Biocompatibility of MBT straight wire appliance with nitrogen-doped titania TiO2-xNx film
) film on MBT straight wire appliance and to evaluate its biocompatibility. Methods Radio-frequency magnetron sputtering was applied to produce TiO
2-x
N
x
film on the surface of raw material of MBT straight wire applian
ce. Biocompatibility was evaluated as follows: (1) The film material (raw material of MBT straight wire appliance with TiO
2-x
N
x
film) and the control material (raw material of MBT straight wire appliance) were fixed on the surface of rat oral mucosa
which were examined macroscopically and histologically after 2 weeks. (2) Mice were given the extracted liquids of the film material and the control material and normal saline by lavage for 1 week
respectively. The clinical symptoms were observed
body weight was recorded and viscera were dissected and observed in the mice to evaluate the short-term systemic toxicity of the film material. (3) The hemolysis of the film material was evaluated with rabbit blood in vitro. (4) Mice were given the extracted liquids of the film material and the control material and normal saline by lavage twice in 24 h
respectively
and the bone marrow slides of the mice were prepared to evaluate the teratogenicity and mutagenicity of the film material. Results The appetite and movement of the rats were normal
and the histological morphology of rat oral mucosa of the film material group showed no significant difference from that of the control material group
which suggested the film material had no oral mucosa irritation response. The growth rate of mouse body weight was not significantly different between the film material group and the control material group (P0.05)
and neither toxic signs nor apparent abnormality of pathological examination was found
revealing that the film material had no obvious short-term systemic toxicity. The hemolytic rate was less than 5%
and the breaking up of red blood cells was not found in the film material group
indicating that the film material had no hemolytic reaction. The micronuclear rates in the polychromatic erythrocytes (PCE) of the mouse bone marrow of the negative control group
the film material group and the control material group were 1.48
2.08 and 2.16
respectively
which were all less than 5
but th
ere was no significant difference between the three groups (P0.05). Conclusion MBT straight wire appliance with nitrogen-doped titania TiO
2-x
N
x
film has good biocompatibility without oral mucosa irritation response