0.05). SNR and CNR were both higher in IDEAL than in 3D FSE-Cube (
P
<
0.01). Conclusion The sequences of IDEAL and 3D FSE-Cube both have the capacity of clearly displaying postganglionic brachial plexus
But the imaging quality of 3D FSE-Cube is better when combined with assistive device.
Expression of ferritin heavy chain 1 in neural differentiation of bone marrow mesenchymal stem cells
Application value of amide proton transfer imaging in differential diagnosis of brain metastasis and high grade glioma
Related Author
FU Yuanqiao1
QIN Yong1
ZHOU Yirui1
HU Jun2
CAI Jinhua1
LUO Jiaxin
ZHANG Gaofu
LI Wei
Related Institution
Department of Radiology, National Clinical Research Center for Child Health and Disorders, Key Laboratory of Child Development and Disorders of Ministry of Education, Chongqing Key Laboratory of Pediatrics, Children’s Hospital of Chongqing Medical University
Department of Neurology, First Affiliated Hospital, Army Medical University
国家儿童健康与疾病临床医学研究中心,儿童发育疾病研究教育部重点实验室,儿科学重庆市重点实验室3
重庆医科大学附属儿童医院:放射科1,肾脏内科与免疫学2
Department of Radiology, the First Affiliated Hospital of Chongqing Medical University