OBJECTIVE: To develop a method for simultaneous determination of irinotecan (CPT-11) and its active metabolite 7-ethyl-10-hydroxycamptothecin (SN-38) in human plasma. METHODS: After precipitated by acetonitrile and acidified with hydrochloric acid, using camptothecin as internal standard, the plasma sample was determined by HPLC-FLD. The determination was performed on Waters Luna C18 column with mobile phase consisted of 0.05 mol/L sodium dihydrogen phosphate-acetonitrile (70 ∶ 30,V/V, adjusted pH to 4.0 by phosphoric acid) at flow rate of 1 mL/min. The excitation wavelength was set at 380 nm; the emission wavelengths of CPT-11 and SN-38 were set at 480 nm and 535 nm, respectively. The column temperature was 25 ℃ and the sample size was 20 μL. RESULTS: The linear ranges were 200-1 000 ng/mL for CPT-11 (r=0.999 4,n=5) and 5-45 ng/mL for SN-38 (r=0.999 2,n=5). RSDs of inter-day and intra-day were 1.68%-5.57%. The relative recoveries of CPT-11 and SN-38 were 90.12%-106.93%(RSD<8%,n=5) and 92.07%-102.56%(RSD<6%,n=5); the extraction recoveries of CPT-11 and SN-38 were 72.23%-86.56%(RSD<6%,n=5) and 71.98%-83.44%(RSD<7%,n=5),respectively. The plasma concentrations of CPT-11 and SN-38 in 5 patients with colon cancer were 431.13-617.19, 13.97-31.89 ng/mL(1 h after intravenous dripping) and 398.14-584.43, 11.61-29.94 ng/mL(2 h after intravenous dripping). CONCLUSIONS: The method is simple, rapid, sensitive, reproducible and suitable for the determination of plasma concentration and pharmacokinetic study of CPT-11 and its metabolite SN-38.