A preliminary study of normal mitral valve structural parameters and their variation patterns with cardiac cycle by three-dimensional transesophageal echocardiography
返回论文页
|更新时间:2023-10-11
|
A preliminary study of normal mitral valve structural parameters and their variation patterns with cardiac cycle by three-dimensional transesophageal echocardiography
Journal of Clinical Ultrasound in MedicineVol. 24, Issue 6, (2022)
作者机构:
空军军医大学
作者简介:
基金信息:
DOI:
CLC:
Published:2022,
Published Online:5 July 2022,
扫 描 看 全 文
A preliminary study of normal mitral valve structural parameters and their variation patterns with cardiac cycle by three-dimensional transesophageal echocardiography. [J]. Journal of Clinical Ultrasound in Medicine 24(6).(2022)
DOI:
A preliminary study of normal mitral valve structural parameters and their variation patterns with cardiac cycle by three-dimensional transesophageal echocardiography. [J]. Journal of Clinical Ultrasound in Medicine 24(6).(2022)DOI:
A preliminary study of normal mitral valve structural parameters and their variation patterns with cardiac cycle by three-dimensional transesophageal echocardiography
Objective: The quantitative mitral valve technique of three-dimensional transesophageal echocardiography was applied to analyze the range of structural parameters of the mitral valve with normal intracardiac structure and function
and summarize their variation patterns with cardiac cycle in order to provide a theoretical basis for the assessment of mitral valve lesions. Methods: 2D images and real-time 3D full-volume images of the mitral valve were collected from 37 subjects with normal intracardiac structure and function
and the images were complete including mitral and aortic valve structures. The data include 18 parameters such as the annular diameter
annular height etc. All the data were recorded in 6 time phases of early
mid
and late diastole and early
mid
and late systole. analyze these data and determine their value reference ranges and analyze their motion regularity. Results: The reference range of mitral valve structures in the 37 subjects was summarized and found to vary regularly with cardiac cycle: some parameters have their maximum values occurring in diastole and their minimum values occurring in systole
such as anterior-posterior and left-right annular diameters
annular area
annular circumference
anterior and posterior annulus length etc. Another parameters have their maximum values occurring in systole and their minimum values occurring in diastole
such as annular height
nonplanar angle
AO-MV angle
and AO-MV midpoint distance. Unlike the previous structures
the maximum and minimum values of Tenting height and Tenting area for the 3 subdivisions of the valve leaflets appeared in early and late systole
respectively. Conclusion: The automated quantitative analysis system of three-dimensional transesophageal echocardiography can reconstruct the mitral valve model relatively quickly and intuitively
and automatically measure several structural parameters. In this study
the time phases of the maximum and minimum values of each parameter are analyzed and helped to deepen the understanding of the mitral valve structure and its motion pattern. The range of reference values summarized in this study provides a theoretical reference for determining mitral valve lesions. and lays the foundation for further research on the structural changes of mitral valve lesions.