Comparison of ultrasound vector flow imaging based on Doppler and speckle tracking
Objective To comparatively evaluate the performance of vector flow imaging based on Doppler and speckle tracking (ST). Methods Based on ultrafast ultrasound imaging technique with multi-angle plane wave transmission
vector flow imaging was performed using Doppler and ST methods
respectively. Performance of the 2 methods was assessed in simulation
phantom
and in vivo study. Results For a simulated artery
the bias of the horizontal velocity component was 3.15% and 4.36% in Doppler and ST methods
respectively. The bias of vertical velocity component was 1.79% and 1.00%
respectively. The bias of the 2 methods was similar with each other. In the phantom experiment with the maximal velocity of 100 cm/s
the bias was -5.00 % and -5.28 %
respectively for the horizontal velocity component
and was 0.53 and 1.46 cm/s respectively for the vertical velocity component. Blood flow estimated with ST method had relatively larger bias
and the peak velocity appeared to be underestimated. In the in vivo experiments
estimated velocity in ST method was smaller than that in Doppler method. Conclusion In simulation
both Doppler and ST methods can achieve considerably accurate vector flow measurements. In the phantom and in vivo studies
the estimated velocity in ST method was smaller than the velocity in Doppler method.