OBJECTIVE: To develop the method for the determination of amoxicillin concentration in dog plasma, and to study its pharmacokinetic characteristics. METHODS: Totally 12 Beagle dogs were collected and given Amoxicillin capsules 250 mg intragastrically. The 1.0-1.5 mL plasma samples were collected before medication, and 0.5, 1, 1.5, 2, 2.5, 3, 3.5, 4, 4.5, 5, 5.5, 6, 8, 10 h after medication. Using amoxicillin-d4 as internal standard (IS), after precipitated with acidic acetonitrile, the plasma concentration of amoxicillin was determined by LC-MS/MS. The separation was performed on an Atlantis dC18 column with mobile phase consisted of acetonitrile (containing 0.1% formic acid)-10 mmol/L ammonium formate water solution (pH 4.5, gradient elution) at the flow rate of 0.3 mL/min. The column temperature was 30 ℃, and sample size was 1 μL. Mass spectrum condition included ESI, positive ion mode, multiple reaction monitoring, ion source spray voltage 5.5 kV, ion source temperature 450 ℃, atomizing gas pressure 30 psi, heater pressure 40 psi, ion pair m/z 366.2→113.9 and m/z 370.2→353.0 for amoxicillin and IS, fragmentation voltage 26 and 16 V, residence time 200 ms. RESULTS: The linear range of amoxicillin was 0.1-16 μg/mL (r=0.998 2). RSDs of intra-day and inter-day were lower than 14.3% and 11.8% (n=5). The accuracy was 93.0%-112.0% (n=3). The extraction recoveries were 58.9%-63.3% (RSD=3.3%, n=3), and matrix effect was 2.8%(n=6). RSDs of stability test was lower than 15% (n=3). By single compartment model, the pharmacokinetic parameters of amoxicillin included that t1/2 was 1.56 h, AUC1-10 h was 24 300 ng·h/mL and cmax was 7 523 ng/mL. CONCLUSIONS: The method is sensitive, accurate, simple and rapid. It is applied for the determination of amoxicillin concentration in the plasma of dog and the pharmacokinetic study of amoxicillin.