Effect of irradiation on content of Norepinephrine Bitartrate Injection
Objective To develop a high performance liquid chromatography (HPLC) method to analyze the content of norepinephrine bitartrate in Norepinephrine Bitartrate Injection
and study the content changes of the injection after exposure to γ-ray irradiation at different intensities. Methods Norepinephrine Bitartrate Injection of the same batch was under the exposure to
60
Co-γ ray radiation at 1
2
4
8
16
and 25 kGy by one time. After that
0.14% heptanesulfonic acid sodium salt solution (pH value adjusted to 3.0 by using phosphate): methanol (65∶35) as mobile phase
flow
rate 1 ml/min
detection wavelength 286 nm
injection volume 20 μl were used to determinate the norepinephrine bitartrate content. Results The linear range of norepinephrine bitartrate was 5 to 25 μg/ml (r=0.999 9). the yield rates at low
medium and high concentrations of norepinephrine bitartrate (8
10 and 12 μg/ml) was 99.685%
96.366%
and 97.557%
respectively. The inter-day (n=3) and intra-day precision (n=3) were less than 2%. The appearance and clarity of the Injection had no significant change
but the content of norepinephrine bitartrate was decreased in an irradiation dose-dependent manner (P0.05). Especially
when the dose was above 8 kGy
and the content was even lower than the range of acceptable quality. Conclusion Our developed HPLC method can be used to determine the active ingredient content in Norepinephrine Bitartrate Injection. Irradiation has a significant impact on the stability of Norepinephrine Bitartrate Injection.