Reliability of shear wave elastography in evaluation on ocular and periocular tissues
返回论文页
|更新时间:2023-10-11
|
Reliability of shear wave elastography in evaluation on ocular and periocular tissues
Journal of Clinical Ultrasound in MedicineVol. 25, Issue 2, (2023)
作者机构:
上海市第一人民医院
作者简介:
基金信息:
DOI:
CLC:
Published:2023,
Published Online:6 March 2023,
扫 描 看 全 文
Reliability of shear wave elastography in evaluation on ocular and periocular tissues. [J]. Journal of Clinical Ultrasound in Medicine 25(2).(2023)
DOI:
Reliability of shear wave elastography in evaluation on ocular and periocular tissues. [J]. Journal of Clinical Ultrasound in Medicine 25(2).(2023)DOI:
Reliability of shear wave elastography in evaluation on ocular and periocular tissues
Objective: To assess the reliability of shear wave elastography (SWE) in evaluation on ocular structure of adult. Methods: Totally 33 healthy adult volunteers and 52 refractive myopia underwent SWE of eye region by different inspectors or same inspector in different time points. retina-choroid-sclera complex
optic nerve
retrobulbar fat tissue
internal rectus muscle and external rectus [ 通讯作者:?张志华,吴蓉 电子信箱:jundesan@163.com,wurong7111@163.com。 作者简介:李芳(1994—),女,硕士;电子信箱: 2403338915@qq.com。 ]muscle were examined
and the intra-observer and inter-observer reproducibility were assessed. Results: Paired test showed no significant difference between the two measured values made by same inspectors on the same day (P>0.05)
intra-class correlation coefficient(ICC) values of retina-choroid-sclera complex
optic nerve
retrobulbar fat tissue
Internal rectus muscle and external rectus muscle was 0.82、0.85、0.86、0.83、0.84
respectively.
indicating very good reproducibility. Paired test showed no significant difference between the two measured values made by same inspectors on the different days (P>0.05)
the intra-class ICC values of retina-choroid-sclera complex
optic nerve
retrobulbar fat tissue
Internal rectus muscle and external rectus muscle was 0.79、0.82、0.82、0.80、0.78
respectively.
indicating very good reproducibility. Paired test showed no significant difference between the two measured values made by different inspectors (P>0.05)
the intra-class correlation coefficient(ICC) values of retina-choroid-sclera complex
optic nerve
retrobulbar fat tissue
Internal rectus muscle and external rectus muscle was 0.71、0.75、0.79、0.73、0.75
respectively.
indicating very good reproducibility. Conclusion: SWE has good repeatability to obtain retina-choroid-sclera complex
optic nerve
retrobulbar fat tissue
Internal rectus muscle and external rectus muscle.
关键词
超声眼部组织眼周组织剪切波弹性成像重复性
Keywords
UltrasonographOcular tissuesPeriocular tissuesShear wave elastographyReproducibility of results
Research progress of echocardiography in evaluation of myocardial fibrosis
The application value of shear wave elastography in evaluating the treatment effect of testicular epididymitis caused by testicular appendage torsion in children
Conventional ultrasound combined with shear wave elastography to distinguish pleomorphic adenoma from Warthin tumor of parotid: comparative with MRI
Application of transperineal shear wave elastography combined with anterior chamber structure parameters in POP screening for puerperaeundergoing natural delivery
Related Author
周玮珺
杨敏
刘勇
徐平
吴兰英
王颖
肖榕
冷晨蕾
Related Institution
安徽医科大学第一附属医院
南京鼓楼医院
Department of ultrasound,Beijing Shijitan Hospital,Capital Medical University