Tan Fabing,Jin Ping,Wu Shuhong,Zhang Wei.Effect of optical imaging powder on accuracy of digital model of full crowns[J].J Amry Med Univ (J Third Mil Med Univ),2012,34(13):1336-1339.
Tan Fabing,Jin Ping,Wu Shuhong,Zhang Wei.Effect of optical imaging powder on accuracy of digital model of full crowns[J].J Amry Med Univ (J Third Mil Med Univ),2012,34(13):1336-1339.DOI:
Effect of optical imaging powder on accuracy of digital model of full crowns
Objective To determine the effect of optical imaging powder on the accuracy of digital model of full crowns in Cerec inlab MC XL system. Methods A upper first right molar in the standard model were prepared
and 8 scanned models made of esthetic-base gypsum were duplicated
of which each model including digital models were digitally imaged by scanning in inEos unit before and after powder spraying. The changing paradigm of each digital model at its marginal area
shoulder area
axial area and pit and fissure of occlusal areas were measured by inlab 3D software (version 3.65). Finally
the results were analyzed statistically by SAS programs. Results No statistical difference was observed between the two reports of the digital model before and after a week (P0.05) in terms of measured values. The resolution of digital model after powder spraying was better than before
and optical imaging powder had certain effect on the model’s marginal area
shoulder area
axial area
and pit and fissure of occlusal areas. The floor impact counter of them were 0.073±0.042
0.106±0.077
0.044±0.025
0.128±0.060 and 0.102±0.056 mm respectively. Such differences were considered with statistical significance (P0.01). Conclusion Optical imaging powder has different effect on the accuracy of digital model in different area
which shows stronger on shoulder and occlusal area than on marginal and axial area. It is suggested that different regional spray powder should be treated differently.